Note:
For thermal stabilization of Pockels cell and harmonic generators, the
system software will not allow the SHUTTER button to open for 30 minutes
after the red AC POWER toggle is first turned on.
4. Measure power coming from the exit port and compare with the value for total
energy indicated in QA.
5. If power is comparable then proceed with optimizing harmonic crystals. See
Harmonics optimization on page 31.
6. If energy is low, see Section 6: Troubleshooting on page 71.