13 Configuration & Explain of the function setting
The settings of the balance can be changed in the setting menu. This way, the balance can be adjusted to
individual weighing requirements. By the factory the user menu has been set in a way that normally no
more changes must be made, only at special conditions of use.
Overview:
In the off state press the [CAL] button and keep it pressed, press the [ ] button turn on the balance. Now
the balance is accessed to the setting menu. Using the [CAL] button the individual menu items (C1-C4) can
be selected one after the other. Press the [TARE] button to change the setting value. After all points have
been configured, press [CAL] button returns the balance automatically into weighing mode.
a) C1 = sensitivity setting - values: 0 1 2 3 4
The higher the value, the lower the sensitivity and higher stability.
Filter 0: Setting for dispensing. Scale reacts very fast but also is very vulnerable to bad environment
conditions.
Filter 1/2: Sensitive and fast, very quiet using location.
Filter 3/4: Robust but slow, busy using location.
b) C2 = filtering strength setting - values: 0 1 2 3
The higher the value, the lower the sensitivity and higher stability.
Filter 0: Setting for dispensing. Filter is deactivated, scale reacts very sensitive but also is very
vulnerable to bad environment conditions.
Filter 1/2: Sensitive and fast, very quiet using location.
Filter 3: Robust but slow, busy using location.
c) C3 = serial port setting - transmission speed (RS-232)
2(600), 3(1200), 4(2400), 5(4800), 6(9600) - Default setting is 2.
d) C4 = device number / communication setting (RS-232)
Press [TARE] button changes the value smaller. Press [
number corresponds to the first data signal to the command control of the computer.
Default setting is 27 (1B
e) C5 = setting the mode for the backlight
0(always on), 1(automatic after the weight has changed), 2(usually off) - Default setting is 0.
f) C6 = setting the internal calibration function
0(Internal calibration function on),
1(Internal calibration function off,press the "CAL" key the internal calibration will not work),
2(change the "CAL" key as the external calibration key,press "CAL" key the balance begin the external
calibration mode) –
Default setting is 0.
).
hex
G&G GmbH | Configuration & Explain of the function setting
] button changes the value bigger. The
16