Fault symptom
• Analyser just continually
beeps
• BAT only shows 65 with
fully charged NiMH
batteries fitted
Causes / Solutions
• Turn dial back to MENU and press ENTER
• This is not a problem and is to be expected as
NiMH batteries only deliver 1.25 V per cell
whereas Alkalines deliver 1.5 V per cell.
Fresh alkalines might give a BAT value of 90
or so.