Responsive Neurostimulator System
System. This could result in loss of therapy, and additional surgery to remove or
replace components of the RNS
administering physician should consult with the physician prescribing the RNS
System.
WARNING:
R
ADIATION
The effects of high radiation sources (such as cobalt 60 or gamma radiation used in
cancer therapy) on the RNS
of radiation may damage the RNS
additional surgery to remove or replace components of the RNS
course of radiation therapy, the radiation oncologist should consult with the physician
prescribing the RNS
WARNING:
E
LECTROLYSIS
The effects of electrolysis on the RNS
the head or neck should be avoided.
WARNING:
C
OMPUTERIZED
For CT procedures on a patient with an implanted RNS
should:
• Ask the patient to have the neurostimulator temporarily shut off with a programmer
while the scan is performed, if possible.
• Minimize x-ray exposure to the implanted electronic medical device by:
• Using the lowest possible x-ray tube current consistent with obtaining the required
• Making sure that the x-ray beam does not dwell over the device for more than a
Important note: For CT procedures that require scanning over the medical device
continuously for more than a few seconds, as with CT perfusion or interventional
exams, attending staff should be ready to take emergency measures to treat adverse
reactions if they occur.
After CT scanning, the operator should:
• Ask the patient to have the neurostimulator turned back on with a programmer if it
had been turned off prior to scanning.
• Advise the patient to contact their healthcare provider as soon as possible if they
have questions or suspect their device is not functioning properly after any medical
procedure.
®
RNS
System User Manual
®
®
System.
T
OMOGRAPHY
image quality.
few seconds.
®
System. Prior to any administration of lithotripsy, the
System have not been studied. Exposure to high levels
®
System. This could result in loss of therapy, and
®
System have not been studied. Electrolysis on
(CT) S
CANS
®
®
System. Prior to any
®
Neurostimulator, the operator
13