Flue Gas Analyser - RIDGID FG-100 Quick Reference

Flue gas analyser
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
"EN 50379 TUV Approved
by RgG 258"
WARNING! Read these instructions and the accompanying
safety booklet and the complete technical manual on the
CD-ROM carefully before using this equipment.
If you are uncertain about any aspect of using this tool,
contact your RIDGID distributor for more information.
Failure to understand and follow all instructions may result in electric shock,
fire, and/or serious personal injury.
SAVE THESE INSTRUCTIONS!
n
The analyzer should be used in environments where the temperature is
between -5°C and +45°C.
n
When the analysis is complete, before turning the instrument off, remove
the probe and wait about 30 second, to clear the circuit from gas.
n
Do not use the analyzer with clogged or humid filters.
n
Before placing the probe in the case, be sure that the probe, the water
trap and the hose are clean and cold.
n
If you have a double hose sampling probe, insert both connectors to the
instrument during the gas analysis (pay attention to the connector
colour code).
n
To empty the water trap from the condensate water, remove the
draining plug. DO NOT OPEN the water trap. DO NOT leave the water
trap without the draining rubber plug.
n
Never store the analyser together with solvents.
n
Do not open the instrument.
n
Handle the instrument properly and according to its intended purpose.
n
For the best efficiency and accuracy, we suggest to calibrate the
instrument within 12 months after purchase.
If the instrument is stored at temperature exceeding the operative limits,
the analyzer will a few needs some minutes to warm-up to the actual
ambient temperature, before starting the operation.
Regular servicing of the FG-100 is the best way to ensure maximal
machine uptime. It is recommended to have your device serviced and
calibrated on an annual basis by an authorised service station.
Contact your RIDGID representative for contact details.