Programming Instructions
Creating a Montage
1. Go to the C
2. From the M
selecting either a non-inverting (+) or inverting (-) input. Each electrode can be assigned to
one or two channels. Use a positive (+) input for electrodes assigned to 2 channels.
3. Assign a G
AIN
The ECoG signal amplification may be assessed following programming by viewing the real-
time ECoGs.
Note:
Using a negative (-) input for electrodes assigned to 2 channels may result in ECoG
artifacts.
Recommended Initial Montage settings
Program the initial ECoG sensing montage on the M
CH 1 = Lead 1: Electrode 1 +, Electrode 2 -
•
CH 2 = Lead 1: Electrode 3 +, Electrode 4 -
•
CH 3 = Lead 2: Electrode 1 +, Electrode 2 -
•
CH 4 = Lead 2: Electrode 3 +, Electrode 4 -
•
All G
settings on "HIGH"
•
AIN
Figure 22: Montage tab
®
RNS
System User Manual
S
tab.
ONFIGURE
ETTINGS
screen, assign an electrode to each C
ONTAGE
for each channel by selecting a high, medium-high, medium-low, or low value.
(CH1 through CH4) by
HANNEL
tab as follows:
ONTAGE
62