If the device is not used for several days running, follow the recommendations indicated
below:
1 Carry out the cleaning cycles for the water circuits using the CLEAN function (see
point "06.1").
2 Disconnect the device from the power mains.
3 Carry out cleaning, disinfection and sterilisation (see Section "06.0")
DANGER: Check regularly that the power cable is intact. If it is damaged, replace it
with an original Mectron spare.
DANGER : Switch off the apparatus.
Always turn off the apparatus by means of the switch (Fig.5 - Ref.F) and disconnect it from
the power outlet before carrying out the following maintenance activities.
1 Insert the flat tip of a screwdriver into the recess in the fuse compartment below the
power socket and use it as a lever (Fig.11 - Ref.A).
2 Pull out the fuse compartment (Fig.11 - Ref.B).
3
DANGER: Replace the fuses, using fuses of the type indicated on the identifica-
tion label on the bottom of the apparatus. Depending on the mains power supply,
these fuses may be rated as follows:
230 VAC
500 mA T
115 VAC
1
100 VAC
1,2 mA T
4 Put the compartment back into place (Fig.11 - Ref.B).
Piezosurgery // GB
07.0 - REGULAR MAINTENANCE
08.0 - REPLACEMENT OF THE FUSES
mA T
A
07.2 - Power-supply Cable
B
Fig. 11
MANUAL OF USE AND MAINTENANCE
07.1 - Shelf Storage
23 / 32
Need help?
Do you have a question about the Piezosurgery II and is the answer not in the manual?