Hameg HMF2525 Manual page 37

25/50 mh z arbitrary function generator
Hide thumbs Also See for HMF2525:
Table of Contents

Advertisement

Available languages

Available languages

the information about the languages available on the stick. With
the soft menu, languages may be added, removed or updated.
Please note the format of the date (YYYY-MM-DD) according to
the multi language norm of ISO 8601.
Attention!
At the time of the update the unit will not respond on
any inputs and the display will be resettet. Does not
switch off the unit during the update process. A inter-
ruption of power supply can destroy the unit!
With the soft key EXIT you can close the update menu.
5.4.2
Interface settings
This menu item is used for the settings of the diverse interfaces:
1. The dual interface HO720 USB/RS-232 (Baud rate, number
of stop bits, parity, handshake on/off)
2. LAN interface HO730 (IP address, sub net mask etc., see
the HO730 manual)
3. IEEE-488 GPIB interface HO740 (GPIB address).
The interface desired for the communication can be selected
with the respective soft menu key. Use the soft menu item
PARAMETER to set the necessary interface parameters. More
information about the selected interface you can find on www.
hameg.com. With the left arrow button you can jump back to
the last menu.
5.4.3
Storing / Recalling of instrument settings
(SAVE/RECALL)
The HMF series can store two different kinds of data:
– Device settings
– Screenshots
Screenshots can only be stored on USB sticks. Device settings
can be stored either on a USB stick or in the instrument's non-
volatile memories.
Device Settings
Push the key SAVE/RECALL for calling the main menu for sto-
rage and load functions. First a listing is shown of the kinds of
data which can be stored and loaded. By pushing the key next
to the top menu item DEVICE SETTINGS this menu will open.
Fig. 5.5: Updating menu
and information window
Fig. 5.6: Basic menu for
instrument settings
E x t e n d e d o p e r a t i n g m o d e s
In this menu, by pushing the respective key, it is possible to
call the menu for storing, the data manager for loading, and
the menu for exporting and importing instrument settings.
Additionally, the menu item DEFAULT SETTINGS will reset the
instrument to the factory settings. The storing menu is opened
by pushing the SAVE key.
Here the storage location (internal memory or front panel USB)
is selected, also a name and a commentary can be added; these
will be stored by pushing the soft menu key next to SAVE. In
order to recall stored instrument settings, call the main instru-
ment settings menu and select LOAD by pushing the respective
soft menu key. The data manager will open, use the menu keys
and the universal knob for navigating.
Here the location is selected from which the settings data
are to be loaded. After the selection in the data manager,
load the settings by pushing the soft menu key LOAD. The
data manager also allows you to erase individual settings
in the internal memory. If a USB stick is plugged in and has
been selected as the location, it is also possible to change or
erase directories. With the left arrow button you can jump
back to the last menu.
Screenshots
The most important method of storing for documentation
purposes is the screen photo. At least one USB stick must be
connected, only then will any settings regarding the destination
and the format be possible. Push the keys SAVE/RECALL and
SCREENSHOTS for opening the appropriate menu.
The second menu item FILE NAME allows you to enter a name
with the respective name entry menu which will open automati-
cally upon selecting this menu item. If FORMAT is selected with
the respective menu key, these formats will be offered and can
be selected with the universal knob:
Fig. 5.7: Saving
instrument settings
Fig. 5.8: Loading
instrument settings
Fig. 5.9: Menu for
screenshots
Subject to change without notice
37