INSTRUCTION MANUAL
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BST-CHK01 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Besantek BST-CHK01

  • Page 1 BST-CHK01 BESANTEK TEST BOX Functional checking for RCD testers, Loop testers and Insulation/Continuity testers INSTRUCTION MANUAL 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentDepot.com...
  • Page 2: Table Of Contents

    Index Page 1. Warning and safety..........2. Specification ............3. Safety notes............4. Checking a RCD test.......... 5. Checking an installation or insulation tester..6-10 Checking for the correct applied test voltage............6-7 Checking the reading accuracy....... Checking the continuity measurement……..8-9 Checking for the AC voltage measurement function............
  • Page 3: Warning And Safety

    BESANTEK 1. Warnings and safety 1. Read the following safety information carefully before attempting to operate or service the test box. 2. The test box should be operated only by a professional with understanding of the test box. 3. Do not expose the test box to direct sunlight, high temperature and humidity.
  • Page 4: Specification

    2. Specification RCD Testing 10mA±4mA Trip,14mA↑ over,150mS±10% Range 30mA±12mA Trip,42mA↑ over,150mS±10% 150mA±60mA Trip,210mA↑ over,30mS±10% Insulation Testing Range 1MΩ / 9.9MΩ / 99MΩ Accuracy ± 1% Max. Testing 1.2kV Continuity Testing Range 0.5Ω / 2Ω / 10Ω Accuracy ± 0.05Ω / ± 0.1Ω / ± 0.2 Ω Max.
  • Page 5: Safety Notes

    BESANTEK 3. Safety notes • Rated environmental conditions: 1. Indoor use. 2. Installation Category Ill. 3. Pollution Degree 2. Altitude up to 2000 Meter. 5. Relative Humidity 80% Max. ° ° 6. Ambient Temperature 0 C-40 • Observe the international electrical symbols listed below: ri=il Meter is protected throughout by double insulation or L==J reinforced insulation.
  • Page 6: Checking A Rcd Test

    4. Check RCD tester A. CHECKING A RCD TESTER RCD tester simulates a current between live and earth to trip TEST BOX. The disconnect time of relay is shown on RCD tester usually in milli-seconds if TEST BOX and RCD tester select the correct current and the correct time range.
  • Page 7 BESANTEK B. CHECKING RCD TEST TIME & CURRENT Plug the TEST BOX into an unprotected supply (no RCD trip) and plug the RCD tester into the left- h and socket marked 'RCD TESTER' on the TEST BOX is shown as below.
  • Page 8: Checking An Installation Or Insulation Tester

    BESANTEK Select the 150mA or 30mA with "x5I" button (fif available). Set the TEST BOX to 150mN30mS and press the "test" button. Record the time displayed on the tester. Allow from 27mS to 33mS, ± 1mS. Tripped is approximate to 45mA tolerance (150mA± 45mA) Over current is over 195mA.
  • Page 9: Checking The Reading Accuracy

    Select the lowest test voltage (100V or 250V) range on the tester. Press the “test” button on the tester. Note the voltage LED (100V or 250V) on the TEST BOX. It should be lit with the same range on the tester. Record the reading on the tester. Repeat the same step on the other test voltage ranges.
  • Page 10: Checking The Continuity Measurement

    BESANTEK C. CHECKING THE CONTINUITY MEASUREMENT The continuity or low resistance measures the resistance in TEST BOX has three low value resistors for p recision measurement. Many testers have a null function to zero out this leads resistance exactly before making measurement. The instruction shall show how to use this function.
  • Page 11: Checking For The Ac Voltage Measurement Function

    Select the fitting range on the tester in order to read precise measurement from each range on the tester. For example, if the tester has a 200Ω, 20Ω and 2Ω ranges. Use the 200Ω range to measure “10Ω”, the 20Ω to measure “2Ω” and the 2Ω...
  • Page 12 BESANTEK Connect the TEST BOX to an unprotected supply. Connect test leads of installation tester to the AC OUTPUT on the TEST BOX as shown below. AC IN 50Hz 210-250V .� l ;,' ;t ,!' & l23DVI: O· ·@ -:,�:I Record voltage value displayed on meter.
  • Page 13: Checking A Loop Tester

    BESANTEK 6. Checking a LOOP tester Loop tester is an instrument of measuring the resistance of the main supply. Its measurement is made in Ohm. Plug the TEST BOX into an unprotected supply (no RCD trip) and plug the loop tester into the central socket marked 'WIRING CHECK' on the TEST BOX is shown as below.
  • Page 14: Limited Warranty

    Common Problems Solution Supply trips out when test TEST BOX must plug into button pressed unprotected supply. A: Faulty loop tester main supply. No power to Loop Tester B: TEST BOX does not plugged into AC supply. Noise on main supply. Try a No reading on tester different main Socket for the TEST BOX.
  • Page 15 All warranties implied by law are hereby limited to a period of twenty four months, and the provisions of the warranty are expressly in lieu of any other warranties expressed or implied. The purchaser agrees to assume all liability for any damages or bodily injury which may result from the use or misuse of the product by the purchaser, or its user, his employees, or others, and the remedies provided for in this warranty are expressly...