Continuum Surelite I 10 Operation And Maintenance Manual page 33

Table of Contents

Advertisement

Note:
For thermal stabilization of Pockels cell and harmonic generators, the
system software will not allow the SHUTTER button to open for 30 minutes
after the red AC POWER toggle is first turned on.
4. Measure power coming from the exit port and compare with the value for total
energy indicated in QA.
5. If power is comparable then proceed with optimizing harmonic crystals. See
Harmonics optimization on page 31.
6. If energy is low, see Section 6: Troubleshooting on page 71.
996-0207
Section 2: Installation
23

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Surelite I 10 and is the answer not in the manual?

Table of Contents

Save PDF