RIDGID FG-100 Quick Reference page 7

Flue gas analyser
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
B. External pressure probe
Draft can be measured using the external pressure probe.
Do not exceed 0.2 mbar input.
REMOVE THE PROBE FROM THE STACK
#
Press
DP
Press
to input the ambient temperature
Connect the external probe to the instrument
be sure that the two elements, sensor and probe, are fixed
and in vertical position. Do not move the probe after the zeroing.
Press
F2
to select the external Sensor input
Press
FI
"ZERO" to
Insert the probe inside the standard boiler holes (see figure). If the boiler
do not have them, insert the probe at 2 diameter from the boiler and in
central position in the stack
W
ait for the measure stabilization.
Press
to save the measured value. The v
Press
F2
to exit to the previous menu.
3. Smoke index procedure
The smoke index can be set.
REMOVE THE PROBE FROM THE STACK
Press:
To avoid condensation inside the smoke pump, aspirate ambient air
before the gas sample by the fireside (not in the exhaust gas connection
pipe) without using the paper filter. Introduce now the filter in the pump. The
gas sample should be drawn to the center of the combustion gas flux. The
correct gas volume is sampled with 10 uniform suctions from the pump.
Compare the darkening of the filter with the smoke index table.
The smoke index found cannot exceed the ones defined by the current
normative.
We suggested to use a retaining cone during gas sampling.
Select the Test (from 1 to 3) for input, pressing
Press:
to confirm the selection.
Enter value ( from 0 to 9) pressing
Press
to confirm changes.
Execute the steps above for other test in the same way.
Select the "Oil derivative" field for input, pressing
Press:
to confirm the selection.
Enter value (Yes or Not) pressing
Press
to confirm changes.
Values will printed on the report.
Press
F2
to exit to the previous menu.
4. Connect the probe to the instrument
Flue gas probe
5. insert the probe to the chimney
6
ATTENTION:
reset the display indication
F1=Internal sensor
F2=External sensor
.
alue will printed on the report.
- applicable for oil burner
Meas. 1: -
Meas. 2: -
Meas. 3: -
Meas. 1: 3
Meas. 2: 3
Meas. 3: 2
Ambient temperature probe
Insert the probe inside the standard
boiler holes (see figure).
If the boiler do not have them, insert
the probe at twice the diameter from
the boiler and in central position in the
stack.
TaE
23.00
INT
EXT
DRAUGHT
Measure
1.0 mbar
Press [ENTER]
To save measure
ZERO
ESCAPE
SMOKE
AVG
-
Oil.Der.: NO
ESCAPE
SMOKE
AVG
-
Oil.Der.: NO
ESCAPE