Electromagnetic Emissions - Tivic Health ClearUP User Manual

Sinus pain relief
Hide thumbs Also See for ClearUP:
Table of Contents

Advertisement

Electromagnetic Compatibility
(EMC) Information
Guidance and manufacturer's declaration-

Electromagnetic emissions

The ClearUP™ is intended for use in the electromagnetic
environment specified below. The customer or the user of the
ClearUP™ should assure that it is used in such an environment.
Emissions
Compliance
test
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic
Compliant
emissions
IEC 61000-3-2
Voltage
Compliant
fluctuations /
flicker emissions
IEC 61000-3-3
22
Electromagnetic
environment -
guidance
RF emissions are low
compared to the FCC
Group 1 limits.
RF emissions are low
compared to the FCC
Class B limits.
Electromagnetic Compatibility
(EMC) Information
Guidance and manufacturer's declaration-
Electromagnetic immunity
The ClearUP™ is intended for use in the electromagnetic
environment specified below. The customer or the user of the
ClearUP™ should assure that it is used in such an environment.
Immunity
Compliance
Compliance
test
level
Electrostatic
+/- 6 kV
discharge
contact
(ESD) IEC
+/-8 kV air
+/-8 kV air
61000 - 4-2
Electrical
+/-2 kV for
+/-2 kV for
power supply
power supply
fast
lines +/-1kV
lines +/-1kV
transient/
for input/
for input/
burst IEC
output lines
output lines
61000-4-4
Surge IEC
+/-1 kV
+/-1 kV
differential
differential
61000-4-5
mode +/-2kV
mode +/-2kV
common
common
mode
mode
Voltage dips.
<5% Ut (>95%
<5% Ut (>95%
dip in Ut) for
dip in Ut) for
Short
0.5 cycle
0.5 cycle
Interruptions
and voltage
40% Ut (60%
40% Ut (60%
variations
dip in Ut) for 5
dip in Ut) for 5
on power
cycles
cycles
supply input
70% Ut (30%
70% Ut (30%
lines IEC
dip in Ut) for 25
dip in Ut) for 25
61000-4-1
cycles
cycles
<5% Ut (>95%
<5% Ut (>95%
dip in Ut) for
dip in Ut) for
5 sec
5 sec
Power
frequency
(50/60 Hz)
30 A/m
magnetic
field IEC
61000-4-8
Conducted
3 Vrms
RF IEC
150kHz to 80
150kHz to 80
MHz
61000-4-6
Radiated RF
3 V/m 80MHz
3 V/m 80MHz
IEC 61000-
to 2.5GHz
to 2.5GHz
4-3
Electromagnetic
level
environment -
guidance
+/- 6 kV
Did not show
contact
sensitivity to test
levels.
Did not show
sensitivity to test
levels.
Did not show
sensitivity to test
levels.
AC use only in
charging mode.
Device did not show
30 A/m
sensitivity to the test
levels.
3 Vrms
Separation distances
do not apply,
MHz
because there is no
RF emitter in the
device. Device uses
haptic technology to
generate the vibration.
23

Advertisement

Table of Contents
loading

Table of Contents