 Replace the rack and plug in the PRESSURE module into the new rack. If the
LED on the module does not light up, replace the rack interface card in the
function box.
If a " LQVWHDG RI D QXPHULF and one or more INOP messages are displayed for the
PRESSURE module, perform the following actions:
3OHDVH QRWH WKDW IRU SURFHHGLQJ WR WKH QH[W VWHSV D WUDQVGXFHU PXVW EH FRQQHFWHG WR WKH
PRGXOH
6WHS  Perform a module self-test (See "Performing the M1006A/B Pressure Module Self-
Test" on page 5-42). If the PRESSURE test numerics read exactly the listed values,
proceed to the next step, if not, replace the module.
6WHS  Refer to the following table. Correct the possible cause identified by the INOP
message and verify that the INOP disappears. If the problem still exists, replace the
module.
7DEOH  ,123 0HVVDJHV IRU 3UHVVXUH
,123 0HVVDJH
<Px> UNPLUGGED
<Px> NO TRANSDUCER
<Px> EQUIP MALF
<Px> ZERO + CHECK CAL Occurs when new trans-
<Px> OVERRANGE
CPP CHECK SOURCES
CPP CHECK UNITS
$JLOHQW 0$ 99 6HULHV 6HUYLFH *XLGH
3RVVLEOH &DXVH
Pressure module is
Perform Step 6 on page 5-10.
unplugged from the rack.
No transducer connected to
Connect transducer to module.
the module.
Malfunction in the pressure
Replace module.
hardware.
Zero the PRESSURE module.
ducer attached to module, or
when the power is off for
longer than one minute and
settings transfer is off.
Measured pressure >361
Replace transducer.
mmHg, <-41 mmHg.
If the INOP persists, replace module.
Occurs for one minute if
Connect module properly to rack and
either the arterial or the
make sure that the PRESSURE parame-
intracranial pressure sources
ter is turned on.
are switched off, or the front
Make sure you get valid numerics for
end modules are removed.
ICP and the other pressure. Make sure
that the preselected pressure label in the
CPP source window is the same as for
the second pressure.
Occurs if two pressure
Use the same unit for all PRESSURE
sources have different units.
modules.
Troubleshooting Plug-in Modules 5-11
&RUUHFWLYH $FWLRQ