Available languages

7.2.4. measurement module ECP1

width of the module itself
connector for measurement
range
sensitivity
output
power supply
resolution
temperature range
humidity range
sensor diameter
weight (sensor + cable)
cable length
min. diameter measure object
temperature stability electronic
long time stability
european standards
protection standard (electronic
and sensor)
linearity
repeatability
8.

operation

8.1.

initiation

First check that your system uses the correct supply voltage (see the label on the
backside of the casing). Please connect the device with the wall outlet by using
the power cord. The piezo-actuators use operation voltages up to +150Volt.
Please be careful for any electric hazards. Do not remove the actuator plug until
the amplifier has been switched off. Piezoelectric actuators such stacks or various
tables work electrically as a capacitance. These elements are able to store
electrical energy over a long period (up to days) and the stored energy may be
dangerous.
When using the ENV 40 CSG/CCP please be sure to notice that before switching
on the system, that the closed-loop switch is set to the "off" position.
Turn all potentiometers to maximal counter clockwise position. Remove external
modulation signals from MOD plug. Make sure all cable from the actuators (OUT
and SENSOR) are plugged into the correct amplifier.
Geschäftsführer: Dr. Bernt Götz • Telefon 03641/66880 • Fax 03641/668866 • www.piezosystem.com
46
6TE = 30,5mm
LEMO 0S.650
200µm or 500µm
100mV/µm or rather 40mV/µm
10V
15VDC / 50mA
0.005% FS
sensor: -50 ... +150 °C
cable: -50 ... +150 °C
electronic: +10 ... +40 °C
5 ... 95% (not condensing)
10mm
56g
1.6m
12mm
 0.005% FS / K
 0.04% FS / Month
EN 50081-1, EN 50082-2
IP 40
typ. < 0.05 % FS
typ. < 0.04 % FS
table 13: technical data measurement module ECP1