Epic Hygiene utilizes a solid state diode as a semiconductor source for invisible infrared radiation. The energy is delivered to the treatment site via a flexible fiber connected at one end to the laser source and the other end to the Handpiece.
Contraindications, Warnings, and Precautions 3.1 CONTRAINDICATIONS All clinical procedures performed with Epic Hygiene must be subjected to the same clinical judgment and care used with traditional techniques. Patient risk must always be considered and fully understood before clinical treatment. Practitioners must completely understand the patient’s medical history prior to treatment and exercise caution for general medical conditions that might contraindicate a local procedure;...
Prior to use, inspect the eyewear for pitting and/or cracking. Replace if damaged; do not use. CAUTION: The specifications for the protective eyewear provided with the Epic Hygiene laser system are marked on the side of the glasses. Always check the eyewear specifications marked on the glasses to ensure they offer the required protection for the specific laser wavelength.
Make appropriate power, pulse length, and interval adjustments to compensate for varying tissue compositions, density, and thickness. Always start treatment at the lowest power setting for the specific indication and increase as required. BIOLASE assumes no responsibility for parameters, techniques, methods or results.
Equipment Description 5.2 GENERAL The Epic Hygiene system consists of three components: Base Console Fiber Delivery System LED Indicators LED Indicators Control Button Fiber Storage Channel Handpiece Fiber Optic Cable Wireless Footswitch Figure 5.1 Epic Hygiene User Manual...
The Handpiece (Figure 5.3) is re-usable. It is a conduit for delivering the optical energy produced by the laser to the treatment site and must be cleaned and sterilized between patients to avoid cross-contamination (See Section 9). Figure 5.3 Epic Hygiene User Manual...
Equipment Description WIRELESS FOOTSWITCH The Epic Hygiene laser will only emit laser energy when the user presses down on the Footswitch Cover while the laser is in Ready mode. It is designed to work using wireless technology and is powered by two AAA batteries which are installed prior to shipment from the factory;...
• Always place the system into Standby mode (by pressing the control button while in Ready mode) before exchanging Handpieces or disposable Tips. • Always ensure that the proper laser parameters are set before the Epic Hygiene laser is used in a procedure.
Safety with the Epic Hygiene LASER WARNING: DO NOT open the console housing at any time. Danger from optical radiation may exist. CAUTION: DO NOT position this equipment so that it is difficult to pull the plug from the power source.
“LASER CURRENT HIGH/LOW”. The laser console will not operate until the system first clears the error and then goes into Ready mode. If the error message persists, please contact BIOLASE Service at 1-800-321-6717 or your BIOLASE-authorized service representative. Figure 6.1: Power Switch, DC...
The system color display with touch screen and LED indicators on the control panel shows the functional conditions of the system. Figure 6.5: Emergency Laser Stop CAUTION: Use only the DC Power Supply Module supplied with the Epic Hygiene laser system. (BIOLASE P/N 2400075) Epic Hygiene User Manual...
7.2 CONNECTING THE FIBER OPTIC CABLE The Epic Hygiene ships with the Fiber Optic Cable already connected to the laser console. CAUTION: Do not connect or disconnect the fiber cable while the laser console is turned ON.
(Figures 7.3, 7.4). If no click is audible, disconnect the Handpiece and then re-install it. Protective Cap Fiber Shaft Handpiece Collar Figure 7.3: Connecting the Handpiece to the Fiber Optic Cable Figure 7.4: Handpiece Assembly fully assembled Epic Hygiene User Manual...
(Figure 7.9). Remove the fiber Tip by twisting the Tip counterclockwise (Figure 7.7). Figure 7.6: Insert the fiber Tip into the Handpiece (only when the Handpiece is connected to the fiber) and twist clockwise until snug Epic Hygiene User Manual...
Page 22
LASER WARNING: Never operate the laser without a fiber Tip attached to the Handpiece. Never point the laser at a person’s eyes. All persons present in the operatory must wear protective eyewear when the laser is in use. Epic Hygiene User Manual...
Turn the power switch at the rear of the console to the ON ( I ) position. The “BIOLASE” logo screen will appear. After three (3) seconds the Epic Hygiene “Welcome” screen will be displayed.
LED indicator light on the laser console will blink when pairing is established. The laser and Footswitch are shipped already paired. However, if pairing is not confirmed, an "x" will appear over the pairing icon located in the upper left hand corner of the touchscreen. Figure 7.20 Epic Hygiene User Manual...
(2) seconds will allow the transition from Standby or Ready mode to Sleep mode. Note that you will not be allowed to go into Ready mode unless you have chosen a treatment module on the HOME screen. Figure 7.25 Figure 7.26 Epic Hygiene User Manual...
ALERT 2, indicating the system is not in Ready mode. Press the control button to switch to Ready mode. When the Footswitch is not in use, it will go into Sleep mode to conserve battery power. It automatically reactivates when pressed. Epic Hygiene User Manual...
(Figure 7.28). By pressing the "X" on the bottom left of the screen the user may cancel the Pulse mode selected. Figure 7.27 NOTE: Operating the laser at a shorter pulse duration typically results in lower tissue temperature. Figure 7.28 Epic Hygiene User Manual...
Installation and Set-Up 7.15 PROCEDURES BUTTON The Epic Hygiene has the ability to store up to 12 pre-set procedures. The laser has 6 pre-programmed procedural presets installed and 6 empty slots for custom use. Pre-programmed settings can be customized to your preference.
Clinical Applications 8.1 OVERVIEW To efficiently treat tissues, it is imperative to understand the nature of the Epic Hygiene device. Please review this section carefully, practice on model tissues, and attend a diode laser training session before using this device in a clinical situation within the applicable scope of practice.
Use disinfectant to wipe down the front panel and Handpiece Holder of the Epic Hygiene system after each procedure. Do not use bleach or abrasive cleansers.
1. Place the Handpiece and fiber Tips in separate single-wrap, self-seal autoclave pouches. 2. Place on an autoclave tray; do not stack other instruments on top of the pouches. 3. Place the tray inside the autoclave chamber and set the appropriate cycle as recommended in Figure 9.1. Epic Hygiene User Manual...
Always disinfect the fiber between patients by wiping it completely with an appropriate disinfecting solution such as Cavicide™ or a similar quaternary ammonium compound product (containing 20% alcohol or less), and follow the manufacturer’s instructions. DO NOT AUTOCLAVE. Epic Hygiene User Manual...
4. Repeat as needed using a new Mirocbrush each time until no residue appears on the swab. Figure 9.4 CAUTION: If blemishes appear below or inside the lens, contact BIOLASE or your authorized service representative to order a replacement lens. Figure 9.5 Epic Hygiene User Manual...
6. Dispose of the used Lithium Ion battery as regulated. Do not discard in a trash bin. Figure 9.6 NOTE: Only use the battery pack supplied by BIOLASE. The battery pack is a separate accessory (BIOLASE P/N 6400457). Epic Hygiene User Manual...
Figure 9.7 battery cover screws Figure 9.8 pairing button (internal view) NOTE: To ensure the longevity of the battery power, only BIOLASE-supplied batteries are recommended as replacements (BIOLASE p/n 2400203); these are industrial-grade batteries which under normal use have a longer life than conventional AAA batteries.
Cover the unit when not in use for extended periods of time. Store the system in a place where it will not be accidentally bumped or banged. The Epic Hygiene is shipped inside a custom shipping box. Please save and store the box in a cool, dry place for use when transporting the laser, or for long-term storage.
Service at 1-800-321-6717 or your Authorized Service Representative to schedule an appointment. Software Specification BIOLASE respects the intellectual property of others, and we ask our users to do the same. Epic Hygiene software is protected by copyright and other intellectual property laws. This product contains proprietary, copyrighted software developed by BIOLASE, Inc.
NOTE: For any on-screen message not listed in Figure 12.1, re-power the laser console; if the message does not clear, call BIOLASE Service at 1-800-321-6717 or your Authorized Service Representative. Screen...
Page 40
Check remote interlock closed Error 12 Battery critically low Battery is critically low Plug in DC supply Error 13 Internal error Internal error occurred Restart unit Footswitch battery Error 14 Footswitch battery Replace Footswitch battery critically low Figure 12.2 Epic Hygiene User Manual...
Appendix A Tip Guide NOTE: All BIOLASE tips for diode lasers are sold non-sterile and are for single-use only. See Section 9 in this User Manual for cleaning and sterilization instructions. NON-INITIATED Length (mm) Part Number Name Diameter E4-4 400µm...
Tip will glow (Figure 3). 4. Repeat initiation process as needed to ensure the Tip is initiated. Figure 2 Figure 3 Please follow local/state regulations to ensure the procedures are within your scope of practice. Epic Hygiene User Manual...
The applied part is not conductive to the patient EMERGENCY LASER STOP SWITCH: The switch used in emergencies to stop laser output LOCATION: Left side of laser console FIBER WARNING: Indicates the system contains a laser LOCATION: Back of laser console Epic Hygiene User Manual...
Page 45
The Footswitch and laser console comply with Part 15 of FCC Rules regarding unlicensed transmissions LOCATION: Bottom of Footswitch INGRESS PROTECTION CODE: IPX6 The Footswitch is water-resistant, protected from splashes of water SINGLE USE ONLY – DO NOT REUSE Epic Hygiene User Manual...
Page 47
DC POWER, USB, REMOTE INTERLOCK LABEL: Identifies input ports POWER INPUT RATING: 12 Volts Direct Current, 5 amps MINI USB INPUT: For external programming REMOTE INTERLOCK: Input for Remote Interlock Connector Epic Hygiene User Manual...
Contact your sales location or BIOLASE if any of these problems are observed. 4. Do not place the batteries in microwave ovens, high-pressure containers, or on induction cookware.
Page 49
• The temperature range over which the battery can be discharged is -20°C to 60°C. Use of the battery outside of this temperature range may damage the performance of the battery or may reduce its life expectancy Epic Hygiene User Manual...
DC Power Supply Module, BIOLASE p/n 2400075 Wireless Footswitch, BIOLASE p/n 6400541 WARNING: The use of accessories other than those supplied or sold by BIOLASE, Inc. as replacement parts for internal or external components may result in increased EMISSIONS or decreased IMMUNITY of the Epic Hygiene diode laser system.
Page 51
Appendix F Electromagnetic Compatibility Epic Hygiene User Manual...
Page 52
Appendix F Electromagnetic Compatibility Epic Hygiene User Manual...
Page 53
Appendix F Electromagnetic Compatibility Epic Hygiene User Manual...
• Increase the separation between the equipment and receiver • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected • Consult the dealer or an experienced radio/TV technician for help Epic Hygiene User Manual...