 By maintenance please notice the information in
 If the balance seems not to operate properly, unplug it from the mains and do not use until checked by
authorized service.
 All repairs and necessary regulations can be made by authorized personnel only.
COUTION!
:
Carefully read this operation manual before setup and commissioning.

5 Transportation and Storage

5.1 Acceptance check

When receiving the appliance, please check packaging immediately, and the appliance itself when
unpacking for possible visible damage.

5.2 Packing

 Keep all parts of the original packaging for a possibly required return
 Only use original packaging for returning.
 Prior to dispatch disconnect all cables and remove loose/mobile parts.
 Reattach possibly supplied transport securing devices.
 Secure all parts such as the glass wind screen, the weighing platform, power unit etc.
against shifting and damage.

6 Unpacking, Setup and Commissioning

6.1 Installation Site, Location of Use

The balances are designed in a way that reliable weighing results are achieved in common conditions of
use. You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
 Major weighing deviations (incorrect weighing results) may be experienced should electro-
magnetic fields (e.g. due to mobile phones or radio equipment), static electricity accumulations
or instable power supply occur. Change location or remove source of interference.
 Place the balance on a firm, level surface;
 Avoid extreme heat as well as temperature fluctuation caused by installing next to a radiator or in the
direct sunlight;
 Protect the balance against direct draughts due to open windows and doors;
 Avoid jarring during weighing;
 Protect the balance against high humidity, vapors and dust;
 Do not expose the device to extreme dampness for longer periods of time.
 Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold
appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected
appliance for ca. 4 hours at room temperature.
 Avoid static charge of goods to be weighed or weighing container.
Chapter "Maintenance and
G&G GmbH | Transportation and Storage
Servicing".
8