Programming Instructions

Programming the Newly selected Settings

Programming the RNS
the neurostimulator. The P
new settings can be programmed into the neurostimulator.
1. Select the P
2. Select R
EVIEW
3. Select the P
ECoG storage triggers are inactive during programming. Programming clears activity
Note:
information and ECoGs stored in the neurostimulator. If there is any new
information, be sure to interrogate the neurostimulator prior to programming.
S
EC
G S
ETTING UP
O
TORAGE
Electrocorticograms (ECoGs) can be stored by the RNS
trigger occurs. The maximum number of ECoGs that can be stored is affected by the number of
channel(s) selected ON for storage, and the capture duration. The oldest stored ECoGs are
continually replaced by the newly stored ECoGs, except as noted below.
The number of ECoGs stored for a particular ECoG storage trigger may be reserved. Reserved
ECoGs are only overwritten by newer ECoGs of the same ECoG storage trigger type. ECoG storage
may also be scheduled at four different times on a 24-hour clock.

Selecting ECoG Storage Settings

1. Go to the C
2. Select the EC
3. Select the C
and stored.
4. Determine the desired length (in seconds) of stored ECoGs.
Figure 23: Illustration of a stored ECoG with a 120 second pre-trigger and a 60 second post-trigger period.
5. Select the ECoG storage T
L
E
ONG
beyond a preset duration (L
detection is ENABLED.
P
ATTERN
with a characteristic pattern (Pattern A or Pattern B) is detected. Selecting one of these
®
RNS
System User Manual
®
Neurostimulator sends all newly selected settings from the programmer to
S
ROGRAM
ETTINGS
S
tab.
ROGRAM
ETTINGS
C
to view changes prior to programming.
HANGES
button.
ROGRAM
S
tab.
ONFIGURE
ETTINGS
G S
tab.
O
TORAGE
(
) (CH1 through CH4). Only the channels selected ON will be recorded
HANNEL
S
RIGGERS
triggers initiate ECoG storage when a detected episode continues
PISODES
ONG
A and P
B triggers initiate ECoG storage when an episode beginning
ATTERN
tab is the only location on the programmer from which
®
Neurostimulator when an ECoG storage
that will initiate storage of an ECoG.
E
L
). This trigger can be used whenever
PISODE
ENGTH
63