Advertisement

Quick Links

20000005492
HS/KS_260_b/c_052017
HS/KS 260 basic
IKA
®
HS/KS 260 control
IKA
®
KS 260 basic
HS 260 control

Advertisement

Table of Contents
loading

Summary of Contents for IKA HS 260 basic

  • Page 1 20000005492 HS/KS_260_b/c_052017 HS/KS 260 basic ® HS/KS 260 control ® KS 260 basic HS 260 control...
  • Page 2: Table Of Contents

    repairs. The machine is to be unplugged from the mains before opening. Contens Live parts inside the machine may still be live for some time after unplug- ging from the mains. Page NOTE ! Covering or parts that are capable of being removed Declaration of conformity from the unit without accessory equipment have to be reat- Guarantee...
  • Page 3: Proper Use

    The original IKA ® cable layout must be restored following servicing! If the target value of the speed is changed too rapidly, the machine (HS/KS 260 basic) will turn itself off automatically.
  • Page 4: Useful Facts

    device is ready to operate after plugging in the mains plug. If these Useful facts procedures are not followed, safe operation cannot be guaranteed and/or the equipment may be damaged. With the purchase of this device, you have acquired a high-quality Observe the ambient conditions (temperature, humidity, etc.) listed product.
  • Page 5: Display

    Description of the LCD-Display - HS/KS 260 control Display Real value Rated value Description of the LED-Display - HS/KS 260 basic Display element for the setting speed Mot 1/min 50 100 150 200 250 300 350 400 450 500 On Off Time min Display element for Continuous...
  • Page 6 back on, both functions must be started again. The upper speed limit d.) If the upper speed limit is fixed, the unit 1/min in the left display can only be set or changed in operating mode A. You can recognice changes to OK.
  • Page 7 the display for operating mode B along with * in the right display. d.) The new speed target value has been set or just stored in the left After the power has been turned off,the device will no longer auto- display (for example 130) is the shown in the unit 1/min and flashing matically start up in operating mode B by itself.
  • Page 8 a.) All display elements in the digital displays are activated for about 2 on the display. In addition, the text is shown on the display seconds. instead of SV . The actual value of the time setting must be understood from this point onward as the time remaining for the agitation process.
  • Page 9: Setting The Agitation Time

    ATTENTION! After a power failure, the unlocked Note: If the time expiration signal has been activated, expiration of the (freely movable) agitation table will continue to perform set time is indicated by a sequence of signals (3 times 5 short tons at a circular motion until the lock function can fasten the intervals of about 10 seconds).
  • Page 10: Setting The Speed

    Control model tion or for several agitation processes in timed mode. 3 4 5 9 5 9 . 9 5 9 : The amount 3 4 5 NOTE: After the power has failed, the set time is erased and the of time requi- device switches into Offline operating mode (standstill).
  • Page 11: Attachments

    410 x 33 x 370 ® Attachments recommended by IKA must be connected to the agita- ATTENTION! The dish attachment is secured in place by four lateral tion table to ensure safe and secure operation. in addition , the usable clamping screws on the agitation table.
  • Page 12: Error Codes

    AS 260.1 Universal attachment Error code Error Cause of the error AS 260.5 Separating funnel attachment Off - LED (red) The potentiometer for -Too rapid a change in the the speed is not assig- speed target value 1x flash / ning any target value -Internal device error 2 sec pause...
  • Page 13: Interface And Outputs

    Control error codes: Interface and outputs ATTENTION! If an error code is encountered, an audio warning signal is generated and the red Off signal light goes on in the display (HS/KS 260 control version only) field. First, please try to continue operating by turning the device off The HS/KS 260 control is equipped with a 15-pin SUB-D connector on the rear and back on again.
  • Page 14 NAMUR Instructions Function Display • RTS: (Pin 7) LOW (positive tension) additional / PC may send • RTS: (Pin 7) HIGH (negative tension) IN_PV_X Reading the real-value / PC may not send OUT_SP_Xn Setting the actual value to n • CTS: (Pin 8) LOW (positive tension) up to a maximum of the set / PC recipient...
  • Page 15: Communication Between Laboratory Instrument And Pc

    Ending of ramp loop IN_TYPE Requiring identification of (HS/KS 260 control version only) laboratory instrument ® PC 5.1 Adapter IKA - Control IN_NAME Requiring the designation name The Adaptor PC 5.1 splits the analog and serial signals. The analog OUT_NAME...
  • Page 16: Motor Protection / Safety Device

    (25-pin plug). only cleansing agents which have been appro- ved by IKA ® to clean IKA ® devices. To remove use: AK 2.1 Cable Dyes isopropyl alcohol This cable is required to connect the 7-pin connector to a chart recor-...
  • Page 17: Accessories

    AS 260.5 Separating funnel attachment 8,8 (control) STICKMAX Adhesive mat HS/KS 260 basic (HS/KS 260 control version only) Speed setting: Analog (infinitely variable) Adapter IKA ® - Control PC 5.1 Rotary knob on front side left PC 1.5 Cable Speed display: 10 LED’s with scale-in...