Page 5
August 2014 Service Screen Temporary release Schavemaker • 6.66 November History Screen Introduction of chapter Quality 2013 Schavemaker Control Settings Screen • Barcode reader/tube detection Quality Control rewritten Trouble shooting Starrsed RS User Manual Page 3 Version 6.70 MRN-071 (A)-EN...
Page 6
Program version 2.22 • 6.30 September Removed H.E. van Dijk Removed the installation section 2008 installation section from the manual and insert this in from manual Installation manual. Page 4 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 7
Correction made in WI 191, 192, Insert section 193, 194 5.6.1.2 Conveyer • motor test Add in settings-carrousel the pipette flow diagram pop-up Insert in the 5.4 the fill and clean screens procedure Starrsed RS User Manual Page 5 Version 6.70 MRN-071 (A)-EN...
Page 8
Make corrections and add the 2004 latest pictures • Update the sections according the latest version of the Graphic User Interface • Preliminary November H.E. van Dijk Start of he manual 2003 Page 6 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Medical requirements: ................. 30 5.2.3. General laboratory practices: ..............30 5.2.4. Specific laboratory practices and requirements: .......... 31 5.3. E.C. Declaration Starrsed RS ................32 5.4. Labels and stickers on containers ................33 5.4.1. Stickers for the onboard reagents bottles ............ 33 5.4.2.
Page 10
6.5.2.6. Separator sensor ..................74 6.5.2.7. Flow sensor .................... 74 6.5.2.8. Measure sensor ..................75 6.5.3. Display error history ..................76 6.5.4. Display maintenance history ................ 76 6.5.5. Maintenance info ..................77 Page 8 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 12
Turn off sequence ..................164 QUALITY CONTROL ....................165 10.1. Control pipettes ....................165 10.2. Monitoring measurement quality with Starrsed Control ........165 10.2.1. Limitations ....................165 10.2.2. Expected value range ................166 Page 10 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 13
Rinse solution spilling over the instrument ..........186 15.1.3. Rinse pump failure ..................186 15.1.4. Sample probe is not washed after aspiration ..........186 15.1.5. Saline dripping in the sample tube adapter ..........187 Starrsed RS User Manual Page 11 Version 6.70 MRN-071 (A)-EN...
Page 14
New rack adapter ..................207 15.21. Rack unit errors .................... 207 15.21.1. Gripper module ..................208 15.21.2. Mixer module ..................209 15.21.3. Check the mixer ..................209 15.22. Communications ................... 210 Page 12 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 15
Communication error (E2) ................211 15.28. Quality control trouble shooting ..............213 APPENDIX FOR STARRSED AUTO-COMPACT ............ 216 WORK INSTRUCTION STARRSED AUTO-COMPACT ........... 268 GLOSSARY OF TERMS ..................321 INDEX........................323 Starrsed RS User Manual Page 13 Version 6.70 MRN-071 (A)-EN...
To be used for all patient populations, not restricted by age or any other anatomical or physiological particulars. 1.2. Indication for use of device For screening and follow-up of patients with (suspected) inflammatory conditions. Page 14 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
EDTA blood sample tubes. The system uses rack adapters and is set up for a specific type of rack. • Windows based System Software is running on an external computer. Starrsed RS User Manual Page 15 Version 6.70 MRN-071 (A)-EN...
Warning sign to prevent personal injury due to hot surfaces. Warning sign to prevent personal injury due to electrical shocks. General warning. General note. This symbol indicates a reference to this or other product documentation Page 16 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Introduction 1.5. EQAS EQAS is RR Mechatronics' comprehensive tool to periodically assess the performance of the Starrsed instrument at your lab. Through EQAS, the test results of the instrument, obtained with Starrsed Control are objectively compared with a worldwide peer group using the same type of instrument.
Results of the test are expressed in millimeters. This data together with the patient ID number is sent to the Laboratory Information System along with the sedimentation time used (60 or 30 minutes), the temperature and the dilution ratio. Starrsed RS User Manual Page 19 Version 6.70 MRN-071 (A)-EN...
Page 22
This model uses large bulk containers for reagent supply instead of the 0.5-liter onboard containers. This means longer operation time without reagent preparation and no cleaning time. This model is delivered with longer level sensors and special reagents cover assembly. Page 20 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
QRR 010931 Starrsed Diluent Reagents used QRR 010947 Starrsed Disinfectant QRR 010933 Starrsed Saline QRR 010934 Starrsed Rinse solution De-ionized water Blood volume: Aspirated blood volume per 1.4 ml in EDTA mode sample Starrsed RS User Manual Page 21 Version 6.70 MRN-071 (A)-EN...
Page 24
20 Gb Hard disk on external PC Storage medium approx. 5 Mb per 1000 samples (results and raw data) Storage capacity indication LIS connectivity: Ethernet and serial, bi-directional according various Communication protocols Page 22 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Table space (incl. PC) 770 mm Height Weight empty (ESR 70 kg unit+rack) Table size (Min.) 1100 mm Width (excl. space for PC) 1500 mm Width (incl. space for PC) 660 mm Depth Starrsed RS User Manual Page 23 Version 6.70 MRN-071 (A)-EN...
Starrsed Disinfectant is a dedicated IVD product, developed exclusively for automatic disinfection of the waste system of Starrsed ESR analyzers. Approximately 0.5 ml disinfectant is used after each pipette rinse. This reagent is delivered in a 5-liter container (QRR 010947) Page 24 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Starrsed ST, Starrsed RS, Starrsed RL and Starrsed TL are instrument names of RR Mechatronics. Until 2014 the Starrsed ST was called Inversa 24M, the Starrsed RS was called AutoCompact, the Starrsed RL was called Interrliner (Standalone). RR Mechatronics changed the product names to benefit better from the worldwide recognizability of the brand name Starrsed.
Example: • A collision with a moving object or a person. • Something falling on the instrument. • Liquids spilling into the instrument. Page 26 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
The instrument must be unpacked, installed and checked by a trained engineer prior to first operation. Detailed installation instructions are given in the Starrsed Auto-Compact Installation manual. Connect the instrument only to a properly earthed power connection. Page 28 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Transfer materials according to federal and international regulations. • Be familiar with written instructions for laboratory procedures and proper responses to emergencies. • Report spills, exposures, illnesses, and injuries immediately. Starrsed RS User Manual Page 29 Version 6.70 MRN-071 (A)-EN...
Departmental Safety Guidelines and Bio safety Policy. Each user is responsible to leave a clean, disinfected and tidy work place. All biohazard waste must be properly disposed. Page 30 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Dispose of gloves into a biohazard container. The use of eye protection is advised while processing samples. Remove and properly dispose of gloves and wash hands before leaving the laboratory. Starrsed RS User Manual Page 31 Version 6.70 MRN-071 (A)-EN...
Standard Operating Procedures (S.O.P.) 5.4. Labels and stickers on containers 5.4.1. Stickers for the onboard reagents bottles Starrsed RS User Manual Page 33 Version 6.70 MRN-071 (A)-EN...
Starrsed RS program 6. STARRSED RS PROGRAM The Starrsed Auto-Compact is controlled via an external computer on which runs the Starrsed Auto-Compact software. The software functions are grouped on six tabbed screens. The software is controlled by mouse pointer or directly via the touch screen. A virtual keyboard is automatically displayed on screen, when numerical or alphanumerical input is required.
The following screens are selectable via the associated tabs: 1. S screen (on page 36) AMPLE 2. H screen (on page 42) ISTORY 3. R screen (on page 64) EAGENTS 4. M screen (on page 67) AINTENANCE Page 36 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 39
When the Compact is running in the NORMAL MODE, the instrument uses the standard saved settings. For instance the ESR time is set back to 60 minutes or 30 minutes according the used method. Starrsed RS User Manual Page 37 Version 6.70 MRN-071 (A)-EN...
This is a graphical representation of the pipette. It is generated from the results of the ESR measurement. It can be used to locate possible air bubbles. Sample mode button: This is the button to start or stop the run mode of the instrument. Page 38 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 41
(EDTA or Citrate), selected method (60 or 30 minute) and symbols that draw attention to certain maintenance conditions or QC sample status (if applicable). Manual input sample ID: Can be used for manual input of a barcode from a sample tube. Starrsed RS User Manual Page 39 Version 6.70 MRN-071 (A)-EN...
A virtual number pad is shown. Type the number of the requested pipette and press the OK button. The pipette information screen is shown. Page 40 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
The sample has been measured and the pipette is marked to be washed when it reaches the rinse station. This indicator is also on when a sample could not be aspirated properly. Starrsed RS User Manual Page 41 Version 6.70 MRN-071 (A)-EN...
END ALL TO HOST: Send only the selected patient result to the HOST. END PATIENT RESULT TO • (on page 47) ISPLAY RACK HISTORY (on page 48) ISPLAY RACK DETAILS Page 42 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
After measuring the pipette, the pipette data is transferred to the sample history files. In the header of the table the names of the columns are shown. Double-click the header of any column to sort the table by this column in ascending order. Starrsed RS User Manual Page 43 Version 6.70 MRN-071 (A)-EN...
In the header of the table the names of the columns are shown. Double-click the header of any column to sort the table by this column in ascending order. Page 44 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Starrsed RS program 6.3.2.1. Display patient results Starrsed RS User Manual Page 45 Version 6.70 MRN-071 (A)-EN...
Page 48
Pipette in which the sample was measured. Error code Shows any ESR error or warning code (e.g. "Too many borders found"). A complete overview of measurement data is shown with activating the option "Display raw data". Page 46 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
ESR was measured, but with fill errors. The Clear rack history button will clear the contents of the rack history file and restart to build-up a new rack history file. Starrsed RS User Manual Page 47 Version 6.70 MRN-071 (A)-EN...
The last 10 racks are stored and can be selected. The Clear rack history button will clear the contents of the rack history file and restart to build-up a new rack history file. Page 48 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
ESR errors (%) Increasing ESR errors may indicate the need for maintenance, see ESR Error (on page 156) • Number of samples This can be used to document variations in work load. Starrsed RS User Manual Page 49 Version 6.70 MRN-071 (A)-EN...
Linked lab ID: The Lab ID is given if it is linked to the Starrsed Control sample ID Sampling date: The date and time when the QC sample was aspirated. Page 50 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 53
Results can be exported to a .CSV file and imported in an MS Excel file for further analyses. ATCH All used batches of Starrsed Control are shown, results are shown for chosen batch ID. LOSE Return to History screen (on page 42). Starrsed RS User Manual Page 51 Version 6.70 MRN-071 (A)-EN...
Starrsed RS program 6.3.5.2. QC normal results screen extended After scrolling the general data from the QC results are shown. 6.3.5.3. QC normal results (graph) Page 52 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
This graph gives a first indication of the measuring stability of the Starrsed Auto-Compact. Further analysis and identification of systematic errors have to be performed in the user's Quality Control System. LOSE Return to History screen (on page 42). 6.3.5.4. QC abnormal results (table) Starrsed RS User Manual Page 53 Version 6.70 MRN-071 (A)-EN...
Page 56
After these columns additional data is shown: pipette number, dilution rate, ESR30, ESR time and Aspect. Scroll to the right. Results are always shown with and without Temperature correction, independent of the setting (ON or OFF). ORRECTION The following options can be selected: Page 54 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
All used batches of Starrsed Control are shown, results are shown for chosen batch ID. LOSE Return to History screen (on page 42). 6.3.5.5. QC abnormal results screen extended After scrolling the general data from the QC results are shown. Starrsed RS User Manual Page 55 Version 6.70 MRN-071 (A)-EN...
This graph gives a first indication of the measuring stability of the Starrsed Auto-Compact. Further analysis and identification of systematic errors have to be performed in the user's Quality Control System. LOSE Return to History screen (on page 42). Page 56 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
44)" screen. Depending on the frequency of QC samples, related patient results may span over multiple days and are listed per date. All general ESR data and errors of QC samples are shown here. Starrsed RS User Manual Page 57 Version 6.70 MRN-071 (A)-EN...
- QC UTOMATICALLY REMOVE LINKED AFTER RESULT ETTINGS 92)) the links can be removed automatically when a usable ESR result has been ETTINGS ON PAGE reported for this particular Lab ID. Page 58 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
By selecting Dilution <= 90 all the samples with a dilution rate <= 90 are displayed in the table. In the header of the table the names of the columns are shown. Double-click the header of any column to sort the table by this column in ascending order. Starrsed RS User Manual Page 59 Version 6.70 MRN-071 (A)-EN...
In the header of the table the names of the columns are shown. Double-click the header of any column to sort the table by this column in ascending order. Page 60 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
In the header of the table the names of the columns are shown. Double-click the header of any column to sort the table by this column in ascending order. Starrsed RS User Manual Page 61 Version 6.70 MRN-071 (A)-EN...
3. A range of days from start date to end date. 6.3.12. History sample analyse option day Make a selection for all of today's results or only today's results between start time and end time. Page 62 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Starrsed RS program 6.3.13. Set start date Enter the Start date and time. 6.3.14. Set end date Enter the End date and time. Starrsed RS User Manual Page 63 Version 6.70 MRN-071 (A)-EN...
! appears. Processing of new samples EAGENTS is stopped. A log is available for all reagents and can be accessed by clicking on D ISPLAY REAGENT (on page 65). HISTORY Page 64 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Select the reagent type on the right side. For external use of the information all the logged reagent data can by exported to EXCEL .CSV format by clicking Export to Excel (CSV). 6.4.2. New reagent input Starrsed RS User Manual Page 65 Version 6.70 MRN-071 (A)-EN...
3. Enter / read the Expiry date (if barcode reader is used: cursor has to be in one of the two boxes) 4. If necessary, adjust the date when the reagent was placed. 5. Check if the preview box shows the correct information, then press OK. Page 66 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
2. C (on page 71) HECK SENSORS 3. D (on page 76) ISPLAY RROR HISTORY 4. D . (on page 76) ISPLAY AINTENANCE HIST 5. M (on page 77) AINTENANCE Starrsed RS User Manual Page 67 Version 6.70 MRN-071 (A)-EN...
When the Starrsed Auto-Compact has been idle for more than eight hours, there might be some backflow of reagents due to gravity. Prime all tubing before sampling by using the function: • RIME ALL UNITS All priming functions are sequentially performed one time. Page 68 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 71
This function fills all pipettes with a cleaning agent and removes the cleaning agent after a specified time. • End-of-day wash: All pipettes will be washed once and needle, fill-nozzle and rinse-nozzle (wash station) are primed. Starrsed RS User Manual Page 69 Version 6.70 MRN-071 (A)-EN...
See chapter Maintenance Fill and clean procedure (on page 179) for more information. 6.5.1.2. End-of-day-wash procedure • End-of-day wash: All pipettes will be washed once and needle, fill-nozzle and rinse-nozzle (wash station) are primed. Page 70 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
After finishing the test, both LED's must be green. • Check Separator sensor: Click the Check button. The green light is shown if the sensor value is in range. Starrsed RS User Manual Page 71 Version 6.70 MRN-071 (A)-EN...
Page 74
"ESR (mm)". When the sensor is out of range and the light is red, the sensor values (on page 73) can be checked by turning on the service mode. Page 72 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
When there is a sensor alarm, an alarm indicator is shown in the tab M AINTENANCE All functions to check the status of the sensors are grouped under button C (on page HECK SENSORS 71). Starrsed RS User Manual Page 73 Version 6.70 MRN-071 (A)-EN...
Check Flow sensor: The vacuum unit switches on and the values must be in this range: For Starrsed Auto-Compact Flow: 0925-0980-1020 Abs: 0300-360-0390 Note: If for example the yellow orifice is blocked the flow will be: Offset: 0045-0050-0055. Page 74 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Clean sensors first before executing this function. When a test pipette is installed at the measuring position the result of the test pipette is displayed in the field "ESR (mm)". Starrsed RS User Manual Page 75 Version 6.70 MRN-071 (A)-EN...
When there is a sensor alarm, an alarm indicator is shown in the tab M AINTENANCE All performed maintenance functions are logged automatically. This log can be saved e.g. to a memory stick by clicking button Save As ... Page 76 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Press the button Info to open the work instruction for a specific maintenance level. When this maintenance is done press the button Done to log the completed work in the maintenance log file. Starrsed RS User Manual Page 77 Version 6.70 MRN-071 (A)-EN...
All pipettes will be washed once, needle, fill-nozzle and rinse-nozzle (wash station) are primed. Select the time of the day in hours and minutes for automatic start of this function. Page 78 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Only once: The function runs only once at the selected time. • Weekdays: The function runs only on working days (monday till friday) at the selected time. • Daily: The function runs on a daily base at the selected time. Starrsed RS User Manual Page 79 Version 6.70 MRN-071 (A)-EN...
If virtual keyboard is switched ON a virtual keyboard pops up for input the data. When Virtual keyboard is switched OFF the arrow keys must be used for input of data. When ready with press S before you continue. AVE SETTINGS Page 80 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 83
ON: The carousel is filled with optimum sample speed. The sample tubes in the rack AST FILLING are faster available for the user but the carousel shows more unused pipettes. Starrsed RS User Manual Page 81 Version 6.70 MRN-071 (A)-EN...
Page 84
Select the language for the software and the "Instructions For Use" by clicking on the appropriate language name. Note: To switch to the selected language the first time, close and restart the software. Page 82 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Starrsed RS program 6.6.1. Carousel control Starrsed RS User Manual Page 83 Version 6.70 MRN-071 (A)-EN...
Page 86
Select the language for the software and the "Instructions For Use" by clicking on the appropriate language name. Note: To switch to the selected language the first time, close and restart the software. Page 84 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
This test is a useful function for checking the pipette position adjustment, vacuum adjustment and filling height sensor positioning adjustment. Each individual pipette is tested and results are displayed in their own line of the diagram. Starrsed RS User Manual Page 85 Version 6.70 MRN-071 (A)-EN...
This test is a useful function for checking the pipette position adjustment, vacuum adjustment and filling height sensor positioning adjustment. Each individual pipette is tested and results are displayed in their own line of the diagram. Page 86 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
This test is a useful function for checking the pipette position adjustment, vacuum adjustment and filling height sensor positioning adjustment. Each individual pipette is tested and results are displayed in their own line of the diagram. Starrsed RS User Manual Page 87 Version 6.70 MRN-071 (A)-EN...
HANGE INSE POSITION key for learning a new position table. EARN CAROUSEL POSITION Note: This option is used when there is a position error. Page 88 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Example: Dilution error detection is set at 10%. When the dilution error is outside the 10% range, in the last column of the report EDTA 079 or EDTA 121 is printed which indicating this sample is 21% under or over diluted. Starrsed RS User Manual Page 89 Version 6.70 MRN-071 (A)-EN...
When the flow sensor is still giving errors after trouble shooting and there are no detectable faults in the liquid flow, use the switch OFF function. This check is now not active, call for service. Page 90 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
YES: always transmit results with bubbles on top warning to the output. set to NO: transmit no results to the output when there is generated a warning for bubbles on top. Starrsed RS User Manual Page 91 Version 6.70 MRN-071 (A)-EN...
Note: If limit error filters in the LIMS are not corresponding with the limit error settings on the instrument it is possible that not all expected results are send to the LIMS. 6.6.4. QC Settings Page 92 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 95
This link will stay active if there is no result or a general ESR error is generated. This setting is useful when the lab issues a new and unique Lab ID for every QC sample. Starrsed RS User Manual Page 93 Version 6.70 MRN-071 (A)-EN...
There are no restrictions for the file AVE SETTINGS AS name. • is used for storing software settings after settings are changed or altered. AVE SETTINGS Page 94 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
This screen gives access to: 1. Mixer motor (on page 96) settings. 2. Conveyer motor (on page 97) settings. 3. Position motor (on page 97). 4. Timing settings (on page 100). Starrsed RS User Manual Page 95 Version 6.70 MRN-071 (A)-EN...
OLD LEVEL 4. H : default 20% OMING SPEED After setting use Save settings to save the new settings. If Save settings is not pressed the old settings will be used Page 96 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Starrsed RS program 6.7.1.2. Conveyer motor 6.7.1.3. Position motor Starrsed RS User Manual Page 97 Version 6.70 MRN-071 (A)-EN...
Page 100
Rack out position. : Service purpose only. This is to move from the home position to any distance in O TO POSITION millimeters from the home position. Page 98 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 101
IX POSITION 3. Go to T UBE POSITION NUMBER After pressing Read barcode the barcode is shown above the button. If barcode label is not read correctly see section Troubleshooting. Starrsed RS User Manual Page 99 Version 6.70 MRN-071 (A)-EN...
For service purpose only. Those settings are set in the factory, no need for altering the timing settings. The timing settings T8 and T11 are used for the mixer setting. Page 100 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
When selected the following keys will pop-up. Checksum On/Off. 30 Minute method On/Off. 16. Advia 120 17. Advia 120 unidirectional 18. InteRRliner For more details see Section 9 Protocols (on page 115) and the Appendix Page 102 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
OFF, this file cannot be selected in the history window. EARCH IN EXAMPLE HISTORY • If S is ON, this can be selected in the history window. EARCH IN EXAMPLE HISTORY Starrsed RS User Manual Page 103 Version 6.70 MRN-071 (A)-EN...
ON, this can be selected in the history window. EARCH IN EXAMPLE HISTORY 6.7.3.7. Outer needle depth Setting for the O . No need to change the default 51.0 mm setting. UTER NEEDLE DEPTH Page 104 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Note: If vacuum pump is OFF rinse solution will spill over the Auto rack unit OFF: LL PUMPS All active pumps will be switched OFF. The waste pump is switched ON for one minute. Starrsed RS User Manual Page 105 Version 6.70 MRN-071 (A)-EN...
: The Rinse nozzle is send to the rinse nozzle home position. INSE NOZZLE DOWN Led status display • Led status check for the fill- and rinse nozzle is displayed. Page 106 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
This list contains a log of all the maintenance what is done which can be useful for the field engineer to check the problems with the instrument in the past. The key C will delete all errors from the list. LEAR Page 108 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
This list contains error numbers what can be useful for the field engineer to check the problems with the instrument in the past. The key C will delete all errors from the list. LEAR Starrsed RS User Manual Page 109 Version 6.70 MRN-071 (A)-EN...
There are no restrictions for the file AVE SETTINGS AS name. • is used for storing software settings after settings are changed or altered. AVE SETTINGS Page 110 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
2. Fill up the containers with the reagents. 3. Place the reagent containers. 4. Replace screw caps with level sensors. Starrsed RS User Manual Page 111 Version 6.70 MRN-071 (A)-EN...
Page 114
4. Install the level sensors and spacers according the following pictures. Make sure to place the appropriate level sensors in the containers by matching the color codes on the tube and on the container: Page 112 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
The latest version of the Material Safety Data Sheet (MSDS) of the used reagents can be found on our web site www.rrmechatronics.com. 7.3.4. De-ionised water Add one or two drops of Starrsed Saline to the deionized water to avoid <bottle empty alarm>. Starrsed RS User Manual Page 113 Version 6.70 MRN-071 (A)-EN...
(biohazardous) material and disposed of according to local regulations. Preferably, discard the complete waste container and replace it with a cleaned one. Clear the error. Disclaimer: Check your local environment rules about discharging the waste. Page 114 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
17. Advia 120 unidirectional The protocol can be set in tab S - LIMS settings. After selecting a protocol, save the new ERVICE settings by pressing the Save setting key. Starrsed RS User Manual Page 115 Version 6.70 MRN-071 (A)-EN...
Limit error Error One of the following limits are out of the setting range: • ESR Time • Column height • Dilution • Bubbles on top • Hazy aspect • Temperature Page 116 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
30 min 80 data characters cs eot Checksum - ON 30 min Data consists, if 30 min. output is switched to ON Positio Description Format Patient identification text Text PPPPPPPPPP Starrsed RS User Manual Page 117 Version 6.70 MRN-071 (A)-EN...
Page 120
Text.: left aligned followed by spaces (ASCII $20). xx.…: number made up of (xx...) digits 0 9 (ASCIII $30 $39) with leading zeros. Leading zeros and non-specified positions are filled with spaces (ASCII $20). Page 118 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
ASCII string sum = the ASCII sum of all preceding characters incl. stx, cr, lf and ext modulo 256 (ASCII string sum) = the remainder of the ASCII string sum when divided by 256. Starrsed RS User Manual Page 119...
Text.: left aligned followed by spaces (ASCII $20). xx.…: number made up of (xx...) digits 0 9 (ASCIII $30 $39) with leading zeros. Leading zeros and non-specified positions are filled with spaces (ASCII $20). Page 120 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
]PPPPPPPPPP[ ]WWWW[ ][ ][ ][ ][ ][ ][ ][ ]AAAAAAAAAAA[cr][lf] - Or -: Result string with Error: pp[ ]PPPPPPPPPP[ ][ ][ ][ ][ ][ ][ ][ ][ ][ ][ ][ ][ ]EEEEEEEEEEE[cr][lf] Starrsed RS User Manual Page 121 Version 6.70 MRN-071 (A)-EN...
Page 124
If the Compact is switched to the 30 min method the output string has the same format. The Compact automatic apply the conversion table to the 60 min method. If temperature correction is switched on the ESR value will be the temperature corrected ESR value. Page 122 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
If temperature correction is switched on, the ESR value will be the temperature corrected ESR value. • If the result has an error, the ESR value will be 4 space characters (ASCII $20). Starrsed RS User Manual Page 123 Version 6.70 MRN-071 (A)-EN...
(xx...) digits 0 9 (ASCII $30 $39) Leading zeros and non-specified positions are filled with spaces (ASCII $20). The following error codes are defined: No cells / plasma found RROR ESR Probably >140 mm RROR Page 124 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 127
If the Compact is switch to the 30 min method the output string has the same format. The Compact automatics apply the conversion table to the 60-minute method. If temperature correction is switched on the ESR value will be the temperature corrected ESR value. Starrsed RS User Manual Page 125 Version 6.70 MRN-071 (A)-EN...
Workorder record from LIMS to analyzer: ESR = Yes (capital Y on position 125) Data "Y" space ESR = No (capital N on position 125) Data space "N" MECHATRONICS result record Page 126 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 129
Error code xxxxx See section ESR error codes Limit error message (results) text See section Limit error message (on page 158) Reserved (spaces) text Checksum [CS] See section Checksum Starrsed RS User Manual Page 127 Version 6.70 MRN-071 (A)-EN...
Page 130
Measure error ARNING Bubbles on top ARNING Limit error ,no result transmitted!, see RROR Limit error messages (on page 158) Note: See analyzer manual for more information about limit error settings! Page 128 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 131
Dilution rate (%) xxxxx Column height (mm) xxxxx Error code xxxxx See section ESR error codes Limit error message (results) text See section Limit error message (on page 158) Starrsed RS User Manual Page 129 Version 6.70 MRN-071 (A)-EN...
Page 132
(digits 0-9) with leading spaces Non-specified positions are filled with spaces Sample codes Aspect codes Sample type Transmitted code Aspect Transmitted code Patient sample Clear QC normal Hazy<10 QC abnormal Hazy<25 Hazy>25 Page 130 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 133
CS cannot be equal to that of the ETX byte (03h). Therefore, if the calculated CS value is 03h, the transmitted CS byte is set to the substitute value 83h in order to avoid erroneous action by the receiving device. Starrsed RS User Manual Page 131 Version 6.70 MRN-071 (A)-EN...
EXOR those 2. Take the result and EXOR this with position 3 ect till the position At position 28: means that this is the Request string Char $02 Char $03 Page 132 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 135
Acknowledge (char $06) if the string is found ok. If either the host computer as well the Starrsed Auto-Compact receives a string it should reply by sending a Non Acknowledge (char $15) if the string is found faulty. Starrsed RS User Manual Page 133 Version 6.70 MRN-071 (A)-EN...
If either the hos t c omputer as well the Starrs ed Auto-Compact r ecei ves a string it should repl y by s endi ng a Non Ac knowl edg e (c har $15) if the string is found faul ty. At position 28: means that this is the Request string Char $02 Char $03 Page 134 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 137
If either the hos t c omputer as well the Starrs ed Auto-Compact r ecei ves a string it should repl y by s endi ng a Non Ac knowl edg e (c har $15) if the string is found faul ty. At position 28: means that this is the Request string Char $02 Char $03 Starrsed RS User Manual Page 135 Version 6.70 MRN-071 (A)-EN...
Page 138
After 3 attempts the Host-computer must stop sending the conformation string and the Starrsed Auto-Compact won’t do an ESR on this sample and goes to the next sample. The Host-computer needs to be ready for the request string of the next sample! Page 136 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 139
If the Starrsed Auto-Compact does not get any response from the host, (timeout 10 sec) it will send the result string again. After 3 attempts it will stop communication and generate an error. Starrsed RS User Manual Page 137 Version 6.70 MRN-071 (A)-EN...
Page 140
ASCII string sum = the ASCII sum of all preceding characters incl. stx, cr, lf and ext modulo 256 (ASCII string sum) = the remainder of the ASCII string sum when divided by 256. OR 128 = setting the MSB-bit to 1. Page 138 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 141
Text.: left aligned followed by spaces (ASCII $20). xx.…: number made up of (xx...) digits 0 9 (ASCIII $30 $39) with leading zeros. Leading zeros and non-specified positions are filled with spaces (ASCII $20). Starrsed RS User Manual Page 139 Version 6.70 MRN-071 (A)-EN...
10100 Dilution rate = 101 12300 30 minute ESR value = 123 Reserved 0000… n.a. n.a. = not applicable R3500 Sample data flag record format (131 bytes) # Of Parameter chars Example Comment Page 140 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 143
0 0 = sample does not exist 1 = sample exists Date ordered "yyyymmdd" Sample ID no. Must be the same as inquiry record Rack no. Tube position no. Inquiry mode Patient ID no. Starrsed RS User Manual Page 141 Version 6.70 MRN-071 (A)-EN...
Page 144
The code appears in the "sample data record" at variable 'RBC' Compact "ERROR" code messages (see General Error codes in protocols (on page 116)) This code appears in the "sample data record" at variable 'HGB'. Page 142 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
10100 Dilution rate = 101 12300 30 minute ESR value = 123 Reserved 0000… n.a. n.a. = not applicable R-3500 Sample data flag record format (131 bytes) # Of Parameter chars Example Comment Starrsed RS User Manual Page 143 Version 6.70 MRN-071 (A)-EN...
Page 146
The code appears in the "sample data record" at variable 'RBC' Compact "ERROR" code messages (see General Error codes in protocols (on page 116)) This code appears in the "sample data record" at variable 'HGB'. Page 144 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Reserved 00.. n.a. n.a. = not applicable * Note: Visible in combination with software version 5.07 and up ** Note: With leading spaces, applicable from software version 5.08 and up Starrsed RS User Manual Page 145 Version 6.70 MRN-071 (A)-EN...
Page 148
The code appears in the "sample data record" at variable 'RBC' Compact "ERROR" code messages (see General Error codes in protocols (on page 116)) This code appears in the "sample data record" at variable 'HGB'. Page 146 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
0 n.a. Reserve 0 n.a. Reserve 0 n.a. Reserve 0 n.a. Reserve 0 n.a. IP message information 000000 n.a. 123000 Esr value = 123 mm 102000 Hazy code = 12 Starrsed RS User Manual Page 147 Version 6.70 MRN-071 (A)-EN...
Page 150
Patient number (same as sample data) Rack no. 1234 Rack number = 1234 Tube position no. 03 Tube position in rack = 3 SE9000 order information record format (171 bytes) Parameter # of Example Comment Page 148 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 151
The code appears in the "sample data record" at variable 'RBC' Compact "ERROR" code messages (see General Error codes in protocols (on page 116)) This code appears in the "sample data record" at variable 'HGB'. Starrsed RS User Manual Page 149 Version 6.70 MRN-071 (A)-EN...
0 n.a. Reserve 0 n.a. Reserve 0 n.a. Reserve 0 n.a. Reserve 0 n.a. IP message information 000000 n.a. 123000 Esr value = 123 mm 102000 Hazy code = 12 Page 150 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 153
The code appears in the "sample data record" at variable 'RBC' Compact "ERROR" code messages (see General Error codes in protocols (on page 116)) This code appears in the "sample data record" at variable 'HGB'. Starrsed RS User Manual Page 151 Version 6.70 MRN-071 (A)-EN...
0 n.a. Reserve 0 n.a. Reserve 0 n.a. Reserve 0 n.a. Reserve 0 n.a. IP message information 000000 n.a. 123000 Esr value = 123 mm 102000 Hazy code = 12 Page 152 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 155
Information status exists Date ordered "yyyymmdd" Sample ID no. Must be the same as inquiry record Rack no. Tube position no. Inquiry mode Patient ID no. Patient name Patient birthday "yyyymmdd" Starrsed RS User Manual Page 153 Version 6.70 MRN-071 (A)-EN...
7. Pip.=Sedimentation pipette number (number on the pipette belt). 8. Time=Actual sedimentation time in minutes. 9. T=Temperature (in degrees Centigrade). 10. Error=Error message (if the Analyser detects an error). 11. EDTA mode. Page 154 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
5. Aspect (clear, hazy). 6. Manually entered code number. 7. Pip.=Sedimentation pipette number (number on the pipette belt). 8. Time=Actual sedimentation time in minutes. 9. T=Temperature (in degrees Centigrade). Starrsed RS User Manual Page 155 Version 6.70 MRN-071 (A)-EN...
ESR "ERROR" and "WARNING" code messages. This code appears in the "sample data record" at column 10. Error: No ESR result is given Warning: Result should be reviewed before release. The following codes are defined: Page 156 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 159
Limit error Error One of the following limits are out of the setting range: • ESR Time • Column height • Dilution • Bubbles on top • Hazy aspect • Temperature Starrsed RS User Manual Page 157 Version 6.70 MRN-071 (A)-EN...
The reporting range in the columns 2, 3 and 4 are in millimeters. The start of the measure range is at the top of the meniscus down to 140 mm. If the detection of cells/plasma is over 140 mm then the report will be >140. Page 158 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Sample is clear. Sample is Hazy < 10 Sample is Hazy < 25 Sample is Hazy > 25 Results with hazy aspect can be suppressed in the menu Limit error settings. Page 160 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
IONIZED WATER tube connected to the fill nozzle cap. 5. P , activates the disinfectant valve. Disinfectant must flow through the small tube RIME ISINFECTANT connected to the pipette wash station. Starrsed RS User Manual Page 161 Version 6.70 MRN-071 (A)-EN...
If the small onboard bottles are used, wash and keep the bottles clean to avoid bacterial growth. The Starrsed Auto-Compact has liquid level sensors. When the level sensor alarm appears, replace reagent as soon as possible. Page 162 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
When the Starrsed Auto-Compact has been idle for more than eight hours, there might be some backflow of reagents due to gravity. Prime all tubing before sampling by using the function: • RIME ALL UNITS All priming functions are sequentially performed one time. Starrsed RS User Manual Page 163 Version 6.70 MRN-071 (A)-EN...
Pressing OK activates the settings. 9.4.2. Turn off sequence • Close the Starrsed Auto-Compact software. • Switch the PC and monitor OFF. • Switch (optional) printer OFF. • Switch the Compact OFF. Page 164 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Starrsed Control should not be used past the expiry date. Mechatronics as supplier of the Starrsed Control shall not be liable for any claimed damages arising from other than intended usage. Starrsed RS User Manual Page 165 Version 6.70 MRN-071 (A)-EN...
The history of QC results is maintained internally. Error messages are generated when the QC results are outside the acceptable range. QC samples can be optionally requested by the LIMS and QC results can be send to the LIMS. Page 166 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Do not re-use empty tubes. The software interface is described in the chapter History screen (on page 42). Starrsed Control should be disposed of as medical waste. Starrsed RS User Manual Page 167 Version 6.70 MRN-071 (A)-EN...
The uncorrected result is within range, but the corrected result is out of range. See Quality control trouble shooting (on page 213) and QC Results screen (on page 50) for more details. Page 168 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
QC sample step before releasing patient results. If results are not send to the LIMS QC Results can be exported to MS Excel CSV files for further analysis in lab's own Quality Control data system. Starrsed RS User Manual Page 169 Version 6.70 MRN-071 (A)-EN...
4. Lift the left cover and place the waste container back into the Compact. Note: If you are re-cycling waste containers, make sure that they are bleached and rinsed thoroughly. Page 170 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
The cause of this problem can be foam or the waste pump is not working. The Compact will continue to measure and send the ESR results on time to the printer, but the rinse and fill sequences are stopped until the error is solved. Starrsed RS User Manual Page 171 Version 6.70 MRN-071 (A)-EN...
7. Rinse head up error. 8. Measure head not home error. 9. Separator full error The explanation of all these messages can be found in Appendix - System messages Starrsed RS Page 172 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
The purpose of the daily maintenance is to keep the instrument clean and contamination as low as possible. Clean all parts that are exposed to blood, wipe the outer surface and the stainless steel plate below the pipette belt. See WI Daily maintenance (on page 292). Starrsed RS User Manual Page 173 Version 6.70 MRN-071 (A)-EN...
A faulty or broken needle can cause a fill time-out error or a dilution error. If necessary replace the sample probe or outer needle. Remove the top cover from the Rack unit. Page 174 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
The purpose of the weekly maintenance is to carry out the daily maintenance and additionally check the optical sensor of the measure head and the vacuum pressure. Detailed instructions of this procedure can be found in the Work Instruction Weekly maintenance. (on page 293) Starrsed RS User Manual Page 175 Version 6.70 MRN-071 (A)-EN...
Detailed instructions of this procedure can be found in the Work Instruction Cleaning liquid separator (on page 282). Symptoms of a dirty separator: 1. Separator errors. 2. Foam in the separator. 3. Waste pump cannot sufficiently remove the waste from the separator. Page 176 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
If the tube is not fitted correctly or is worn the following symptoms can occur. • Liquid flowing back into the container. • First glass tube on the pipette belt is not washed sufficiently. Note: The wider bore tube is for the rinse pump. Starrsed RS User Manual Page 177 Version 6.70 MRN-071 (A)-EN...
Detailed instructions of this procedure can be found in WI Cleaning liquid separator (on page 282). As part of the Cleaning liquid separator procedure the bacterial Hepa filter QWLV040002 is replaced with a new one. Exchange bacterial filter QWLV040001 on the waste bottle assembly. Page 178 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
3. Place the two caps on the adapter and mix well. 4. Put the adapter in the Input-pool. 5. Select M , button F AINTENANCE RIME LEAN AND CLEAN Start Fill and clean procedure: 1. Select button OK. Starrsed RS User Manual Page 179 Version 6.70 MRN-071 (A)-EN...
Page 182
'hazy' aspects, check the dispenser system carefully and clean it if there is any doubt. Symptoms for a contaminated instrument: 1. If more than 3… 5 out of 20 measurements are reported HAZY. 2. Poor meniscus may be reported. 3. Incorrect results. Page 180 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
1. All tubing ESRI079200 and additional tube set. 2. Waste pump motor ESRI090920. 3. Waste pump cassette ESRI090921. 4. Blue Vacuum filter disc. Part no QWLV040003. 5. Fill block washer. Part no ESRI030906. Starrsed RS User Manual Page 181 Version 6.70 MRN-071 (A)-EN...
1. Remove level sensors and spacers from the empty container. 2. Remove empty container. 3. Place new container. 4. Remove the container screw caps and pull the necks of the bottle packs out of the cardboard box. Page 182 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 185
After each reagents change, the fluid system must be primed: 1. Select M -> P (on page 68). AINTENANCE RIME LEAN 2. Perform the applicable prime step to fill the relevant tubes with reagent and remove air. Starrsed RS User Manual Page 183 Version 6.70 MRN-071 (A)-EN...
Year\Month\Day, for example D:\2014\01\03-01-2014. Each result is stored with the sample ID as document name. QC results are stored on D:\QC\NORMAL\batchXXX and D:\QC\ABNORMAL\batchYYY. Page 184 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
See WI-162 rinse tube replacement (on page 269). 2. Check rinse solution level in rinse solution container. If the level is insufficient, a message is displayed and the alarm sounds! Starrsed RS User Manual Page 185 Version 6.70 MRN-071 (A)-EN...
It may be worn or leak. (See Level 4 maintenance). An incorrectly fitted saline peristaltic pump tube may cause the same problem. One of the tubes carrying saline maybe blocked or bended. Page 186 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
1. Liquid in the container is not detected. This occurs sometimes with the D bottle IONIZED WATER and is caused by a very low conductivity. 2. Add one or two drops of S to the D to increase the conductivity. ALINE IONIZED WATER Starrsed RS User Manual Page 187 Version 6.70 MRN-071 (A)-EN...
Trouble shooting 15.3. Compact stalls When the Compact is not working again check the main fuse. The main fuse can be found at the main input socket from the Compact. Page 188 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
3. Check all tubes are still connected. 4. Check if the 2 tubes (ESRI070508) are still connected with the tubes ESRI070406. 5. Check if the 2 tubes positioned correctly in the pinch valve. Starrsed RS User Manual Page 189 Version 6.70 MRN-071 (A)-EN...
1. Check the diluter syringe tip. 2. Check on the Mixing-piece joints and connectors for leaks and replace if necessary. 3. Check the connectors on the EDTA flow sensor for leaks and replace if necessary. Page 190 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 193
Diluent to the tubing and fill system manually. 1. Check pick-up tube at the Diluent container, it may be kinked. 2. One of the Diluent lines has become blocked or kinked. Starrsed RS User Manual Page 191 Version 6.70 MRN-071 (A)-EN...
WI-179 Replace blue air filter (on page 279)) 6. Defective flow sensor board. Needs replacement, fatal error. 7. Outer needle valve is not functioning correctly and vacuum is leaking away, check outer needle valve. Page 192 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Fatal error, technical assistance is needed. 4. Bad vacuum. Adjustment needed on the vacuum. Fatal error, technical assistance is needed. 5. Blockage in the 3 way vacuum manifold. Fatal error, technical assistance is needed. Starrsed RS User Manual Page 193 Version 6.70 MRN-071 (A)-EN...
After each wash cycle, approximately 0.5 ml of disinfectant will be flushed through the wash station. If no disinfectant flows; 1. Unscrew the cap of the disinfectant container to check. 2. One of the disinfectant lines may be blocked or kinked. Page 194 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
If the sample probe has being set too deep, it will touch the bottom of the sample tube. The sample probe then pushes the sample tube slightly downwards, and the aspiration cycle will be aborted. A broken outer needle may cause a similar fault. Starrsed RS User Manual Page 195 Version 6.70 MRN-071 (A)-EN...
O-ring. 3. Check for air in diluter system. 4. Check that the sample probe O-ring is not leaking. 5. Check transparent mixing- piece block for cracks. Page 196 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
1. If just one air bubble is found about 5mm below the meniscus, the filling speed may be too high. 2. The blood column should not exceed the filling height sensor by more than 10mm. One air bubble can result in ESR Error 3. Starrsed RS User Manual Page 197 Version 6.70 MRN-071 (A)-EN...
Check the fill nozzle for damage. If necessary, replace the fill nozzle. Sample tube is leaking on the fill nozzle side. Replace the silicon sample tube Small air bubbles result in ESR Error 3. Page 198 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
3. A scratched valve body. 4. Valve on top of the pipette is dirty or damaged. 5. Check pipette bottom, glass may be chipped. 6. Check the pipette valve for contamination or wear. Starrsed RS User Manual Page 199 Version 6.70 MRN-071 (A)-EN...
First run an extra Fill & Clean sequence, then check after a day's run if haziness is decreased. When there are still many reports, it is recommended to fill the diluter system with a 5% chlorine solution. See WI-178 Hazy problem (on page 278). Page 200 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
<default 90 sec> ETTINGS ENERAL SETTINGS IXING INTERVAL 2. Filling procedure stopped by operator. 3. Insufficient sample volume. Should be at least 1.4 ml. 4. Faulty fill nozzle or fill nozzle washer/O-ring. Starrsed RS User Manual Page 201 Version 6.70 MRN-071 (A)-EN...
INSE POSITION Enter the right value of the pipette number at the rinse station. Close pop-up menu. 2. Select S ET RINSE POSITION HANGE INSE POSITION Press L EARN CAROUSEL POSITION Page 202 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 205
3. Check mechanical connection potentiometers. If the position error is not solved after all the mentioned actions, the positioning device needs replacement. This is a fatal error, call distributor. Starrsed RS User Manual Page 203 Version 6.70 MRN-071 (A)-EN...
Waste pump failure. Exchange the waste pump cassette. If the error returns, call for service. Electrical bridge between the waste-level Clean liquid separator, see Cleaning liquid electrodes. separator (on page 282) Page 204 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
The fill nozzle did not reach the fill position in a certain time limit. Possible reason; • Fill nozzle motor is faulty. • Fill nozzle motor driver is faulty. • Fill nozzle is blocked. • Fatal error, call distributor. Starrsed RS User Manual Page 205 Version 6.70 MRN-071 (A)-EN...
Piercing position sensor was not triggered, check the piercing position sensor Tube 1 not aligned properly, check the tube 1 position setting and save the setting For other error numbers and details See the Appendix - Error list Starrsed RS (on page 236) Page 206 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
15.21. Rack unit errors The rack unit can give errors on the • Gripper module (on page 208) • Mixer module (on page 209) • Mixer module (on page 209) Starrsed RS User Manual Page 207 Version 6.70 MRN-071 (A)-EN...
IN position. Could not send rack to the grabber. Causes and solutions • Check the position motor For more error numbers and details see the Appendix - Error list Starrsed RS (on page 236) Page 208 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Mixer motor send to home position , but position grabber motor is not on the MIX position. • Check the position motor. For error numbers and details see the Appendix - Error list Starrsed RS (on page 236) 15.21.3. Check the mixer The mixer is integrated with the rack grabber and will start mixing as soon the first barcode label is read.
TUBE# "2", the sensor indicator must not light up, “No tube !” is shown next POSITION to the grey block. 8. Repeat this for all applicable tube positions. If tube detection errors are persisting, call for service. Page 210 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
The other codes (command: XX, TSWR: XX and E:XX) are indicators for status/actions from software and I2C connections. Each board has its own range of commands and can not be explained in one way. Starrsed RS User Manual Page 211 Version 6.70 MRN-071 (A)-EN...
Page 214
Power cable not connected on the communication PCB mounted on the back panel. • An I2C cable not connected • Serial cable not connected • No power on one of the PCB's • Short circuit or fault on one of the PCB's Page 212 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Check general ESR data, QC result with ESR error no ESR Result is given see ESR Error (on page 156) • Check sample tube volume • Try new QC sample tube Starrsed RS User Manual Page 213 Version 6.70 MRN-071 (A)-EN...
Page 216
QC, press "No" to return and take appropriate action. General errors Extra information Action • Check barcode Barcode is not accepted Barcode cannot be read Data is incorrect Page 214 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 217
Note: QC Error messages are only shown and stored in QC results and not send to LIMS. QC result is given with the same general errors and warnings as a normal patient ESR-result Starrsed RS User Manual Page 215 Version 6.70 MRN-071 (A)-EN...
Appendix for Starrsed Auto-Compact 16. APPENDIX FOR STARRSED AUTO-COMPACT Appendix section Page 216 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 219
Appendix for Starrsed Auto-Compact Appendix - Article reference list Starrsed RS The Starrsed RS is delivered with a complete accessories kit ESRI 110992. This reference list is for article order numbers only. Part number Description ESRI010246 Pinch valve tube ESRI030903...
Page 220
Spacer for cubitainer When using Plastic container for Deionized water storage: ESRI110106 Deionized water sticker 5l QWFG010902 Container 5L for Deionized water QTLV020004 Cap bottle 5l ESRI110949 Level sensor 5l assembly Page 218 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 221
Appendix for Starrsed Auto-Compact Appendix - Cable connection diagram Starrsed RS User Manual Page 219 Version 6.70 MRN-071 (A)-EN...
Page 222
Appendix for Starrsed Auto-Compact Appendix - Tube connection Starrsed Auto-Compact Page 220 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 223
In this sample, the dilution rate has a dilution failure of 21% and that is printed as EDTA 079. 905007 Sample results with a text error. This sample gives Too many borders found. Result of a pipette possibly filled with air bubbles. Starrsed RS User Manual Page 221 Version 6.70 MRN-071 (A)-EN...
Page 224
Limit error Error One of the following limits are out of the setting range: • ESR Time • Column height • Dilution • Bubbles on top • Hazy aspect • Temperature Page 222 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 225
ESR "ERROR" and "WARNING" code messages. This code appears in the "sample data record" at column 10. Error: No ESR result is given Warning: Result should be reviewed before release. The following codes are defined: Starrsed RS User Manual Page 223 Version 6.70 MRN-071 (A)-EN...
Page 226
Limit error Error One of the following limits are out of the setting range: • ESR Time • Column height • Dilution • Bubbles on top • Hazy aspect • Temperature Page 224 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 227
Diluter settings [S ] > D ETTINGS ILUTER SETTINGS Dilution adjust Automatic Automatic Dilution error detect Auto dilution adjust Dilution flow check General settings [S ] > L ETTINGS IMIT ERROR SETTINGS Starrsed RS User Manual Page 225 Version 6.70 MRN-071 (A)-EN...
Page 228
Position motor depending on Rack selection customer Set tube 1 position 22 mm Actual position 0.0 mm Always read barcode Set serial output comport [S ] > ERVICE SERIAL OUTPUT SETTINGS Page 226 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 229
IMIT ERROR SETTINGS Starrsed Auto-Compact settings Starrsed Auto-Compact Stand alone Via LIMS Client Settings is operational as Send results when time [ No ] 2 Minutes [ Yes ] exceeded Starrsed RS User Manual Page 227 Version 6.70 MRN-071 (A)-EN...
Page 230
Note: When an option is set to YES and this limit error occurs, results will be printed/send to the LIMS. Note: When an option is set to NO and this limit error occurs, results are not printed/send to the LIMS. Page 228 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 231
The cause of this problem can be foam or the waste pump is not working. The Compact will continue to measure and send the ESR results on time to the printer, but the rinse and fill sequences are stopped until the error is solved. Starrsed RS User Manual Page 229 Version 6.70 MRN-071 (A)-EN...
Page 232
Compact was not able to position the measure-unit within a certain time limit. Check for mechanical obstructions. 3. Rinse-unit. Compact was not able to position the rinse-unit within a certain time limit. Check for mechanical obstructions. 4. Fill-nozzle unit. Page 230 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 233
Check in screen M if vacuum is available. AINTENANCE HECK SENSORS Check in screen M the value of flow sensor. AINTENANCE HECK SENSORS Fatal error, call distributor. 5. Position error. Starrsed RS User Manual Page 231 Version 6.70 MRN-071 (A)-EN...
Page 234
9. Separator full error It takes too long for the waste pump to empty the liquid separator. Check separator assembly on air leaks. Replace waste tubes. Exchange waste pump cassette. Clean liquid separator. Page 232 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 235
To set-up the barcode readers to client specifications, use the set-up tools for the barcode readers. Detailed information can be found in Keyence Configuration manual MRN-011 or in Opticon Configuration manual MRN-015. Starrsed RS User Manual Page 233 Version 6.70 MRN-071 (A)-EN...
Page 236
Power cable comes from the power supply. • When the printer is used connect the printer cable onto the 25 pin connector. • The host serial cable is connected to the connecter with the arrow. Page 234 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 237
7. Com1: Compact serial cable comes from the Interface RS232 to IIC. 8. USB connection not used 9. USB connection for touch screen. 10. USB connection not used. 11. Ethernet connection. Starrsed RS User Manual Page 235 Version 6.70 MRN-071 (A)-EN...
Page 238
Program stopped and shut down. Check settings changed settings which were computer needed to be before continuing! not saved to disk are lost. reset. • Computer reset after power failure. Page 236 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 239
(home) home (top) sensor within a • Outer needle motor is certain time limit. faulty. • Outer needle motor driver on needle board is faulty. • Outer needle is blocked. Starrsed RS User Manual Page 237 Version 6.70 MRN-071 (A)-EN...
Page 240
E26: "Diluent Flow" sensor out The EDTA Flow sensor has Check and/or clean the of range. Check/clean this reached a critical level. EDTA Flow sensor. sensor ! Continuing could result in Page 238 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 241
Position settings in Eeprom do Configuration file maybe Settings loaded from Eeprom. not match settings saved to corrupted. Check settings before file. Settings in Eeprom OK sampling! and loaded from Eeprom. Starrsed RS User Manual Page 239 Version 6.70 MRN-071 (A)-EN...
Page 242
T8 or T11 are too low distance). T11. or mixer (gripper) maybe • Mixer is blocked. blocked. • Mixer motor is faulty. • Mixer motor driver on rack board is faulty. Page 240 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 243
Mixer motor is faulty • Mixer motor is blocked. • Mixer motor driver on rack board is faulty. • Check the mixer motor drives the gripper.(Loose belt, loose pulleys ) Starrsed RS User Manual Page 241 Version 6.70 MRN-071 (A)-EN...
Page 244
Mixer motor driver on rack board is faulty. • Check mixer motor belt. E69: Position motor error! Not Gripper did not reach the MIX See E61 on position! (MIX) position. Starting point: IN position. Page 242 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 245
See E61 on position! (IN) (catch) position. Starting point: MIX position. E81: Position motor error! Not Gripper did not reach the MIX See E61 on position! (MIX) position. Starting point: IN Starrsed RS User Manual Page 243 Version 6.70 MRN-071 (A)-EN...
Page 246
Please exit and then restart succeed due to a possible Starrsed RS. If this error fatal hardware error. occurs repeatedly, please note the error code before seeking assistance!
Page 247
Not on position! (FILL) FILL position. Starting point: last sampled tube. E102: Position motor error! Gripper did not reached the See E61 Not on position! (MIX) MIX position. Starting point: FILL position. Starrsed RS User Manual Page 245 Version 6.70 MRN-071 (A)-EN...
Page 248
Faulty Sample probe motor. • Check if needles are blocked. • Faulty motor drivers on the needle board. E116-118 Quality Control Errors See section Quality control trouble shooting (on page 213) Page 246 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 249
Perform End-of-day wash Clean outside aspiration needle Check tubing/diluent syringe Clean outside instrument Clean Fill nozzle Clean Liquid separator Check sensors Replace Fill nozzle O-ring QWLV050004 Replace bacterial air filter (HEPA) QWLV040002 Starrsed RS User Manual Page 247 Version 6.70 MRN-071 (A)-EN...
Page 250
Check Rack transport unit For older models Flat washer Fill nozzle EDTA QWLV060001 Teflon top repair set (Version I diluter 2 rings) ESRI060911 One-way check valve (Version I diluter 2 rings) QWLV010104 Page 248 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 251
Appendix for Starrsed Auto-Compact Appendix - Rack adapters Starrsed RS The Starrsed Auto-Compact uses dedicated Mechatronics rack adapter sets to accommodate one of the following rack types. Use rack adapters with clips when racks are used in which sample tubes are not fixed.
Page 252
Appendix for Starrsed Auto-Compact For manual patient id handling: Beckman Coulter Page 250 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 253
Appendix for Starrsed Auto-Compact Abbott Bayer Starrsed RS User Manual Page 251 Version 6.70 MRN-071 (A)-EN...
Page 254
Appendix for Starrsed Auto-Compact Horiba Horiba ABX Page 252 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 255
ABX closed, Pentra 120 VERA149010 Coulter HmX VERA149011 Horiba VERA149021 Beckman Coulter DXH-800 VERA149022 Roche Rack adapters for other rack types can be developed on request. Please contact support@rrmechatronics.com for details. Starrsed RS User Manual Page 253 Version 6.70 MRN-071 (A)-EN...
Page 256
The right row (JP2) is for the communication port settings and is set on default 4800 baud. To change the port setting, replace the jumper to the required setting you want. Page 254 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 257
Input Transmit Data (a.k.a TxD, Tx). Sending data from DTE. (RXD) Output Request To Send (ASYNC) (RTS) Input Clear To Send (ASYNC) (CTS) Input Data Set Ready (ASYNC) Signal Ground (SG) Input Carrier Detect (a.k.a DCD). Starrsed RS User Manual Page 255 Version 6.70 MRN-071 (A)-EN...
Page 258
Interface connections to PC Example connection to P.C. Example to HOST computer Example connection to P.C. without handshake system including handshake including hardware handshake Hardware no handshaking for DB 25 connector Page 256 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 260
SD (TXD) -> RD (RXD) GND (SG) 6 + 8 DSR + CD 6 + 8 DSR + CD CS (CTS) ->RS (RTS) RS (RTS) -> CS (CTS) Hardware handshaking for DB 25 connectors Page 258 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 261
500 milli-seconds. If the data transmission is interrupted for more than 500 milliseconds by the host computer then the instrument will stop sending data. This data is then lost to the host computer. Starrsed RS User Manual Page 259 Version 6.70 MRN-071 (A)-EN...
Page 262
To overcome this problem and still be able to use a cheap null modem communication cable with only three lines in it, a fake null modem cable layout has been defined. The null modem cable with loop back handshaking resulted from this. Page 260 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 263
All specified RS232 signals are completely isolated from the 'computer earth' and the 'mains supply'. Handshake: Hardware handshake is not applied. Change the data transmission parameters: From the transmission parameters, only the baud rate can be selected from 1200 up to 9600. Starrsed RS User Manual Page 261 Version 6.70 MRN-071 (A)-EN...
Page 264
To overcome this problem and still be able to use a cheap null modem communication cable with only three lines in it, a fake null modem cable layout has been defined. The null modem cable with loop back handshaking resulted from this. Page 262 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 265
All specified RS232 signals are completely isolated from the 'computer earth' and the 'mains supply'. Handshake: Hardware handshake is not applied. Change the data transmission parameters: From the transmission parameters, only the baud rate can be selected from 1200 up to 9600. Starrsed RS User Manual Page 263 Version 6.70 MRN-071 (A)-EN...
Page 266
Input Clear To Send. Raised by DCE in response to RTS from DTE Input Ring Indicator. Set when incoming ring detected - used for auto- answer application. DTE raised DTR to answer. Female side Male side Pin no Pin no Page 264 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 267
In case of preparation of a cable the following connections must be made. Female side Male side Pin no Pin no If you have no result you can try changing the following pin wires. Female side Male side Pin no Pin no Starrsed RS User Manual Page 265 Version 6.70 MRN-071 (A)-EN...
Page 268
Not in use In-0 Motor sensor Waste ERROR Measure. motor tacho In-2 Check position. 51-111 Rinse head top Measure. head top In-3 Check 10 pos Rinse head bottom Measure. head bottom Page 266 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 269
000C - FFFF Board address Selector Picture 2 Current Jumper Picture 1 Forward reverse Selector Fusses In mA In mA In mA Fuse A 1600 Fuse B 1600 1600 Fuse C Starrsed RS User Manual Page 267 Version 6.70 MRN-071 (A)-EN...
Work instruction Starrsed Auto-Compact 17. WORK INSTRUCTION STARRSED AUTO-COMPACT Work instruction section Page 268 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 271
6. Place one end of the tube in the pump plate holder. 7. Pull the new tube over the peristaltic pump rotor. 8. Pull pump tube slightly downwards and at the same time towards the back of the Starrsed Auto-Compact. Starrsed RS User Manual Page 269 Version 6.70 MRN-071 (A)-EN...
Page 272
6. Place one end of the tube in the pump plate holder. 7. Pull the new tube over the peristaltic pump rotor. 8. Pull pump tube slightly downwards and at the same time towards the back of the Starrsed Auto-Compact. Starrsed RS User Manual Page 270 Version 6.70 MRN-071 (A)-EN...
Page 273
11. Release the stainless steel vacuum tube. 12. Replace the waste container. 13. Close left cover. Prime Saline 1. Select P . Repeat the prime saline until the liquid separator is empty. RIME ALINE Starrsed RS User Manual Page 271 Version 6.70 MRN-071 (A)-EN...
Page 274
6. Insert new waste pump cassette ESRI090921 until it clicks into place. 7. Remove the protection caps on from the tubes. 8. Connect the two tubes to new waste pump cassette. Page 272 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 275
6. Check for leakage with Fill & Clean. Replace the top cover: 1. Put the cover carefully over the instrument. 2. Fasten the two rear screws of the top cover. (If present/if needed). Starrsed RS User Manual Page 273 Version 6.70 MRN-071 (A)-EN...
Page 276
MS 40..50..60 by using the C HECK function. MEASURE SENSOR 3. If not in range repeat cleaning the inner part of the measuring sensor. 4. When in range switch OFF the Starrsed Auto-Compact. Starrsed RS User Manual Page 274 Version 6.70 MRN-071 (A)-EN...
Page 277
Replace the top cover: 1. Put the cover carefully over the instrument. 2. Fasten the two rear screws of the top cover. (If present/if needed). Switch ON the Starrsed Auto-Compact. Starrsed RS User Manual Page 275 Version 6.70 MRN-071 (A)-EN...
Page 278
Remove the valve body and check the silicon tube position, it must be exactly centred. 8. Fit the bottom tube clamp and V-seal ring. 9. Check the position of the valve. If incorrect, disassemble pipette valve and tube and re-assemble again. Starrsed RS User Manual Page 276 Version 6.70 MRN-071 (A)-EN...
Page 279
6. Check for leakage with Fill & Clean. Replace the top cover: 1. Put the cover carefully over the instrument. 2. Fasten the two rear screws of the top cover. (If present/if needed). Starrsed RS User Manual Page 277 Version 6.70 MRN-071 (A)-EN...
Page 280
1. Prepare a Fill and Clean arrangement. 2. Run the fill and clean sequence. When all the pipettes are filled the needle goes back to the home position. 3. Remove the Fill and clean arrangement. Starrsed RS User Manual Page 278 Version 6.70 MRN-071 (A)-EN...
Page 281
3. Reconnect the tube connectors on the filter Replace the top cover: 1. Put the cover carefully over the instrument. 2. Fasten the two rear screws of the top cover. (If present/if needed). Starrsed RS User Manual Page 279 Version 6.70 MRN-071 (A)-EN...
Page 282
3. Calculate the mean of the 10 samples. 4. Make dilution adjustment if necessary in S > D ETTINGS ILUTER SETTINGS 5. Go to tab S and select D > G ETTINGS ENERAL SETTINGS ISPLAY DILUTION Starrsed RS User Manual Page 280 Version 6.70 MRN-071 (A)-EN...
Page 283
3. Refill the diluent bottle with new diluent solution. 4. Perform another [P RIME 5. After the prime sequence stops press [P ] again to fill the dispenser system with the new RIME diluent solution. Starrsed RS User Manual Page 281 Version 6.70 MRN-071 (A)-EN...
Page 284
C. A little silicon grease on the rim of separator will make the assembling and adjustment easier. D. If applicable replace the bacterial HEPA filter (For Maintenance Level 4: Exchange filter QWLV040002) Starrsed RS User Manual Page 282 Version 6.70 MRN-071 (A)-EN...
Page 285
The following items need to be replaced annually: 1. Outer needle and sample probe (if needed) 2. Pipette valve bodies (84 pieces) QTST040001. Check: 1. Adjustment of fill nozzle and rinse nozzle. Starrsed RS User Manual Page 283 Version 6.70 MRN-071 (A)-EN...
Page 286
Safety: Bio Hazard area Instrument: Compact Revision: 001, February 2005 Instructions for disassembling and assembling the fill nozzle Note: For the O-ring replacing, only take the top part away from the fill nozzle. Starrsed RS User Manual Page 284 Version 6.70 MRN-071 (A)-EN...
Page 287
The use of a toothbrush and detergent is recommended. 1. Carefully scrub the fill nozzle inner part. 2. Use a tissue to dry the fill nozzle. Starrsed RS User Manual Page 285 Version 6.70 MRN-071 (A)-EN...
Page 288
1. Carefully scrub the fill nozzle inner part. 2. Use a tissue to dry the fill nozzle. Disassemble fill nozzle holder: 1. Turn the holder to the right. 2. The holder can now be removed Starrsed RS User Manual Page 286 Version 6.70 MRN-071 (A)-EN...
Page 289
1. Connect the silicon tube to the fill nozzle. 2. Put the fill nozzle into the holder. 3. Push the fill nozzle upwards and turn the holder to the left. Starrsed RS User Manual Page 287 Version 6.70 MRN-071 (A)-EN...
Page 290
Work instruction Number 205 Page 1 of 1 Purpose: Replace the pinch valve tube ESRI010246 Safety: Bio Hazard area Instrument: Compact Revision: 001, February 2005 Replace the pinch valve tube ESRI010246 Starrsed RS User Manual Page 288 Version 6.70 MRN-071 (A)-EN...
Page 291
12. Check sensors and vacuum pressure settings 13. Check instrument settings 14. Check communication interface settings 15. Perform aspiration with test blood before normal operation, check if all steps are performed without errors Starrsed RS User Manual Page 289 Version 6.70 MRN-071 (A)-EN...
Page 292
1. Remove inner- and outer needle assembly and dispose of as biohazardous waste 2. Remove filter from separator and (if applicable) waste container, dispose of as biohazardous waste 3. Discard instrument according local regulations Starrsed RS User Manual Page 290 Version 6.70 MRN-071 (A)-EN...
Page 293
2. Run 10 samples through the instrument and make a note of the dilution rate. 3. Calculate the mean value of the 10 samples. 4. Make adjustment if necessary in [S ] -> [D ] -> [D ETTINGS ILUTER SETTINGS ILUTION ADJUSTMENT Starrsed RS User Manual Page 291 Version 6.70 MRN-071 (A)-EN...
Page 294
], click button [P 68)] AINTENANCE RIME LEAN ON PAGE and perform the [P ] function. RIME ILUENT ILUTER 7. Wipe outer surface and stainless steel plate below the pipettes with disinfectant. Starrsed RS User Manual Page 292 Version 6.70 MRN-071 (A)-EN...
Page 295
2. The fill-nozzle can now be removed. removed. 3. Disconnect the silicon tube 3. Disconnect the silicon tube from the fill nozzle. from the fill nozzle. 2. Clean Liquid separator Removing Starrsed RS User Manual Page 293 Version 6.70 MRN-071 (A)-EN...
Page 296
C. A little silicon grease on the rim of separator will make the assembling and adjustment easier. D. If applicable replace the bacterial HEPA filter (For Maintenance Level 4: Exchange filter QWLV040002) Page 294 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 297
], click button [P 68)] AINTENANCE RIME LEAN ON PAGE and perform the [P ] function. RIME ILUENT ILUTER 7. Wipe outer surface and stainless steel plate below the pipettes with disinfectant. Starrsed RS User Manual Page 295 Version 6.70 MRN-071 (A)-EN...
Page 298
Press test. When test is finished, signal Down and signal Up must be green. Separator check • Go to tab M -> C . Select C box. AINTENANCE HECK SENSOR HECK EPARATOR SENSOR Separator sensor <200 600 >700 Starrsed RS User Manual Page 296 Version 6.70 MRN-071 (A)-EN...
Page 299
8. Replace the correct tubes on the outer needle. 9. Fasten the outer needle bolt.(Do not over-tighten the screw) 10. Replace the cover on the Rack unit. Starrsed RS User Manual Page 297 Version 6.70 MRN-071 (A)-EN...
Page 300
4. When all the pipettes are filled, the needle goes back to the home position and the adapter is moved to the release position and will be transported into the End-pool. Starrsed RS User Manual Page 298 Version 6.70 MRN-071 (A)-EN...
Page 301
The use of a toothbrush and detergent is recommended. 1. Carefully scrub the fill nozzle inner part. 2. Use a tissue to dry the fill nozzle. Replace O-ring: Remove the O-ring. (QWLV050004) Install new O-ring. (QWLV050004) Starrsed RS User Manual Page 299 Version 6.70 MRN-071 (A)-EN...
Page 302
2. Clean Liquid Separator and exchange filters Removing Cleaning A. Clean all parts with hot water and a brush. B. Use some acid free vaseline on the screw-thread of the glass jar. Page 300 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 303
6. Place one end of the tube in the pump plate holder. 7. Pull the new tube over the peristaltic pump rotor. 8. Pull pump tube slightly downwards and at the same time towards the back of the Starrsed Auto-Compact. Starrsed RS User Manual Page 301 Version 6.70 MRN-071 (A)-EN...
Page 304
ILUTER START SENSOR Diluter start sensor 400-700 Measure sensor check • Go to tab M -> C . Select C box. AINTENANCE HECK SENSOR HECK MEASURE SENSOR Measure sensor MS 40..50..60 Page 302 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 305
], click button [P 68)] AINTENANCE RIME LEAN ON PAGE and perform the [P ] function. RIME ILUENT ILUTER 7. Wipe outer surface and stainless steel plate below the pipettes with disinfectant. Starrsed RS User Manual Page 303 Version 6.70 MRN-071 (A)-EN...
Page 306
The use of a toothbrush and detergent is recommended. 1. Carefully scrub the fill nozzle inner part. 2. Use a tissue to dry the fill nozzle. Replace O-ring: Remove the O-ring. (QWLV050004) Install new O-ring. (QWLV050004) Starrsed RS User Manual Page 304 Version 6.70 MRN-071 (A)-EN...
Page 307
6. Place one end of the tube in the pump plate holder. 7. Pull the new tube over the peristaltic pump rotor. 8. Pull pump tube slightly downwards and at the same time towards the back of the Starrsed Auto-Compact. Starrsed RS User Manual Page 305 Version 6.70 MRN-071 (A)-EN...
Page 308
4. When all the pipettes are filled, the needle goes back to the home position and the adapter is moved to the release position and will be transported into the End-pool. Page 306 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 309
Be careful, as there may be blood in the cassette. First, make up some disinfectant and put this in the liquid separator. Press P to pump disinfectant through the pump cassette. RIME ISINFECTANT Starrsed RS User Manual Page 307 Version 6.70 MRN-071 (A)-EN...
Page 310
7. Remove the protection caps on from the tubes. 8. Connect the two tubes to new waste pump cassette. 7. Exchange Pinch valve tube Replace the pinch valve tube ESRI010246 Page 308 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 311
7. Wipe outer surface and stainless steel plate below the pipettes with disinfectant. 10. Exchange air filter Air filter replacement QWLV040003 1. Pull both tube connectors out of the blue filter. 2. Place new blue filter Starrsed RS User Manual Page 309 Version 6.70 MRN-071 (A)-EN...
Page 312
3. Reconnect the tube connectors on the filter If applicable: Check your on-board diluent container Recommendation is to clean the bottle when you replace the 5 liter container Diluent. Page 310 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 313
The use of a toothbrush and detergent is recommended. 1. Carefully scrub the fill nozzle inner part. 2. Use a tissue to dry the fill nozzle. Replace O-ring: Remove the O-ring. (QWLV050004) Install new O-ring. (QWLV050004) Starrsed RS User Manual Page 311 Version 6.70 MRN-071 (A)-EN...
Page 314
6. Place one end of the tube in the pump plate holder. 7. Pull the new tube over the peristaltic pump rotor. 8. Pull pump tube slightly downwards and at the same time towards the back of the Starrsed Auto-Compact. Page 312 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 315
ILUTER START SENSOR Diluter start sensor 400-700 Measure sensor check • Go to tab M -> C . Select C box. AINTENANCE HECK SENSOR HECK MEASURE SENSOR Measure sensor MS 40..50..60 Starrsed RS User Manual Page 313 Version 6.70 MRN-071 (A)-EN...
Page 316
10. Push the liquid separator towards the rear, with the sensor connectors in the holes. 11. Release the stainless steel vacuum tube. 12. Replace the waste container. 13. Close left cover. Page 314 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 317
Fill Stop sensor check • Go to tab M -> C . Select C box. AINTENANCE HECK SENSOR HECK ILL STOP SENSOR Fill stop sensor FS 90..140..165 Diluter Start sensor check Starrsed RS User Manual Page 315 Version 6.70 MRN-071 (A)-EN...
Page 318
10. Push the liquid separator towards the rear, with the sensor connectors in the holes. 11. Release the stainless steel vacuum tube. 12. Replace the waste container. 13. Close left cover. Page 316 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 319
7. Remove the protection caps on from the tubes. 8. Connect the two tubes to new waste pump cassette. 7. Exchange Pinch valve tube Replace the pinch valve tube ESRI010246 10. Replace diluter syringe tip Replace diluter tip Starrsed RS User Manual Page 317 Version 6.70 MRN-071 (A)-EN...
Page 320
D ETTINGS ENERAL SETTINGS ISPLAY DILUTION If applicable: 6. Check your on-board diluent container Recommendation is to clean the bottle when you replace the 5 liter container Diluent. Page 318 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...
Page 321
8. Check if the information is the same as on the barcode label. If also the barcode could not be detected, the tube position is logged as empty. In that case the barcode reader itself must be adjusted. Starrsed RS User Manual Page 319 Version 6.70 MRN-071 (A)-EN...
18.3°C. instrument. 18.1.1.1.4. ESR is short for Erythrocyte Sedimentation Rate. It is the amount of sedimentation (setting) of erythrocytes (red blood cells) in a blood column during a specified time. Starrsed RS User Manual Page 321 Version 6.70 MRN-071 (A)-EN...
Page 324
Starrsed Auto-Compact to the HOST. Only sample results and result related messages are send. 18.1.1.1.12. WI is short for Work Instruction and is used with an index number for a range of work instructions. Page 322 Starrsed RS User Manual Version 6.70 MRN-071 (A)-EN...