USER MANUAL
STARRSED RS
Version 6.70 MRN-071_1-EN
Master Registration Number: MRN-071_1-EN

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the STARRSED RS and is the answer not in the manual?

Questions and answers

Summary of Contents for RR Mechatronics STARRSED RS

  • Page 1 USER MANUAL STARRSED RS Version 6.70 MRN-071_1-EN Master Registration Number: MRN-071_1-EN...
  • Page 3 Postal address: 1620 AE Hoorn The Netherlands De Corantijn 13 1689 AN Zwaag Office address: The Netherlands © Copyright R&R Mechatronics Manufacturing B.V. All rights reserved. Subject to changes without prior notice. Issued by the Support Department of RR Mechatronics...
  • Page 5 August 2014 Service Screen Temporary release Schavemaker • 6.66 November History Screen Introduction of chapter Quality 2013 Schavemaker Control Settings Screen • Barcode reader/tube detection Quality Control rewritten Trouble shooting Starrsed RS User Manual Page 3 Version 6.70 MRN-071 (A)-EN...
  • Page 6 Program version 2.22 • 6.30 September Removed H.E. van Dijk Removed the installation section 2008 installation section from the manual and insert this in from manual Installation manual. Page 4 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 7 Correction made in WI 191, 192, Insert section 193, 194 5.6.1.2 Conveyer • motor test Add in settings-carrousel the pipette flow diagram pop-up Insert in the 5.4 the fill and clean screens procedure Starrsed RS User Manual Page 5 Version 6.70 MRN-071 (A)-EN...
  • Page 8 Make corrections and add the 2004 latest pictures • Update the sections according the latest version of the Graphic User Interface • Preliminary November H.E. van Dijk Start of he manual 2003 Page 6 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 9: Table Of Contents

    Medical requirements: ................. 30 5.2.3. General laboratory practices: ..............30 5.2.4. Specific laboratory practices and requirements: .......... 31 5.3. E.C. Declaration Starrsed RS ................32 5.4. Labels and stickers on containers ................33 5.4.1. Stickers for the onboard reagents bottles ............ 33 5.4.2.
  • Page 10 6.5.2.6. Separator sensor ..................74 6.5.2.7. Flow sensor .................... 74 6.5.2.8. Measure sensor ..................75 6.5.3. Display error history ..................76 6.5.4. Display maintenance history ................ 76 6.5.5. Maintenance info ..................77 Page 8 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 11 Advanced ....................110 GETTING STARTED ....................111 7.1. Liquid levels ......................111 7.2. Limit filter settings ....................111 7.3. Reagents preparation ..................111 7.3.1. Starrsed Rinse solution QRR 010934 ............113 Starrsed RS User Manual Page 9 Version 6.70 MRN-071 (A)-EN...
  • Page 12 Turn off sequence ..................164 QUALITY CONTROL ....................165 10.1. Control pipettes ....................165 10.2. Monitoring measurement quality with Starrsed Control ........165 10.2.1. Limitations ....................165 10.2.2. Expected value range ................166 Page 10 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 13 Rinse solution spilling over the instrument ..........186 15.1.3. Rinse pump failure ..................186 15.1.4. Sample probe is not washed after aspiration ..........186 15.1.5. Saline dripping in the sample tube adapter ..........187 Starrsed RS User Manual Page 11 Version 6.70 MRN-071 (A)-EN...
  • Page 14 New rack adapter ..................207 15.21. Rack unit errors .................... 207 15.21.1. Gripper module ..................208 15.21.2. Mixer module ..................209 15.21.3. Check the mixer ..................209 15.22. Communications ................... 210 Page 12 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 15 Communication error (E2) ................211 15.28. Quality control trouble shooting ..............213 APPENDIX FOR STARRSED AUTO-COMPACT ............ 216 WORK INSTRUCTION STARRSED AUTO-COMPACT ........... 268 GLOSSARY OF TERMS ..................321 INDEX........................323 Starrsed RS User Manual Page 13 Version 6.70 MRN-071 (A)-EN...
  • Page 16: Introduction

    To be used for all patient populations, not restricted by age or any other anatomical or physiological particulars. 1.2. Indication for use of device For screening and follow-up of patients with (suspected) inflammatory conditions. Page 14 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 17: Instrument Overview

    EDTA blood sample tubes. The system uses rack adapters and is set up for a specific type of rack. • Windows based System Software is running on an external computer. Starrsed RS User Manual Page 15 Version 6.70 MRN-071 (A)-EN...
  • Page 18: Explanation Of Symbols

    Warning sign to prevent personal injury due to hot surfaces. Warning sign to prevent personal injury due to electrical shocks. General warning. General note. This symbol indicates a reference to this or other product documentation Page 16 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 19: Eqas

    Introduction 1.5. EQAS EQAS is RR Mechatronics' comprehensive tool to periodically assess the performance of the Starrsed instrument at your lab. Through EQAS, the test results of the instrument, obtained with Starrsed Control are objectively compared with a worldwide peer group using the same type of instrument.
  • Page 20: Sedimentation Measurement Principle

    (darkest = 100 % and the lightest = 0 % absorption respectively). By definition the levels are: 87.5% Cells/ plasma separation 75.0% Hazy detection 50.0% Meniscus detection Graphic showing typical absorption values of a sample Page 18 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 21: Instrument Description

    Results of the test are expressed in millimeters. This data together with the patient ID number is sent to the Laboratory Information System along with the sedimentation time used (60 or 30 minutes), the temperature and the dilution ratio. Starrsed RS User Manual Page 19 Version 6.70 MRN-071 (A)-EN...
  • Page 22 This model uses large bulk containers for reagent supply instead of the 0.5-liter onboard containers. This means longer operation time without reagent preparation and no cleaning time. This model is delivered with longer level sensors and special reagents cover assembly. Page 20 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 23: Technical Specifications

    QRR 010931 Starrsed Diluent Reagents used QRR 010947 Starrsed Disinfectant QRR 010933 Starrsed Saline QRR 010934 Starrsed Rinse solution De-ionized water Blood volume: Aspirated blood volume per 1.4 ml in EDTA mode sample Starrsed RS User Manual Page 21 Version 6.70 MRN-071 (A)-EN...
  • Page 24 20 Gb Hard disk on external PC Storage medium approx. 5 Mb per 1000 samples (results and raw data) Storage capacity indication LIS connectivity: Ethernet and serial, bi-directional according various Communication protocols Page 22 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 25: Technical Specifications Contd

    Table space (incl. PC) 770 mm Height Weight empty (ESR 70 kg unit+rack) Table size (Min.) 1100 mm Width (excl. space for PC) 1500 mm Width (incl. space for PC) 660 mm Depth Starrsed RS User Manual Page 23 Version 6.70 MRN-071 (A)-EN...
  • Page 26: Used Reagents

    Starrsed Disinfectant is a dedicated IVD product, developed exclusively for automatic disinfection of the waste system of Starrsed ESR analyzers. Approximately 0.5 ml disinfectant is used after each pipette rinse. This reagent is delivered in a 5-liter container (QRR 010947) Page 24 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 27: Starrsed Cleaning Agent

    Starrsed ST, Starrsed RS, Starrsed RL and Starrsed TL are instrument names of RR Mechatronics. Until 2014 the Starrsed ST was called Inversa 24M, the Starrsed RS was called AutoCompact, the Starrsed RL was called Interrliner (Standalone). RR Mechatronics changed the product names to benefit better from the worldwide recognizability of the brand name Starrsed.
  • Page 28: General Safety Instructions

    Example: • A collision with a moving object or a person. • Something falling on the instrument. • Liquids spilling into the instrument. Page 26 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 29: Starrsed Incident Report

    General safety instructions 3.2. StaRRsed Incident Report Starrsed RS User Manual Page 27 Version 6.70 MRN-071 (A)-EN...
  • Page 30: Installation

    The instrument must be unpacked, installed and checked by a trained engineer prior to first operation. Detailed installation instructions are given in the Starrsed Auto-Compact Installation manual. Connect the instrument only to a properly earthed power connection. Page 28 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 31: Standard Operating Procedures (S.o.p.)

    Transfer materials according to federal and international regulations. • Be familiar with written instructions for laboratory procedures and proper responses to emergencies. • Report spills, exposures, illnesses, and injuries immediately. Starrsed RS User Manual Page 29 Version 6.70 MRN-071 (A)-EN...
  • Page 32: For Working With Bio Hazardous Materials

    Departmental Safety Guidelines and Bio safety Policy. Each user is responsible to leave a clean, disinfected and tidy work place. All biohazard waste must be properly disposed. Page 30 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 33: Specific Laboratory Practices And Requirements

    Dispose of gloves into a biohazard container. The use of eye protection is advised while processing samples. Remove and properly dispose of gloves and wash hands before leaving the laboratory. Starrsed RS User Manual Page 31 Version 6.70 MRN-071 (A)-EN...
  • Page 34: Declaration Starrsed Rs

    Standard Operating Procedures (S.O.P.) 5.3. E.C. Declaration Starrsed RS Page 32 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 35: Labels And Stickers On Containers

    Standard Operating Procedures (S.O.P.) 5.4. Labels and stickers on containers 5.4.1. Stickers for the onboard reagents bottles Starrsed RS User Manual Page 33 Version 6.70 MRN-071 (A)-EN...
  • Page 36: Stickers Of The Reagents Containers

    Standard Operating Procedures (S.O.P.) 5.4.2. Stickers of the reagents containers Page 34 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 37: Starrsed Rs Program

    Starrsed RS program 6. STARRSED RS PROGRAM The Starrsed Auto-Compact is controlled via an external computer on which runs the Starrsed Auto-Compact software. The software functions are grouped on six tabbed screens. The software is controlled by mouse pointer or directly via the touch screen. A virtual keyboard is automatically displayed on screen, when numerical or alphanumerical input is required.
  • Page 38: Software Version

    The following screens are selectable via the associated tabs: 1. S screen (on page 36) AMPLE 2. H screen (on page 42) ISTORY 3. R screen (on page 64) EAGENTS 4. M screen (on page 67) AINTENANCE Page 36 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 39 When the Compact is running in the NORMAL MODE, the instrument uses the standard saved settings. For instance the ESR time is set back to 60 minutes or 30 minutes according the used method. Starrsed RS User Manual Page 37 Version 6.70 MRN-071 (A)-EN...
  • Page 40: Sample Screen Sections Explanation

    This is a graphical representation of the pipette. It is generated from the results of the ESR measurement. It can be used to locate possible air bubbles. Sample mode button: This is the button to start or stop the run mode of the instrument. Page 38 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 41 (EDTA or Citrate), selected method (60 or 30 minute) and symbols that draw attention to certain maintenance conditions or QC sample status (if applicable). Manual input sample ID: Can be used for manual input of a barcode from a sample tube. Starrsed RS User Manual Page 39 Version 6.70 MRN-071 (A)-EN...
  • Page 42: Sample Screen With Keyboard

    A virtual number pad is shown. Type the number of the requested pipette and press the OK button. The pipette information screen is shown. Page 40 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 43: Pipette Information

    The sample has been measured and the pipette is marked to be washed when it reaches the rinse station. This indicator is also on when a sample could not be aspirated properly. Starrsed RS User Manual Page 41 Version 6.70 MRN-071 (A)-EN...
  • Page 44: History Screen

    END ALL TO HOST: Send only the selected patient result to the HOST. END PATIENT RESULT TO • (on page 47) ISPLAY RACK HISTORY (on page 48) ISPLAY RACK DETAILS Page 42 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 45: Display Pipette Data

    After measuring the pipette, the pipette data is transferred to the sample history files. In the header of the table the names of the columns are shown. Double-click the header of any column to sort the table by this column in ascending order. Starrsed RS User Manual Page 43 Version 6.70 MRN-071 (A)-EN...
  • Page 46: Display Sample History

    In the header of the table the names of the columns are shown. Double-click the header of any column to sort the table by this column in ascending order. Page 44 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 47: Display Patient Results

    Starrsed RS program 6.3.2.1. Display patient results Starrsed RS User Manual Page 45 Version 6.70 MRN-071 (A)-EN...
  • Page 48 Pipette in which the sample was measured. Error code Shows any ESR error or warning code (e.g. "Too many borders found"). A complete overview of measurement data is shown with activating the option "Display raw data". Page 46 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 49: Display Rack History

    ESR was measured, but with fill errors. The Clear rack history button will clear the contents of the rack history file and restart to build-up a new rack history file. Starrsed RS User Manual Page 47 Version 6.70 MRN-071 (A)-EN...
  • Page 50: Display Rack Status

    The last 10 racks are stored and can be selected. The Clear rack history button will clear the contents of the rack history file and restart to build-up a new rack history file. Page 48 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 51: Esr Statistics Screens

    ESR errors (%) Increasing ESR errors may indicate the need for maintenance, see ESR Error (on page 156) • Number of samples This can be used to document variations in work load. Starrsed RS User Manual Page 49 Version 6.70 MRN-071 (A)-EN...
  • Page 52: Qc Results Screens

    Linked lab ID: The Lab ID is given if it is linked to the Starrsed Control sample ID Sampling date: The date and time when the QC sample was aspirated. Page 50 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 53 Results can be exported to a .CSV file and imported in an MS Excel file for further analyses. ATCH All used batches of Starrsed Control are shown, results are shown for chosen batch ID. LOSE Return to History screen (on page 42). Starrsed RS User Manual Page 51 Version 6.70 MRN-071 (A)-EN...
  • Page 54: Qc Normal Results Screen Extended

    Starrsed RS program 6.3.5.2. QC normal results screen extended After scrolling the general data from the QC results are shown. 6.3.5.3. QC normal results (graph) Page 52 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 55: Qc Abnormal Results (Table)

    This graph gives a first indication of the measuring stability of the Starrsed Auto-Compact. Further analysis and identification of systematic errors have to be performed in the user's Quality Control System. LOSE Return to History screen (on page 42). 6.3.5.4. QC abnormal results (table) Starrsed RS User Manual Page 53 Version 6.70 MRN-071 (A)-EN...
  • Page 56 After these columns additional data is shown: pipette number, dilution rate, ESR30, ESR time and Aspect. Scroll to the right. Results are always shown with and without Temperature correction, independent of the setting (ON or OFF). ORRECTION The following options can be selected: Page 54 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 57: Qc Abnormal Results Screen Extended

    All used batches of Starrsed Control are shown, results are shown for chosen batch ID. LOSE Return to History screen (on page 42). 6.3.5.5. QC abnormal results screen extended After scrolling the general data from the QC results are shown. Starrsed RS User Manual Page 55 Version 6.70 MRN-071 (A)-EN...
  • Page 58: Qc Abnormal Results (Graph)

    This graph gives a first indication of the measuring stability of the Starrsed Auto-Compact. Further analysis and identification of systematic errors have to be performed in the user's Quality Control System. LOSE Return to History screen (on page 42). Page 56 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 59: Display Sample History (Qc)

    44)" screen. Depending on the frequency of QC samples, related patient results may span over multiple days and are listed per date. All general ESR data and errors of QC samples are shown here. Starrsed RS User Manual Page 57 Version 6.70 MRN-071 (A)-EN...
  • Page 60: Linked Qc Id's

    - QC UTOMATICALLY REMOVE LINKED AFTER RESULT ETTINGS 92)) the links can be removed automatically when a usable ESR result has been ETTINGS ON PAGE reported for this particular Lab ID. Page 58 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 61: Qc Result Analysis

    By selecting Dilution <= 90 all the samples with a dilution rate <= 90 are displayed in the table. In the header of the table the names of the columns are shown. Double-click the header of any column to sort the table by this column in ascending order. Starrsed RS User Manual Page 59 Version 6.70 MRN-071 (A)-EN...
  • Page 62: History Analyse Results High Dilution

    In the header of the table the names of the columns are shown. Double-click the header of any column to sort the table by this column in ascending order. Page 60 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 63: History Analyse Error

    In the header of the table the names of the columns are shown. Double-click the header of any column to sort the table by this column in ascending order. Starrsed RS User Manual Page 61 Version 6.70 MRN-071 (A)-EN...
  • Page 64: History Sample Analyse Option

    3. A range of days from start date to end date. 6.3.12. History sample analyse option day Make a selection for all of today's results or only today's results between start time and end time. Page 62 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 65: Set Start Date

    Starrsed RS program 6.3.13. Set start date Enter the Start date and time. 6.3.14. Set end date Enter the End date and time. Starrsed RS User Manual Page 63 Version 6.70 MRN-071 (A)-EN...
  • Page 66: Reagents Screen

    ! appears. Processing of new samples EAGENTS is stopped. A log is available for all reagents and can be accessed by clicking on D ISPLAY REAGENT (on page 65). HISTORY Page 64 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 67: Display Reagent History

    Select the reagent type on the right side. For external use of the information all the logged reagent data can by exported to EXCEL .CSV format by clicking Export to Excel (CSV). 6.4.2. New reagent input Starrsed RS User Manual Page 65 Version 6.70 MRN-071 (A)-EN...
  • Page 68: New Reagent Input (Cont)

    3. Enter / read the Expiry date (if barcode reader is used: cursor has to be in one of the two boxes) 4. If necessary, adjust the date when the reagent was placed. 5. Check if the preview box shows the correct information, then press OK. Page 66 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 69: Maintenance Screen

    2. C (on page 71) HECK SENSORS 3. D (on page 76) ISPLAY RROR HISTORY 4. D . (on page 76) ISPLAY AINTENANCE HIST 5. M (on page 77) AINTENANCE Starrsed RS User Manual Page 67 Version 6.70 MRN-071 (A)-EN...
  • Page 70: Prime / Clean

    When the Starrsed Auto-Compact has been idle for more than eight hours, there might be some backflow of reagents due to gravity. Prime all tubing before sampling by using the function: • RIME ALL UNITS All priming functions are sequentially performed one time. Page 68 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 71 This function fills all pipettes with a cleaning agent and removes the cleaning agent after a specified time. • End-of-day wash: All pipettes will be washed once and needle, fill-nozzle and rinse-nozzle (wash station) are primed. Starrsed RS User Manual Page 69 Version 6.70 MRN-071 (A)-EN...
  • Page 72: Fill And Clean Screen

    See chapter Maintenance Fill and clean procedure (on page 179) for more information. 6.5.1.2. End-of-day-wash procedure • End-of-day wash: All pipettes will be washed once and needle, fill-nozzle and rinse-nozzle (wash station) are primed. Page 70 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 73: Check Sensors

    After finishing the test, both LED's must be green. • Check Separator sensor: Click the Check button. The green light is shown if the sensor value is in range. Starrsed RS User Manual Page 71 Version 6.70 MRN-071 (A)-EN...
  • Page 74 "ESR (mm)". When the sensor is out of range and the light is red, the sensor values (on page 73) can be checked by turning on the service mode. Page 72 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 75: Check Sensors In Service Mode

    When there is a sensor alarm, an alarm indicator is shown in the tab M AINTENANCE All functions to check the status of the sensors are grouped under button C (on page HECK SENSORS 71). Starrsed RS User Manual Page 73 Version 6.70 MRN-071 (A)-EN...
  • Page 76: Fill Stop Sensor

    Check Flow sensor: The vacuum unit switches on and the values must be in this range: For Starrsed Auto-Compact Flow: 0925-0980-1020 Abs: 0300-360-0390 Note: If for example the yellow orifice is blocked the flow will be: Offset: 0045-0050-0055. Page 74 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 77: Measure Sensor

    Clean sensors first before executing this function. When a test pipette is installed at the measuring position the result of the test pipette is displayed in the field "ESR (mm)". Starrsed RS User Manual Page 75 Version 6.70 MRN-071 (A)-EN...
  • Page 78: Display Error History

    When there is a sensor alarm, an alarm indicator is shown in the tab M AINTENANCE All performed maintenance functions are logged automatically. This log can be saved e.g. to a memory stick by clicking button Save As ... Page 76 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 79: Maintenance Info

    Press the button Info to open the work instruction for a specific maintenance level. When this maintenance is done press the button Done to log the completed work in the maintenance log file. Starrsed RS User Manual Page 77 Version 6.70 MRN-071 (A)-EN...
  • Page 80: Close

    All pipettes will be washed once, needle, fill-nozzle and rinse-nozzle (wash station) are primed. Select the time of the day in hours and minutes for automatic start of this function. Page 78 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 81: End-Of-Day-Wash Options

    Only once: The function runs only once at the selected time. • Weekdays: The function runs only on working days (monday till friday) at the selected time. • Daily: The function runs on a daily base at the selected time. Starrsed RS User Manual Page 79 Version 6.70 MRN-071 (A)-EN...
  • Page 82: General Settings

    If virtual keyboard is switched ON a virtual keyboard pops up for input the data. When Virtual keyboard is switched OFF the arrow keys must be used for input of data. When ready with press S before you continue. AVE SETTINGS Page 80 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 83 ON: The carousel is filled with optimum sample speed. The sample tubes in the rack AST FILLING are faster available for the user but the carousel shows more unused pipettes. Starrsed RS User Manual Page 81 Version 6.70 MRN-071 (A)-EN...
  • Page 84 Select the language for the software and the "Instructions For Use" by clicking on the appropriate language name. Note: To switch to the selected language the first time, close and restart the software. Page 82 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 85: Carousel Control

    Starrsed RS program 6.6.1. Carousel control Starrsed RS User Manual Page 83 Version 6.70 MRN-071 (A)-EN...
  • Page 86 Select the language for the software and the "Instructions For Use" by clicking on the appropriate language name. Note: To switch to the selected language the first time, close and restart the software. Page 84 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 87: Flow Test Potentiometer Mean

    This test is a useful function for checking the pipette position adjustment, vacuum adjustment and filling height sensor positioning adjustment. Each individual pipette is tested and results are displayed in their own line of the diagram. Starrsed RS User Manual Page 85 Version 6.70 MRN-071 (A)-EN...
  • Page 88: Flow Test Absolute

    This test is a useful function for checking the pipette position adjustment, vacuum adjustment and filling height sensor positioning adjustment. Each individual pipette is tested and results are displayed in their own line of the diagram. Page 86 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 89: Flow Test Fill Sensor Air

    This test is a useful function for checking the pipette position adjustment, vacuum adjustment and filling height sensor positioning adjustment. Each individual pipette is tested and results are displayed in their own line of the diagram. Starrsed RS User Manual Page 87 Version 6.70 MRN-071 (A)-EN...
  • Page 90: Flow Test Start

    HANGE INSE POSITION key for learning a new position table. EARN CAROUSEL POSITION Note: This option is used when there is a position error. Page 88 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 91: Diluter Settings

    Example: Dilution error detection is set at 10%. When the dilution error is outside the 10% range, in the last column of the report EDTA 079 or EDTA 121 is printed which indicating this sample is 21% under or over diluted. Starrsed RS User Manual Page 89 Version 6.70 MRN-071 (A)-EN...
  • Page 92: Auto Dilution Adjust

    When the flow sensor is still giving errors after trouble shooting and there are no detectable faults in the liquid flow, use the switch OFF function. This check is now not active, call for service. Page 90 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 93: Limit Error Settings

    YES: always transmit results with bubbles on top warning to the output. set to NO: transmit no results to the output when there is generated a warning for bubbles on top. Starrsed RS User Manual Page 91 Version 6.70 MRN-071 (A)-EN...
  • Page 94: Qc Settings

    Note: If limit error filters in the LIMS are not corresponding with the limit error settings on the instrument it is possible that not all expected results are send to the LIMS. 6.6.4. QC Settings Page 92 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 95 This link will stay active if there is no result or a general ESR error is generated. This setting is useful when the lab issues a new and unique Lab ID for every QC sample. Starrsed RS User Manual Page 93 Version 6.70 MRN-071 (A)-EN...
  • Page 96: Service Screen

    There are no restrictions for the file AVE SETTINGS AS name. • is used for storing software settings after settings are changed or altered. AVE SETTINGS Page 94 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 97: Rack Unit Setting

    This screen gives access to: 1. Mixer motor (on page 96) settings. 2. Conveyer motor (on page 97) settings. 3. Position motor (on page 97). 4. Timing settings (on page 100). Starrsed RS User Manual Page 95 Version 6.70 MRN-071 (A)-EN...
  • Page 98: Mixer Motor

    OLD LEVEL 4. H : default 20% OMING SPEED After setting use Save settings to save the new settings. If Save settings is not pressed the old settings will be used Page 96 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 99: Conveyer Motor

    Starrsed RS program 6.7.1.2. Conveyer motor 6.7.1.3. Position motor Starrsed RS User Manual Page 97 Version 6.70 MRN-071 (A)-EN...
  • Page 100 Rack out position. : Service purpose only. This is to move from the home position to any distance in O TO POSITION millimeters from the home position. Page 98 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 101 IX POSITION 3. Go to T UBE POSITION NUMBER After pressing Read barcode the barcode is shown above the button. If barcode label is not read correctly see section Troubleshooting. Starrsed RS User Manual Page 99 Version 6.70 MRN-071 (A)-EN...
  • Page 102: Timing Settings

    For service purpose only. Those settings are set in the factory, no need for altering the timing settings. The timing settings T8 and T11 are used for the mixer setting. Page 100 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 103: Lims Settings

    4. Party bit: Selectable none, space, mark, even and odd 5. Stop bits: Selectable 1, 1.5 or 2 stop bits 6. Flow control: Selectable for; None XON/XOFF RTS/CTS DTR/DSR XON/XOFF & RTS/CTS XON/XOFF & DTR/DSR Starrsed RS User Manual Page 101 Version 6.70 MRN-071 (A)-EN...
  • Page 104: Set Protocol Settings

    When selected the following keys will pop-up. Checksum On/Off. 30 Minute method On/Off. 16. Advia 120 17. Advia 120 unidirectional 18. InteRRliner For more details see Section 9 Protocols (on page 115) and the Appendix Page 102 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 105: Compact Settings

    OFF, this file cannot be selected in the history window. EARCH IN EXAMPLE HISTORY • If S is ON, this can be selected in the history window. EARCH IN EXAMPLE HISTORY Starrsed RS User Manual Page 103 Version 6.70 MRN-071 (A)-EN...
  • Page 106: Printer Port

    ON, this can be selected in the history window. EARCH IN EXAMPLE HISTORY 6.7.3.7. Outer needle depth Setting for the O . No need to change the default 51.0 mm setting. UTER NEEDLE DEPTH Page 104 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 107: Manual Control

    Note: If vacuum pump is OFF rinse solution will spill over the Auto rack unit OFF: LL PUMPS All active pumps will be switched OFF. The waste pump is switched ON for one minute. Starrsed RS User Manual Page 105 Version 6.70 MRN-071 (A)-EN...
  • Page 108: Needle Control

    : The Rinse nozzle is send to the rinse nozzle home position. INSE NOZZLE DOWN Led status display • Led status check for the fill- and rinse nozzle is displayed. Page 106 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 109: Valve Control

    Actuators 1. R : Rinse solenoid active, rinse valve-block down. INSE ACTUATOR 2. F : Fill solenoid active, fill valve-block down. ILL ACTUATOR Starrsed RS User Manual Page 107 Version 6.70 MRN-071 (A)-EN...
  • Page 110: Display Maintenance History (Service)

    This list contains a log of all the maintenance what is done which can be useful for the field engineer to check the problems with the instrument in the past. The key C will delete all errors from the list. LEAR Page 108 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 111: Display Error History (Service)

    This list contains error numbers what can be useful for the field engineer to check the problems with the instrument in the past. The key C will delete all errors from the list. LEAR Starrsed RS User Manual Page 109 Version 6.70 MRN-071 (A)-EN...
  • Page 112: Advanced

    There are no restrictions for the file AVE SETTINGS AS name. • is used for storing software settings after settings are changed or altered. AVE SETTINGS Page 110 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 113: Getting Started

    2. Fill up the containers with the reagents. 3. Place the reagent containers. 4. Replace screw caps with level sensors. Starrsed RS User Manual Page 111 Version 6.70 MRN-071 (A)-EN...
  • Page 114 4. Install the level sensors and spacers according the following pictures. Make sure to place the appropriate level sensors in the containers by matching the color codes on the tube and on the container: Page 112 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 115: Starrsed Rinse Solution Qrr 010934

    The latest version of the Material Safety Data Sheet (MSDS) of the used reagents can be found on our web site www.rrmechatronics.com. 7.3.4. De-ionised water Add one or two drops of Starrsed Saline to the deionized water to avoid <bottle empty alarm>. Starrsed RS User Manual Page 113 Version 6.70 MRN-071 (A)-EN...
  • Page 116: Starrsed Disinfectant Qrr 010947

    (biohazardous) material and disposed of according to local regulations. Preferably, discard the complete waste container and replace it with a cleaned one. Clear the error. Disclaimer: Check your local environment rules about discharging the waste. Page 114 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 117: Reporting

    17. Advia 120 unidirectional The protocol can be set in tab S - LIMS settings. After selecting a protocol, save the new ERVICE settings by pressing the Save setting key. Starrsed RS User Manual Page 115 Version 6.70 MRN-071 (A)-EN...
  • Page 118: General Error Codes In Protocols

    Limit error Error One of the following limits are out of the setting range: • ESR Time • Column height • Dilution • Bubbles on top • Hazy aspect • Temperature Page 116 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 119: String Format For Starrsed

    30 min 80 data characters cs eot Checksum - ON 30 min Data consists, if 30 min. output is switched to ON Positio Description Format Patient identification text Text PPPPPPPPPP Starrsed RS User Manual Page 117 Version 6.70 MRN-071 (A)-EN...
  • Page 120 Text.: left aligned followed by spaces (ASCII $20). xx.…: number made up of (xx...) digits 0 9 (ASCIII $30 $39) with leading zeros. Leading zeros and non-specified positions are filled with spaces (ASCII $20). Page 118 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 121: String Format For Starrsed (V14)

    ASCII string sum = the ASCII sum of all preceding characters incl. stx, cr, lf and ext modulo 256 (ASCII string sum) = the remainder of the ASCII string sum when divided by 256. Starrsed RS User Manual Page 119...
  • Page 122: In Mm. Half Hour Method

    Text.: left aligned followed by spaces (ASCII $20). xx.…: number made up of (xx...) digits 0 9 (ASCIII $30 $39) with leading zeros. Leading zeros and non-specified positions are filled with spaces (ASCII $20). Page 120 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 123: Sedimatic 100 Serial Output

    ]PPPPPPPPPP[ ]WWWW[ ][ ][ ][ ][ ][ ][ ][ ]AAAAAAAAAAA[cr][lf] - Or -: Result string with Error: pp[ ]PPPPPPPPPP[ ][ ][ ][ ][ ][ ][ ][ ][ ][ ][ ][ ][ ]EEEEEEEEEEE[cr][lf] Starrsed RS User Manual Page 121 Version 6.70 MRN-071 (A)-EN...
  • Page 124 If the Compact is switched to the 30 min method the output string has the same format. The Compact automatic apply the conversion table to the 60 min method. If temperature correction is switched on the ESR value will be the temperature corrected ESR value. Page 122 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 125: Sedimatic 15 Serial Output

    If temperature correction is switched on, the ESR value will be the temperature corrected ESR value. • If the result has an error, the ESR value will be 4 space characters (ASCII $20). Starrsed RS User Manual Page 123 Version 6.70 MRN-071 (A)-EN...
  • Page 126: Vesmatic Serial Output

    (xx...) digits 0 9 (ASCII $30 $39) Leading zeros and non-specified positions are filled with spaces (ASCII $20). The following error codes are defined: No cells / plasma found RROR ESR Probably >140 mm RROR Page 124 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 127 If the Compact is switch to the 30 min method the output string has the same format. The Compact automatics apply the conversion table to the 60-minute method. If temperature correction is switched on the ESR value will be the temperature corrected ESR value. Starrsed RS User Manual Page 125 Version 6.70 MRN-071 (A)-EN...
  • Page 128: Text.: Left Aligned Followed By Spaces (Ascii $20)

    Workorder record from LIMS to analyzer: ESR = Yes (capital Y on position 125) Data "Y" space ESR = No (capital N on position 125) Data space "N" MECHATRONICS result record Page 126 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 129 Error code xxxxx See section ESR error codes Limit error message (results) text See section Limit error message (on page 158) Reserved (spaces) text Checksum [CS] See section Checksum Starrsed RS User Manual Page 127 Version 6.70 MRN-071 (A)-EN...
  • Page 130 Measure error ARNING Bubbles on top ARNING Limit error ,no result transmitted!, see RROR Limit error messages (on page 158) Note: See analyzer manual for more information about limit error settings! Page 128 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 131 Dilution rate (%) xxxxx Column height (mm) xxxxx Error code xxxxx See section ESR error codes Limit error message (results) text See section Limit error message (on page 158) Starrsed RS User Manual Page 129 Version 6.70 MRN-071 (A)-EN...
  • Page 132 (digits 0-9) with leading spaces Non-specified positions are filled with spaces Sample codes Aspect codes Sample type Transmitted code Aspect Transmitted code Patient sample Clear QC normal Hazy<10 QC abnormal Hazy<25 Hazy>25 Page 130 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 133 CS cannot be equal to that of the ETX byte (03h). Therefore, if the calculated CS value is 03h, the transmitted CS byte is set to the substitute value 83h in order to avoid erroneous action by the receiving device. Starrsed RS User Manual Page 131 Version 6.70 MRN-071 (A)-EN...
  • Page 134: Compact Bi-Directional Protocol

    EXOR those 2. Take the result and EXOR this with position 3 ect till the position At position 28: means that this is the Request string Char $02 Char $03 Page 132 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 135 Acknowledge (char $06) if the string is found ok. If either the host computer as well the Starrsed Auto-Compact receives a string it should reply by sending a Non Acknowledge (char $15) if the string is found faulty. Starrsed RS User Manual Page 133 Version 6.70 MRN-071 (A)-EN...
  • Page 136: Protocol Opus Bidirectional

    If either the hos t c omputer as well the Starrs ed Auto-Compact r ecei ves a string it should repl y by s endi ng a Non Ac knowl edg e (c har $15) if the string is found faul ty. At position 28: means that this is the Request string Char $02 Char $03 Page 134 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 137 If either the hos t c omputer as well the Starrs ed Auto-Compact r ecei ves a string it should repl y by s endi ng a Non Ac knowl edg e (c har $15) if the string is found faul ty. At position 28: means that this is the Request string Char $02 Char $03 Starrsed RS User Manual Page 135 Version 6.70 MRN-071 (A)-EN...
  • Page 138 After 3 attempts the Host-computer must stop sending the conformation string and the Starrsed Auto-Compact won’t do an ESR on this sample and goes to the next sample. The Host-computer needs to be ready for the request string of the next sample! Page 136 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 139 If the Starrsed Auto-Compact does not get any response from the host, (timeout 10 sec) it will send the result string again. After 3 attempts it will stop communication and generate an error. Starrsed RS User Manual Page 137 Version 6.70 MRN-071 (A)-EN...
  • Page 140 ASCII string sum = the ASCII sum of all preceding characters incl. stx, cr, lf and ext modulo 256 (ASCII string sum) = the remainder of the ASCII string sum when divided by 256. OR 128 = setting the MSB-bit to 1. Page 138 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 141 Text.: left aligned followed by spaces (ASCII $20). xx.…: number made up of (xx...) digits 0 9 (ASCIII $30 $39) with leading zeros. Leading zeros and non-specified positions are filled with spaces (ASCII $20). Starrsed RS User Manual Page 139 Version 6.70 MRN-071 (A)-EN...
  • Page 142: Protocol Sysmex R-3500

    10100 Dilution rate = 101 12300 30 minute ESR value = 123 Reserved 0000… n.a. n.a. = not applicable R3500 Sample data flag record format (131 bytes) # Of Parameter chars Example Comment Page 140 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 143 0 0 = sample does not exist 1 = sample exists Date ordered "yyyymmdd" Sample ID no. Must be the same as inquiry record Rack no. Tube position no. Inquiry mode Patient ID no. Starrsed RS User Manual Page 141 Version 6.70 MRN-071 (A)-EN...
  • Page 144 The code appears in the "sample data record" at variable 'RBC' Compact "ERROR" code messages (see General Error codes in protocols (on page 116)) This code appears in the "sample data record" at variable 'HGB'. Page 142 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 145: Protocol Sysmex R-3500 Unidirectional

    10100 Dilution rate = 101 12300 30 minute ESR value = 123 Reserved 0000… n.a. n.a. = not applicable R-3500 Sample data flag record format (131 bytes) # Of Parameter chars Example Comment Starrsed RS User Manual Page 143 Version 6.70 MRN-071 (A)-EN...
  • Page 146 The code appears in the "sample data record" at variable 'RBC' Compact "ERROR" code messages (see General Error codes in protocols (on page 116)) This code appears in the "sample data record" at variable 'HGB'. Page 144 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 147: Protocol Sysmex R-3500 Epu

    Reserved 00.. n.a. n.a. = not applicable * Note: Visible in combination with software version 5.07 and up ** Note: With leading spaces, applicable from software version 5.08 and up Starrsed RS User Manual Page 145 Version 6.70 MRN-071 (A)-EN...
  • Page 148 The code appears in the "sample data record" at variable 'RBC' Compact "ERROR" code messages (see General Error codes in protocols (on page 116)) This code appears in the "sample data record" at variable 'HGB'. Page 146 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 149: Protocol Sysmex Se-9000

    0 n.a. Reserve 0 n.a. Reserve 0 n.a. Reserve 0 n.a. Reserve 0 n.a. IP message information 000000 n.a. 123000 Esr value = 123 mm 102000 Hazy code = 12 Starrsed RS User Manual Page 147 Version 6.70 MRN-071 (A)-EN...
  • Page 150 Patient number (same as sample data) Rack no. 1234 Rack number = 1234 Tube position no. 03 Tube position in rack = 3 SE9000 order information record format (171 bytes) Parameter # of Example Comment Page 148 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 151 The code appears in the "sample data record" at variable 'RBC' Compact "ERROR" code messages (see General Error codes in protocols (on page 116)) This code appears in the "sample data record" at variable 'HGB'. Starrsed RS User Manual Page 149 Version 6.70 MRN-071 (A)-EN...
  • Page 152: Protocol Sysmex Se-9000 Unidirectional

    0 n.a. Reserve 0 n.a. Reserve 0 n.a. Reserve 0 n.a. Reserve 0 n.a. IP message information 000000 n.a. 123000 Esr value = 123 mm 102000 Hazy code = 12 Page 150 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 153 The code appears in the "sample data record" at variable 'RBC' Compact "ERROR" code messages (see General Error codes in protocols (on page 116)) This code appears in the "sample data record" at variable 'HGB'. Starrsed RS User Manual Page 151 Version 6.70 MRN-071 (A)-EN...
  • Page 154: Protocol Tdlims

    0 n.a. Reserve 0 n.a. Reserve 0 n.a. Reserve 0 n.a. Reserve 0 n.a. IP message information 000000 n.a. 123000 Esr value = 123 mm 102000 Hazy code = 12 Page 152 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 155 Information status exists Date ordered "yyyymmdd" Sample ID no. Must be the same as inquiry record Rack no. Tube position no. Inquiry mode Patient ID no. Patient name Patient birthday "yyyymmdd" Starrsed RS User Manual Page 153 Version 6.70 MRN-071 (A)-EN...
  • Page 156: Result Printout

    7. Pip.=Sedimentation pipette number (number on the pipette belt). 8. Time=Actual sedimentation time in minutes. 9. T=Temperature (in degrees Centigrade). 10. Error=Error message (if the Analyser detects an error). 11. EDTA mode. Page 154 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 157: Report 30 Minute Mode

    5. Aspect (clear, hazy). 6. Manually entered code number. 7. Pip.=Sedimentation pipette number (number on the pipette belt). 8. Time=Actual sedimentation time in minutes. 9. T=Temperature (in degrees Centigrade). Starrsed RS User Manual Page 155 Version 6.70 MRN-071 (A)-EN...
  • Page 158: Esr Error

    ESR "ERROR" and "WARNING" code messages. This code appears in the "sample data record" at column 10. Error: No ESR result is given Warning: Result should be reviewed before release. The following codes are defined: Page 156 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 159 Limit error Error One of the following limits are out of the setting range: • ESR Time • Column height • Dilution • Bubbles on top • Hazy aspect • Temperature Starrsed RS User Manual Page 157 Version 6.70 MRN-071 (A)-EN...
  • Page 160: Limit Error Settings

    The reporting range in the columns 2, 3 and 4 are in millimeters. The start of the measure range is at the top of the meniscus down to 140 mm. If the detection of cells/plasma is over 140 mm then the report will be >140. Page 158 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 161: Aspect Hazy

    (darkest = 100 % and the lightest = 0 % absorption respectively). By definition the levels are: 87.5% Cells/ plasma separation 75.0% Hazy detection 50.0% Meniscus detection Graphic showing typical absorption values of a sample Starrsed RS User Manual Page 159 Version 6.70 MRN-071 (A)-EN...
  • Page 162: Analyser "Hazy" Code Messages

    Sample is clear. Sample is Hazy < 10 Sample is Hazy < 25 Sample is Hazy > 25 Results with hazy aspect can be suppressed in the menu Limit error settings. Page 160 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 163: Operation

    IONIZED WATER tube connected to the fill nozzle cap. 5. P , activates the disinfectant valve. Disinfectant must flow through the small tube RIME ISINFECTANT connected to the pipette wash station. Starrsed RS User Manual Page 161 Version 6.70 MRN-071 (A)-EN...
  • Page 164: Power Up Sequence

    If the small onboard bottles are used, wash and keep the bottles clean to avoid bacterial growth. The Starrsed Auto-Compact has liquid level sensors. When the level sensor alarm appears, replace reagent as soon as possible. Page 162 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 165: Checks During Operation

    When the Starrsed Auto-Compact has been idle for more than eight hours, there might be some backflow of reagents due to gravity. Prime all tubing before sampling by using the function: • RIME ALL UNITS All priming functions are sequentially performed one time. Starrsed RS User Manual Page 163 Version 6.70 MRN-071 (A)-EN...
  • Page 166: Turn Off

    Pressing OK activates the settings. 9.4.2. Turn off sequence • Close the Starrsed Auto-Compact software. • Switch the PC and monitor OFF. • Switch (optional) printer OFF. • Switch the Compact OFF. Page 164 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 167: Quality Control

    Starrsed Control should not be used past the expiry date. Mechatronics as supplier of the Starrsed Control shall not be liable for any claimed damages arising from other than intended usage. Starrsed RS User Manual Page 165 Version 6.70 MRN-071 (A)-EN...
  • Page 168: Expected Value Range

    The history of QC results is maintained internally. Error messages are generated when the QC results are outside the acceptable range. QC samples can be optionally requested by the LIMS and QC results can be send to the LIMS. Page 166 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 169: Quality Control Procedure

    Do not re-use empty tubes. The software interface is described in the chapter History screen (on page 42). Starrsed Control should be disposed of as medical waste. Starrsed RS User Manual Page 167 Version 6.70 MRN-071 (A)-EN...
  • Page 170: Qc Results

    The uncorrected result is within range, but the corrected result is out of range. See Quality control trouble shooting (on page 213) and QC Results screen (on page 50) for more details. Page 168 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 171: Qc Result Analysis

    QC sample step before releasing patient results. If results are not send to the LIMS QC Results can be exported to MS Excel CSV files for further analysis in lab's own Quality Control data system. Starrsed RS User Manual Page 169 Version 6.70 MRN-071 (A)-EN...
  • Page 172: Waste Disposal

    4. Lift the left cover and place the waste container back into the Compact. Note: If you are re-cycling waste containers, make sure that they are bleached and rinsed thoroughly. Page 170 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 173: Compact System Messages

    The cause of this problem can be foam or the waste pump is not working. The Compact will continue to measure and send the ESR results on time to the printer, but the rinse and fill sequences are stopped until the error is solved. Starrsed RS User Manual Page 171 Version 6.70 MRN-071 (A)-EN...
  • Page 174: Test Messages

    7. Rinse head up error. 8. Measure head not home error. 9. Separator full error The explanation of all these messages can be found in Appendix - System messages Starrsed RS Page 172 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 175: Maintenance

    The purpose of the daily maintenance is to keep the instrument clean and contamination as low as possible. Clean all parts that are exposed to blood, wipe the outer surface and the stainless steel plate below the pipette belt. See WI Daily maintenance (on page 292). Starrsed RS User Manual Page 173 Version 6.70 MRN-071 (A)-EN...
  • Page 176: Check Or Replace Sample Probe Or Outer Needle

    A faulty or broken needle can cause a fill time-out error or a dilution error. If necessary replace the sample probe or outer needle. Remove the top cover from the Rack unit. Page 174 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 177: Weekly

    The purpose of the weekly maintenance is to carry out the daily maintenance and additionally check the optical sensor of the measure head and the vacuum pressure. Detailed instructions of this procedure can be found in the Work Instruction Weekly maintenance. (on page 293) Starrsed RS User Manual Page 175 Version 6.70 MRN-071 (A)-EN...
  • Page 178: Check The Sensors In Service Mode

    Detailed instructions of this procedure can be found in the Work Instruction Cleaning liquid separator (on page 282). Symptoms of a dirty separator: 1. Separator errors. 2. Foam in the separator. 3. Waste pump cannot sufficiently remove the waste from the separator. Page 176 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 179: Level 4 Maintenance

    If the tube is not fitted correctly or is worn the following symptoms can occur. • Liquid flowing back into the container. • First glass tube on the pipette belt is not washed sufficiently. Note: The wider bore tube is for the rinse pump. Starrsed RS User Manual Page 177 Version 6.70 MRN-071 (A)-EN...
  • Page 180: Saline-Pump Tube Replacement

    Detailed instructions of this procedure can be found in WI Cleaning liquid separator (on page 282). As part of the Cleaning liquid separator procedure the bacterial Hepa filter QWLV040002 is replaced with a new one. Exchange bacterial filter QWLV040001 on the waste bottle assembly. Page 178 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 181: Fill-Nozzle O-Ring Replacement

    3. Place the two caps on the adapter and mix well. 4. Put the adapter in the Input-pool. 5. Select M , button F AINTENANCE RIME LEAN AND CLEAN Start Fill and clean procedure: 1. Select button OK. Starrsed RS User Manual Page 179 Version 6.70 MRN-071 (A)-EN...
  • Page 182 'hazy' aspects, check the dispenser system carefully and clean it if there is any doubt. Symptoms for a contaminated instrument: 1. If more than 3… 5 out of 20 measurements are reported HAZY. 2. Poor meniscus may be reported. 3. Incorrect results. Page 180 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 183: Level 3 Maintenance

    1. All tubing ESRI079200 and additional tube set. 2. Waste pump motor ESRI090920. 3. Waste pump cassette ESRI090921. 4. Blue Vacuum filter disc. Part no QWLV040003. 5. Fill block washer. Part no ESRI030906. Starrsed RS User Manual Page 181 Version 6.70 MRN-071 (A)-EN...
  • Page 184: Reagents Replacement

    1. Remove level sensors and spacers from the empty container. 2. Remove empty container. 3. Place new container. 4. Remove the container screw caps and pull the necks of the bottle packs out of the cardboard box. Page 182 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 185 After each reagents change, the fluid system must be primed: 1. Select M -> P (on page 68). AINTENANCE RIME LEAN 2. Perform the applicable prime step to fill the relevant tubes with reagent and remove air. Starrsed RS User Manual Page 183 Version 6.70 MRN-071 (A)-EN...
  • Page 186: Data Safety Management

    Year\Month\Day, for example D:\2014\01\03-01-2014. Each result is stored with the sample ID as document name. QC results are stored on D:\QC\NORMAL\batchXXX and D:\QC\ABNORMAL\batchYYY. Page 184 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 187: Trouble Shooting

    See WI-162 rinse tube replacement (on page 269). 2. Check rinse solution level in rinse solution container. If the level is insufficient, a message is displayed and the alarm sounds! Starrsed RS User Manual Page 185 Version 6.70 MRN-071 (A)-EN...
  • Page 188: Rinse Solution Spilling Over The Instrument

    It may be worn or leak. (See Level 4 maintenance). An incorrectly fitted saline peristaltic pump tube may cause the same problem. One of the tubes carrying saline maybe blocked or bended. Page 186 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 189: Saline Dripping In The Sample Tube Adapter

    1. Liquid in the container is not detected. This occurs sometimes with the D bottle IONIZED WATER and is caused by a very low conductivity. 2. Add one or two drops of S to the D to increase the conductivity. ALINE IONIZED WATER Starrsed RS User Manual Page 187 Version 6.70 MRN-071 (A)-EN...
  • Page 190: Compact Stalls

    Trouble shooting 15.3. Compact stalls When the Compact is not working again check the main fuse. The main fuse can be found at the main input socket from the Compact. Page 188 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 191: Diluter

    3. Check all tubes are still connected. 4. Check if the 2 tubes (ESRI070508) are still connected with the tubes ESRI070406. 5. Check if the 2 tubes positioned correctly in the pinch valve. Starrsed RS User Manual Page 189 Version 6.70 MRN-071 (A)-EN...
  • Page 192: Diluter Errors

    1. Check the diluter syringe tip. 2. Check on the Mixing-piece joints and connectors for leaks and replace if necessary. 3. Check the connectors on the EDTA flow sensor for leaks and replace if necessary. Page 190 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 193 Diluent to the tubing and fill system manually. 1. Check pick-up tube at the Diluent container, it may be kinked. 2. One of the Diluent lines has become blocked or kinked. Starrsed RS User Manual Page 191 Version 6.70 MRN-071 (A)-EN...
  • Page 194: Vacuum

    WI-179 Replace blue air filter (on page 279)) 6. Defective flow sensor board. Needs replacement, fatal error. 7. Outer needle valve is not functioning correctly and vacuum is leaking away, check outer needle valve. Page 192 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 195: Vacuum Error

    Fatal error, technical assistance is needed. 4. Bad vacuum. Adjustment needed on the vacuum. Fatal error, technical assistance is needed. 5. Blockage in the 3 way vacuum manifold. Fatal error, technical assistance is needed. Starrsed RS User Manual Page 193 Version 6.70 MRN-071 (A)-EN...
  • Page 196: Flushing Liquids

    After each wash cycle, approximately 0.5 ml of disinfectant will be flushed through the wash station. If no disinfectant flows; 1. Unscrew the cap of the disinfectant container to check. 2. One of the disinfectant lines may be blocked or kinked. Page 194 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 197: Needle System

    If the sample probe has being set too deep, it will touch the bottom of the sample tube. The sample probe then pushes the sample tube slightly downwards, and the aspiration cycle will be aborted. A broken outer needle may cause a similar fault. Starrsed RS User Manual Page 195 Version 6.70 MRN-071 (A)-EN...
  • Page 198: Air Bubbles

    O-ring. 3. Check for air in diluter system. 4. Check that the sample probe O-ring is not leaking. 5. Check transparent mixing- piece block for cracks. Page 196 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 199: Pipette Looks Like Zebra Crossing

    1. If just one air bubble is found about 5mm below the meniscus, the filling speed may be too high. 2. The blood column should not exceed the filling height sensor by more than 10mm. One air bubble can result in ESR Error 3. Starrsed RS User Manual Page 197 Version 6.70 MRN-071 (A)-EN...
  • Page 200: One Air Bubble Rising In Pipette

    Check the fill nozzle for damage. If necessary, replace the fill nozzle. Sample tube is leaking on the fill nozzle side. Replace the silicon sample tube Small air bubbles result in ESR Error 3. Page 198 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 201: Random Air Bubbles In Pipette

    3. A scratched valve body. 4. Valve on top of the pipette is dirty or damaged. 5. Check pipette bottom, glass may be chipped. 6. Check the pipette valve for contamination or wear. Starrsed RS User Manual Page 199 Version 6.70 MRN-071 (A)-EN...
  • Page 202: Rinse Nozzle (Wash Station) Alignment

    First run an extra Fill & Clean sequence, then check after a day's run if haziness is decreased. When there are still many reports, it is recommended to fill the diluter system with a 5% chlorine solution. See WI-178 Hazy problem (on page 278). Page 200 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 203: Hazy Reports After Cleaning

    <default 90 sec> ETTINGS ENERAL SETTINGS IXING INTERVAL 2. Filling procedure stopped by operator. 3. Insufficient sample volume. Should be at least 1.4 ml. 4. Faulty fill nozzle or fill nozzle washer/O-ring. Starrsed RS User Manual Page 201 Version 6.70 MRN-071 (A)-EN...
  • Page 204: Position Error

    INSE POSITION Enter the right value of the pipette number at the rinse station. Close pop-up menu. 2. Select S ET RINSE POSITION HANGE INSE POSITION Press L EARN CAROUSEL POSITION Page 202 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 205 3. Check mechanical connection potentiometers. If the position error is not solved after all the mentioned actions, the positioning device needs replacement. This is a fatal error, call distributor. Starrsed RS User Manual Page 203 Version 6.70 MRN-071 (A)-EN...
  • Page 206: Separator Error

    Waste pump failure. Exchange the waste pump cassette. If the error returns, call for service. Electrical bridge between the waste-level Clean liquid separator, see Cleaning liquid electrodes. separator (on page 282) Page 204 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 207: Reagents

    The fill nozzle did not reach the fill position in a certain time limit. Possible reason; • Fill nozzle motor is faulty. • Fill nozzle motor driver is faulty. • Fill nozzle is blocked. • Fatal error, call distributor. Starrsed RS User Manual Page 205 Version 6.70 MRN-071 (A)-EN...
  • Page 208: Fill Nozzle Not At Home Position

    Piercing position sensor was not triggered, check the piercing position sensor Tube 1 not aligned properly, check the tube 1 position setting and save the setting For other error numbers and details See the Appendix - Error list Starrsed RS (on page 236) Page 206 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 209: New Rack Adapter

    15.21. Rack unit errors The rack unit can give errors on the • Gripper module (on page 208) • Mixer module (on page 209) • Mixer module (on page 209) Starrsed RS User Manual Page 207 Version 6.70 MRN-071 (A)-EN...
  • Page 210: Gripper Module

    IN position. Could not send rack to the grabber. Causes and solutions • Check the position motor For more error numbers and details see the Appendix - Error list Starrsed RS (on page 236) Page 208 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 211: Mixer Module

    Mixer motor send to home position , but position grabber motor is not on the MIX position. • Check the position motor. For error numbers and details see the Appendix - Error list Starrsed RS (on page 236) 15.21.3. Check the mixer The mixer is integrated with the rack grabber and will start mixing as soon the first barcode label is read.
  • Page 212: Communications

    TUBE# "2", the sensor indicator must not light up, “No tube !” is shown next POSITION to the grey block. 8. Repeat this for all applicable tube positions. If tube detection errors are persisting, call for service. Page 210 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 213: Tube Not In 1 Position

    The other codes (command: XX, TSWR: XX and E:XX) are indicators for status/actions from software and I2C connections. Each board has its own range of commands and can not be explained in one way. Starrsed RS User Manual Page 211 Version 6.70 MRN-071 (A)-EN...
  • Page 214 Power cable not connected on the communication PCB mounted on the back panel. • An I2C cable not connected • Serial cable not connected • No power on one of the PCB's • Short circuit or fault on one of the PCB's Page 212 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 215: Quality Control Trouble Shooting

    Check general ESR data, QC result with ESR error no ESR Result is given see ESR Error (on page 156) • Check sample tube volume • Try new QC sample tube Starrsed RS User Manual Page 213 Version 6.70 MRN-071 (A)-EN...
  • Page 216 QC, press "No" to return and take appropriate action. General errors Extra information Action • Check barcode Barcode is not accepted Barcode cannot be read Data is incorrect Page 214 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 217 Note: QC Error messages are only shown and stored in QC results and not send to LIMS. QC result is given with the same general errors and warnings as a normal patient ESR-result Starrsed RS User Manual Page 215 Version 6.70 MRN-071 (A)-EN...
  • Page 218: Appendix For Starrsed Auto-Compact

    Appendix for Starrsed Auto-Compact 16. APPENDIX FOR STARRSED AUTO-COMPACT Appendix section Page 216 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 219 Appendix for Starrsed Auto-Compact Appendix - Article reference list Starrsed RS The Starrsed RS is delivered with a complete accessories kit ESRI 110992. This reference list is for article order numbers only. Part number Description ESRI010246 Pinch valve tube ESRI030903...
  • Page 220 Spacer for cubitainer When using Plastic container for Deionized water storage: ESRI110106 Deionized water sticker 5l QWFG010902 Container 5L for Deionized water QTLV020004 Cap bottle 5l ESRI110949 Level sensor 5l assembly Page 218 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 221 Appendix for Starrsed Auto-Compact Appendix - Cable connection diagram Starrsed RS User Manual Page 219 Version 6.70 MRN-071 (A)-EN...
  • Page 222 Appendix for Starrsed Auto-Compact Appendix - Tube connection Starrsed Auto-Compact Page 220 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 223 In this sample, the dilution rate has a dilution failure of 21% and that is printed as EDTA 079. 905007 Sample results with a text error. This sample gives Too many borders found. Result of a pipette possibly filled with air bubbles. Starrsed RS User Manual Page 221 Version 6.70 MRN-071 (A)-EN...
  • Page 224 Limit error Error One of the following limits are out of the setting range: • ESR Time • Column height • Dilution • Bubbles on top • Hazy aspect • Temperature Page 222 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 225 ESR "ERROR" and "WARNING" code messages. This code appears in the "sample data record" at column 10. Error: No ESR result is given Warning: Result should be reviewed before release. The following codes are defined: Starrsed RS User Manual Page 223 Version 6.70 MRN-071 (A)-EN...
  • Page 226 Limit error Error One of the following limits are out of the setting range: • ESR Time • Column height • Dilution • Bubbles on top • Hazy aspect • Temperature Page 224 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 227 Diluter settings [S ] > D ETTINGS ILUTER SETTINGS Dilution adjust Automatic Automatic Dilution error detect Auto dilution adjust Dilution flow check General settings [S ] > L ETTINGS IMIT ERROR SETTINGS Starrsed RS User Manual Page 225 Version 6.70 MRN-071 (A)-EN...
  • Page 228 Position motor depending on Rack selection customer Set tube 1 position 22 mm Actual position 0.0 mm Always read barcode Set serial output comport [S ] > ERVICE SERIAL OUTPUT SETTINGS Page 226 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 229 IMIT ERROR SETTINGS Starrsed Auto-Compact settings Starrsed Auto-Compact Stand alone Via LIMS Client Settings is operational as Send results when time [ No ] 2 Minutes [ Yes ] exceeded Starrsed RS User Manual Page 227 Version 6.70 MRN-071 (A)-EN...
  • Page 230 Note: When an option is set to YES and this limit error occurs, results will be printed/send to the LIMS. Note: When an option is set to NO and this limit error occurs, results are not printed/send to the LIMS. Page 228 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 231 The cause of this problem can be foam or the waste pump is not working. The Compact will continue to measure and send the ESR results on time to the printer, but the rinse and fill sequences are stopped until the error is solved. Starrsed RS User Manual Page 229 Version 6.70 MRN-071 (A)-EN...
  • Page 232 Compact was not able to position the measure-unit within a certain time limit. Check for mechanical obstructions. 3. Rinse-unit. Compact was not able to position the rinse-unit within a certain time limit. Check for mechanical obstructions. 4. Fill-nozzle unit. Page 230 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 233 Check in screen M if vacuum is available. AINTENANCE HECK SENSORS Check in screen M the value of flow sensor. AINTENANCE HECK SENSORS Fatal error, call distributor. 5. Position error. Starrsed RS User Manual Page 231 Version 6.70 MRN-071 (A)-EN...
  • Page 234 9. Separator full error It takes too long for the waste pump to empty the liquid separator. Check separator assembly on air leaks. Replace waste tubes. Exchange waste pump cassette. Clean liquid separator. Page 232 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 235 To set-up the barcode readers to client specifications, use the set-up tools for the barcode readers. Detailed information can be found in Keyence Configuration manual MRN-011 or in Opticon Configuration manual MRN-015. Starrsed RS User Manual Page 233 Version 6.70 MRN-071 (A)-EN...
  • Page 236 Power cable comes from the power supply. • When the printer is used connect the printer cable onto the 25 pin connector. • The host serial cable is connected to the connecter with the arrow. Page 234 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 237 7. Com1: Compact serial cable comes from the Interface RS232 to IIC. 8. USB connection not used 9. USB connection for touch screen. 10. USB connection not used. 11. Ethernet connection. Starrsed RS User Manual Page 235 Version 6.70 MRN-071 (A)-EN...
  • Page 238 Program stopped and shut down. Check settings changed settings which were computer needed to be before continuing! not saved to disk are lost. reset. • Computer reset after power failure. Page 236 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 239 (home) home (top) sensor within a • Outer needle motor is certain time limit. faulty. • Outer needle motor driver on needle board is faulty. • Outer needle is blocked. Starrsed RS User Manual Page 237 Version 6.70 MRN-071 (A)-EN...
  • Page 240 E26: "Diluent Flow" sensor out The EDTA Flow sensor has Check and/or clean the of range. Check/clean this reached a critical level. EDTA Flow sensor. sensor ! Continuing could result in Page 238 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 241 Position settings in Eeprom do Configuration file maybe Settings loaded from Eeprom. not match settings saved to corrupted. Check settings before file. Settings in Eeprom OK sampling! and loaded from Eeprom. Starrsed RS User Manual Page 239 Version 6.70 MRN-071 (A)-EN...
  • Page 242 T8 or T11 are too low distance). T11. or mixer (gripper) maybe • Mixer is blocked. blocked. • Mixer motor is faulty. • Mixer motor driver on rack board is faulty. Page 240 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 243 Mixer motor is faulty • Mixer motor is blocked. • Mixer motor driver on rack board is faulty. • Check the mixer motor drives the gripper.(Loose belt, loose pulleys ) Starrsed RS User Manual Page 241 Version 6.70 MRN-071 (A)-EN...
  • Page 244 Mixer motor driver on rack board is faulty. • Check mixer motor belt. E69: Position motor error! Not Gripper did not reach the MIX See E61 on position! (MIX) position. Starting point: IN position. Page 242 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 245 See E61 on position! (IN) (catch) position. Starting point: MIX position. E81: Position motor error! Not Gripper did not reach the MIX See E61 on position! (MIX) position. Starting point: IN Starrsed RS User Manual Page 243 Version 6.70 MRN-071 (A)-EN...
  • Page 246 Please exit and then restart succeed due to a possible Starrsed RS. If this error fatal hardware error. occurs repeatedly, please note the error code before seeking assistance!
  • Page 247 Not on position! (FILL) FILL position. Starting point: last sampled tube. E102: Position motor error! Gripper did not reached the See E61 Not on position! (MIX) MIX position. Starting point: FILL position. Starrsed RS User Manual Page 245 Version 6.70 MRN-071 (A)-EN...
  • Page 248 Faulty Sample probe motor. • Check if needles are blocked. • Faulty motor drivers on the needle board. E116-118 Quality Control Errors See section Quality control trouble shooting (on page 213) Page 246 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 249 Perform End-of-day wash Clean outside aspiration needle Check tubing/diluent syringe Clean outside instrument Clean Fill nozzle Clean Liquid separator Check sensors Replace Fill nozzle O-ring QWLV050004 Replace bacterial air filter (HEPA) QWLV040002 Starrsed RS User Manual Page 247 Version 6.70 MRN-071 (A)-EN...
  • Page 250 Check Rack transport unit For older models Flat washer Fill nozzle EDTA QWLV060001 Teflon top repair set (Version I diluter 2 rings) ESRI060911 One-way check valve (Version I diluter 2 rings) QWLV010104 Page 248 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 251 Appendix for Starrsed Auto-Compact Appendix - Rack adapters Starrsed RS The Starrsed Auto-Compact uses dedicated Mechatronics rack adapter sets to accommodate one of the following rack types. Use rack adapters with clips when racks are used in which sample tubes are not fixed.
  • Page 252 Appendix for Starrsed Auto-Compact For manual patient id handling: Beckman Coulter Page 250 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 253 Appendix for Starrsed Auto-Compact Abbott Bayer Starrsed RS User Manual Page 251 Version 6.70 MRN-071 (A)-EN...
  • Page 254 Appendix for Starrsed Auto-Compact Horiba Horiba ABX Page 252 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 255 ABX closed, Pentra 120 VERA149010 Coulter HmX VERA149011 Horiba VERA149021 Beckman Coulter DXH-800 VERA149022 Roche Rack adapters for other rack types can be developed on request. Please contact support@rrmechatronics.com for details. Starrsed RS User Manual Page 253 Version 6.70 MRN-071 (A)-EN...
  • Page 256 The right row (JP2) is for the communication port settings and is set on default 4800 baud. To change the port setting, replace the jumper to the required setting you want. Page 254 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 257 Input Transmit Data (a.k.a TxD, Tx). Sending data from DTE. (RXD) Output Request To Send (ASYNC) (RTS) Input Clear To Send (ASYNC) (CTS) Input Data Set Ready (ASYNC) Signal Ground (SG) Input Carrier Detect (a.k.a DCD). Starrsed RS User Manual Page 255 Version 6.70 MRN-071 (A)-EN...
  • Page 258 Interface connections to PC Example connection to P.C. Example to HOST computer Example connection to P.C. without handshake system including handshake including hardware handshake Hardware no handshaking for DB 25 connector Page 256 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 259 6 + 8 + 20 DSR + CD +DTR 6 + 8 + 20 DSR + CD + DTR 4 +5 CS (CTS) + RS (RTS) RS (RTS) + CS (CTS) Starrsed RS User Manual Page 257 Version 6.70 MRN-071 (A)-EN...
  • Page 260 SD (TXD) -> RD (RXD) GND (SG) 6 + 8 DSR + CD 6 + 8 DSR + CD CS (CTS) ->RS (RTS) RS (RTS) -> CS (CTS) Hardware handshaking for DB 25 connectors Page 258 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 261 500 milli-seconds. If the data transmission is interrupted for more than 500 milliseconds by the host computer then the instrument will stop sending data. This data is then lost to the host computer. Starrsed RS User Manual Page 259 Version 6.70 MRN-071 (A)-EN...
  • Page 262 To overcome this problem and still be able to use a cheap null modem communication cable with only three lines in it, a fake null modem cable layout has been defined. The null modem cable with loop back handshaking resulted from this. Page 260 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 263 All specified RS232 signals are completely isolated from the 'computer earth' and the 'mains supply'. Handshake: Hardware handshake is not applied. Change the data transmission parameters: From the transmission parameters, only the baud rate can be selected from 1200 up to 9600. Starrsed RS User Manual Page 261 Version 6.70 MRN-071 (A)-EN...
  • Page 264 To overcome this problem and still be able to use a cheap null modem communication cable with only three lines in it, a fake null modem cable layout has been defined. The null modem cable with loop back handshaking resulted from this. Page 262 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 265 All specified RS232 signals are completely isolated from the 'computer earth' and the 'mains supply'. Handshake: Hardware handshake is not applied. Change the data transmission parameters: From the transmission parameters, only the baud rate can be selected from 1200 up to 9600. Starrsed RS User Manual Page 263 Version 6.70 MRN-071 (A)-EN...
  • Page 266 Input Clear To Send. Raised by DCE in response to RTS from DTE Input Ring Indicator. Set when incoming ring detected - used for auto- answer application. DTE raised DTR to answer. Female side Male side Pin no Pin no Page 264 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 267 In case of preparation of a cable the following connections must be made. Female side Male side Pin no Pin no If you have no result you can try changing the following pin wires. Female side Male side Pin no Pin no Starrsed RS User Manual Page 265 Version 6.70 MRN-071 (A)-EN...
  • Page 268 Not in use In-0 Motor sensor Waste ERROR Measure. motor tacho In-2 Check position. 51-111 Rinse head top Measure. head top In-3 Check 10 pos Rinse head bottom Measure. head bottom Page 266 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 269 000C - FFFF Board address Selector Picture 2 Current Jumper Picture 1 Forward reverse Selector Fusses In mA In mA In mA Fuse A 1600 Fuse B 1600 1600 Fuse C Starrsed RS User Manual Page 267 Version 6.70 MRN-071 (A)-EN...
  • Page 270: Work Instruction Starrsed Auto-Compact

    Work instruction Starrsed Auto-Compact 17. WORK INSTRUCTION STARRSED AUTO-COMPACT Work instruction section Page 268 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 271 6. Place one end of the tube in the pump plate holder. 7. Pull the new tube over the peristaltic pump rotor. 8. Pull pump tube slightly downwards and at the same time towards the back of the Starrsed Auto-Compact. Starrsed RS User Manual Page 269 Version 6.70 MRN-071 (A)-EN...
  • Page 272 6. Place one end of the tube in the pump plate holder. 7. Pull the new tube over the peristaltic pump rotor. 8. Pull pump tube slightly downwards and at the same time towards the back of the Starrsed Auto-Compact. Starrsed RS User Manual Page 270 Version 6.70 MRN-071 (A)-EN...
  • Page 273 11. Release the stainless steel vacuum tube. 12. Replace the waste container. 13. Close left cover. Prime Saline 1. Select P . Repeat the prime saline until the liquid separator is empty. RIME ALINE Starrsed RS User Manual Page 271 Version 6.70 MRN-071 (A)-EN...
  • Page 274 6. Insert new waste pump cassette ESRI090921 until it clicks into place. 7. Remove the protection caps on from the tubes. 8. Connect the two tubes to new waste pump cassette. Page 272 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 275 6. Check for leakage with Fill & Clean. Replace the top cover: 1. Put the cover carefully over the instrument. 2. Fasten the two rear screws of the top cover. (If present/if needed). Starrsed RS User Manual Page 273 Version 6.70 MRN-071 (A)-EN...
  • Page 276 MS 40..50..60 by using the C HECK function. MEASURE SENSOR 3. If not in range repeat cleaning the inner part of the measuring sensor. 4. When in range switch OFF the Starrsed Auto-Compact. Starrsed RS User Manual Page 274 Version 6.70 MRN-071 (A)-EN...
  • Page 277 Replace the top cover: 1. Put the cover carefully over the instrument. 2. Fasten the two rear screws of the top cover. (If present/if needed). Switch ON the Starrsed Auto-Compact. Starrsed RS User Manual Page 275 Version 6.70 MRN-071 (A)-EN...
  • Page 278 Remove the valve body and check the silicon tube position, it must be exactly centred. 8. Fit the bottom tube clamp and V-seal ring. 9. Check the position of the valve. If incorrect, disassemble pipette valve and tube and re-assemble again. Starrsed RS User Manual Page 276 Version 6.70 MRN-071 (A)-EN...
  • Page 279 6. Check for leakage with Fill & Clean. Replace the top cover: 1. Put the cover carefully over the instrument. 2. Fasten the two rear screws of the top cover. (If present/if needed). Starrsed RS User Manual Page 277 Version 6.70 MRN-071 (A)-EN...
  • Page 280 1. Prepare a Fill and Clean arrangement. 2. Run the fill and clean sequence. When all the pipettes are filled the needle goes back to the home position. 3. Remove the Fill and clean arrangement. Starrsed RS User Manual Page 278 Version 6.70 MRN-071 (A)-EN...
  • Page 281 3. Reconnect the tube connectors on the filter Replace the top cover: 1. Put the cover carefully over the instrument. 2. Fasten the two rear screws of the top cover. (If present/if needed). Starrsed RS User Manual Page 279 Version 6.70 MRN-071 (A)-EN...
  • Page 282 3. Calculate the mean of the 10 samples. 4. Make dilution adjustment if necessary in S > D ETTINGS ILUTER SETTINGS 5. Go to tab S and select D > G ETTINGS ENERAL SETTINGS ISPLAY DILUTION Starrsed RS User Manual Page 280 Version 6.70 MRN-071 (A)-EN...
  • Page 283 3. Refill the diluent bottle with new diluent solution. 4. Perform another [P RIME 5. After the prime sequence stops press [P ] again to fill the dispenser system with the new RIME diluent solution. Starrsed RS User Manual Page 281 Version 6.70 MRN-071 (A)-EN...
  • Page 284 C. A little silicon grease on the rim of separator will make the assembling and adjustment easier. D. If applicable replace the bacterial HEPA filter (For Maintenance Level 4: Exchange filter QWLV040002) Starrsed RS User Manual Page 282 Version 6.70 MRN-071 (A)-EN...
  • Page 285 The following items need to be replaced annually: 1. Outer needle and sample probe (if needed) 2. Pipette valve bodies (84 pieces) QTST040001. Check: 1. Adjustment of fill nozzle and rinse nozzle. Starrsed RS User Manual Page 283 Version 6.70 MRN-071 (A)-EN...
  • Page 286 Safety: Bio Hazard area Instrument: Compact Revision: 001, February 2005 Instructions for disassembling and assembling the fill nozzle Note: For the O-ring replacing, only take the top part away from the fill nozzle. Starrsed RS User Manual Page 284 Version 6.70 MRN-071 (A)-EN...
  • Page 287 The use of a toothbrush and detergent is recommended. 1. Carefully scrub the fill nozzle inner part. 2. Use a tissue to dry the fill nozzle. Starrsed RS User Manual Page 285 Version 6.70 MRN-071 (A)-EN...
  • Page 288 1. Carefully scrub the fill nozzle inner part. 2. Use a tissue to dry the fill nozzle. Disassemble fill nozzle holder: 1. Turn the holder to the right. 2. The holder can now be removed Starrsed RS User Manual Page 286 Version 6.70 MRN-071 (A)-EN...
  • Page 289 1. Connect the silicon tube to the fill nozzle. 2. Put the fill nozzle into the holder. 3. Push the fill nozzle upwards and turn the holder to the left. Starrsed RS User Manual Page 287 Version 6.70 MRN-071 (A)-EN...
  • Page 290 Work instruction Number 205 Page 1 of 1 Purpose: Replace the pinch valve tube ESRI010246 Safety: Bio Hazard area Instrument: Compact Revision: 001, February 2005 Replace the pinch valve tube ESRI010246 Starrsed RS User Manual Page 288 Version 6.70 MRN-071 (A)-EN...
  • Page 291 12. Check sensors and vacuum pressure settings 13. Check instrument settings 14. Check communication interface settings 15. Perform aspiration with test blood before normal operation, check if all steps are performed without errors Starrsed RS User Manual Page 289 Version 6.70 MRN-071 (A)-EN...
  • Page 292 1. Remove inner- and outer needle assembly and dispose of as biohazardous waste 2. Remove filter from separator and (if applicable) waste container, dispose of as biohazardous waste 3. Discard instrument according local regulations Starrsed RS User Manual Page 290 Version 6.70 MRN-071 (A)-EN...
  • Page 293 2. Run 10 samples through the instrument and make a note of the dilution rate. 3. Calculate the mean value of the 10 samples. 4. Make adjustment if necessary in [S ] -> [D ] -> [D ETTINGS ILUTER SETTINGS ILUTION ADJUSTMENT Starrsed RS User Manual Page 291 Version 6.70 MRN-071 (A)-EN...
  • Page 294 ], click button [P 68)] AINTENANCE RIME LEAN ON PAGE and perform the [P ] function. RIME ILUENT ILUTER 7. Wipe outer surface and stainless steel plate below the pipettes with disinfectant. Starrsed RS User Manual Page 292 Version 6.70 MRN-071 (A)-EN...
  • Page 295 2. The fill-nozzle can now be removed. removed. 3. Disconnect the silicon tube 3. Disconnect the silicon tube from the fill nozzle. from the fill nozzle. 2. Clean Liquid separator Removing Starrsed RS User Manual Page 293 Version 6.70 MRN-071 (A)-EN...
  • Page 296 C. A little silicon grease on the rim of separator will make the assembling and adjustment easier. D. If applicable replace the bacterial HEPA filter (For Maintenance Level 4: Exchange filter QWLV040002) Page 294 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 297 ], click button [P 68)] AINTENANCE RIME LEAN ON PAGE and perform the [P ] function. RIME ILUENT ILUTER 7. Wipe outer surface and stainless steel plate below the pipettes with disinfectant. Starrsed RS User Manual Page 295 Version 6.70 MRN-071 (A)-EN...
  • Page 298 Press test. When test is finished, signal Down and signal Up must be green. Separator check • Go to tab M -> C . Select C box. AINTENANCE HECK SENSOR HECK EPARATOR SENSOR Separator sensor <200 600 >700 Starrsed RS User Manual Page 296 Version 6.70 MRN-071 (A)-EN...
  • Page 299 8. Replace the correct tubes on the outer needle. 9. Fasten the outer needle bolt.(Do not over-tighten the screw) 10. Replace the cover on the Rack unit. Starrsed RS User Manual Page 297 Version 6.70 MRN-071 (A)-EN...
  • Page 300 4. When all the pipettes are filled, the needle goes back to the home position and the adapter is moved to the release position and will be transported into the End-pool. Starrsed RS User Manual Page 298 Version 6.70 MRN-071 (A)-EN...
  • Page 301 The use of a toothbrush and detergent is recommended. 1. Carefully scrub the fill nozzle inner part. 2. Use a tissue to dry the fill nozzle. Replace O-ring: Remove the O-ring. (QWLV050004) Install new O-ring. (QWLV050004) Starrsed RS User Manual Page 299 Version 6.70 MRN-071 (A)-EN...
  • Page 302 2. Clean Liquid Separator and exchange filters Removing Cleaning A. Clean all parts with hot water and a brush. B. Use some acid free vaseline on the screw-thread of the glass jar. Page 300 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 303 6. Place one end of the tube in the pump plate holder. 7. Pull the new tube over the peristaltic pump rotor. 8. Pull pump tube slightly downwards and at the same time towards the back of the Starrsed Auto-Compact. Starrsed RS User Manual Page 301 Version 6.70 MRN-071 (A)-EN...
  • Page 304 ILUTER START SENSOR Diluter start sensor 400-700 Measure sensor check • Go to tab M -> C . Select C box. AINTENANCE HECK SENSOR HECK MEASURE SENSOR Measure sensor MS 40..50..60 Page 302 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 305 ], click button [P 68)] AINTENANCE RIME LEAN ON PAGE and perform the [P ] function. RIME ILUENT ILUTER 7. Wipe outer surface and stainless steel plate below the pipettes with disinfectant. Starrsed RS User Manual Page 303 Version 6.70 MRN-071 (A)-EN...
  • Page 306 The use of a toothbrush and detergent is recommended. 1. Carefully scrub the fill nozzle inner part. 2. Use a tissue to dry the fill nozzle. Replace O-ring: Remove the O-ring. (QWLV050004) Install new O-ring. (QWLV050004) Starrsed RS User Manual Page 304 Version 6.70 MRN-071 (A)-EN...
  • Page 307 6. Place one end of the tube in the pump plate holder. 7. Pull the new tube over the peristaltic pump rotor. 8. Pull pump tube slightly downwards and at the same time towards the back of the Starrsed Auto-Compact. Starrsed RS User Manual Page 305 Version 6.70 MRN-071 (A)-EN...
  • Page 308 4. When all the pipettes are filled, the needle goes back to the home position and the adapter is moved to the release position and will be transported into the End-pool. Page 306 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 309 Be careful, as there may be blood in the cassette. First, make up some disinfectant and put this in the liquid separator. Press P to pump disinfectant through the pump cassette. RIME ISINFECTANT Starrsed RS User Manual Page 307 Version 6.70 MRN-071 (A)-EN...
  • Page 310 7. Remove the protection caps on from the tubes. 8. Connect the two tubes to new waste pump cassette. 7. Exchange Pinch valve tube Replace the pinch valve tube ESRI010246 Page 308 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 311 7. Wipe outer surface and stainless steel plate below the pipettes with disinfectant. 10. Exchange air filter Air filter replacement QWLV040003 1. Pull both tube connectors out of the blue filter. 2. Place new blue filter Starrsed RS User Manual Page 309 Version 6.70 MRN-071 (A)-EN...
  • Page 312 3. Reconnect the tube connectors on the filter If applicable: Check your on-board diluent container Recommendation is to clean the bottle when you replace the 5 liter container Diluent. Page 310 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 313 The use of a toothbrush and detergent is recommended. 1. Carefully scrub the fill nozzle inner part. 2. Use a tissue to dry the fill nozzle. Replace O-ring: Remove the O-ring. (QWLV050004) Install new O-ring. (QWLV050004) Starrsed RS User Manual Page 311 Version 6.70 MRN-071 (A)-EN...
  • Page 314 6. Place one end of the tube in the pump plate holder. 7. Pull the new tube over the peristaltic pump rotor. 8. Pull pump tube slightly downwards and at the same time towards the back of the Starrsed Auto-Compact. Page 312 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 315 ILUTER START SENSOR Diluter start sensor 400-700 Measure sensor check • Go to tab M -> C . Select C box. AINTENANCE HECK SENSOR HECK MEASURE SENSOR Measure sensor MS 40..50..60 Starrsed RS User Manual Page 313 Version 6.70 MRN-071 (A)-EN...
  • Page 316 10. Push the liquid separator towards the rear, with the sensor connectors in the holes. 11. Release the stainless steel vacuum tube. 12. Replace the waste container. 13. Close left cover. Page 314 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 317 Fill Stop sensor check • Go to tab M -> C . Select C box. AINTENANCE HECK SENSOR HECK ILL STOP SENSOR Fill stop sensor FS 90..140..165 Diluter Start sensor check Starrsed RS User Manual Page 315 Version 6.70 MRN-071 (A)-EN...
  • Page 318 10. Push the liquid separator towards the rear, with the sensor connectors in the holes. 11. Release the stainless steel vacuum tube. 12. Replace the waste container. 13. Close left cover. Page 316 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 319 7. Remove the protection caps on from the tubes. 8. Connect the two tubes to new waste pump cassette. 7. Exchange Pinch valve tube Replace the pinch valve tube ESRI010246 10. Replace diluter syringe tip Replace diluter tip Starrsed RS User Manual Page 317 Version 6.70 MRN-071 (A)-EN...
  • Page 320 D ETTINGS ENERAL SETTINGS ISPLAY DILUTION If applicable: 6. Check your on-board diluent container Recommendation is to clean the bottle when you replace the 5 liter container Diluent. Page 318 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 321 8. Check if the information is the same as on the barcode label. If also the barcode could not be detected, the tube position is logged as empty. In that case the barcode reader itself must be adjusted. Starrsed RS User Manual Page 319 Version 6.70 MRN-071 (A)-EN...
  • Page 323: Glossary Of Terms

    18.3°C. instrument. 18.1.1.1.4. ESR is short for Erythrocyte Sedimentation Rate. It is the amount of sedimentation (setting) of erythrocytes (red blood cells) in a blood column during a specified time. Starrsed RS User Manual Page 321 Version 6.70 MRN-071 (A)-EN...
  • Page 324 Starrsed Auto-Compact to the HOST. Only sample results and result related messages are send. 18.1.1.1.12. WI is short for Work Instruction and is used with an index number for a range of work instructions. Page 322 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
  • Page 325: Index

    Appendix - 30 minutes reporting • 224 Communication error (E2) • 212 Appendix - 60 minutes reporting • 222 Communications • 211 Appendix - Article reference list Starrsed RS • Compact bi-directional protocol • 132 25, 218 Compact connected to • 104 Appendix - Cable connection diagram •...
  • Page 326 Display sample history (QC) • 57 GETTING STARTED • 111 Document history overview • 3 Gripper module • 208, 209 E.C. Declaration Starrsed RS • 32 Hazy • 323 EDTA mode • 323 Hazy reports • 201 End-of-day-wash options • 79 Hazy reports after cleaning •...
  • Page 327 Rinse solution spilling over the instrument • Protocol MECHATRONICS-02 unidirectional • 129 Rinse-pump tube replacement • 177 Protocol Opus bidirectional • 134 Protocol Sysmex R-3500 • 140 Protocol Sysmex R-3500 EPU • 145 Starrsed RS User Manual Page 325 Version 6.70 MRN-071 (A)-EN...
  • Page 328 Starrsed line of ESR instruments • 25 Starrsed Rinse solution • 24 Starrsed Rinse solution QRR 010934 • 113 STARRSED RS PROGRAM • 35 Starrsed Saline • 24 Starrsed Saline QRR 010933 • 113 Stickers for the onboard reagents bottles • 33 Stickers of the reagents containers •...