  • EN

Available languages

  • ENGLISH, page 1
6. View the gas analysis values from the 6 different pages
Change page on display pressing
Zoom in and out by pressing.
7. Print the report
All measured and calculated values can be printed.
Press:
Report includes all analysis data, the draft and
smoke index values if performed.
The CO dilution system can be enabled or disabled from the Configuration
page.
Press the
key to open the configuration menu.
Press "
NEXT
"
Select and confirm the "Dilution" option
Select the desired option:
AUTO
(the dilution automatically start when the CO
concentration is higher than 2000 ppm)
ON
(the dilution is always enabled)
OFF
(the dilution is always disabled)
Tflow/Treturn can be measured using the standard thermocouple input.
Connect the Tc K to the instrument.
Select and confirm the "
Press:
Select Flow/Return parameter for changing, pressing
Press
F1
"Copy" to copy the actual temperature to the Flow/Return parameter.
Execute the steps above for other parameter in the same way.
Press
to print the values.
Press
F2
to exit to the previous menu.
Two pressure network leak test can be performed using a pressure decay
program.
The supplied syringe with the optional leak test kit is intended for
THE USAGE OF A PRESSURE HIGHER THEN ±300 HPA WILL DAMAGE
THE THE PRESSURE SENSOR OF THE INSTRUMENT PERMANENTLY.
Select and confirm the "
Press:
A. Test A - one minute Let-by test
Select and confirm the "
Change and confirm test type to "Test A" for one minute let-by test.
Press the "ZERO" function (F1) to zeroing the pressure sensor.
Connect the pressure input
Connect the kit to the gas network. If required, use the rubber blanking
plugs for connecting.
Open the valve on "T" connector.
Pressurize the network at the testing pressure.
Close the valve on "T" connector.
Select the "START" option pressing
CO Dilution
Temperature
TEMPERATURE
Leak Test
applications in Italy and Germany only.
ATTENTION
TIGHTNESS
Test type
P1
to the gas leak test kit (cat. 26028).
" option.
NOTE
" option.
" option.
CONFIGURATION P1
--- FUEL ---
> LPG
CO SENS.
Includ.
FUEL PARAMETERS
NEXT
ESCAPE
CONFIGURATION P2
ENGINEER UNITS
ALARMS
O2 REFERENCES
Dilution AUTO
NEXT
ESCAPE
TEMPERATURE
Temper.:
40.0 °C
Flow...:
40.0 °C
Return.:
--.- °C
Differ.:
--.- °C
COPY
ESCAPE
SELECT TEST
0.3 hPa
Test Type : A
START TEST
ZERO
ESCAPE
7