Download Print this page
Keysight Technologies U1252B Modification Recommended Service Note

Keysight Technologies U1252B Modification Recommended Service Note

Handheld digit multimeter
Hide thumbs Also See for U1252B:

Advertisement

Quick Links

Modification Recommended
Service Note
U1252B Handheld Digit Multimeter
Serial Numbers: "See Appendix (A)"
The Problem – Capacitance measurement may be out of specification at 100nF range and below.
Parts Required:
P/N
Description
Qty.
Pomona 2BC-36 double banana plug both ends cable assembly
ADMINISTRATIVE INFORMATION
ACTION
[X] ON SPECIFIED FAILURE
CATEGORY:
[[]] AGREEABLE TIME
LOCATION
[X] CUSTOMER INSTALLABLE
CATEGORY:
[[]] ON-SITE
(active On-site contract required)
[[]] SERVICE CENTER
[[]] CHANNEL PARTNERS
AVAILABILITY: PRODUCT'S SUPPORT LIFE
[[]] Calibration Required
[X] Calibration NOT Required
ADDITIONAL INFORMATION:
STANDARDS
LABOR: 0.1 Hours
SERVICE:
[[]] RETURN
INVENTORY:
[[]] SCRAP
[x] SEE TEXT
NO CHARGE AVAILABLE UNTIL: 20 April 2017
PRODUCT LINE: WC
AUTHOR: SJ
© Keysight Technologies 2015-2016
Modification Recommended Service Note
Version 1.2 | April 25, 2016| Page 1 of 3
U1252B-06
Supersedes:
NONE
USED
[[]] RETURN
PARTS:
[[]] SCRAP
[x] SEE TEXT

Advertisement

loading
Need help?

Need help?

Do you have a question about the U1252B and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Keysight Technologies U1252B

  • Page 1 U1252B-06 Modification Recommended Service Note Supersedes: NONE U1252B Handheld Digit Multimeter Serial Numbers: “See Appendix (A)" The Problem – Capacitance measurement may be out of specification at 100nF range and below. Parts Required: Description Qty. Pomona 2BC-36 double banana plug both ends cable assembly...
  • Page 2 10. Verify the “Open” adjustment has been performed correctly by repeating step 1 to step 4. Customers who have purchased the U1252B Multimeter within the serial number range stated in this service note will receive a complementary Pomona 2BC-36 cable assembly.
  • Page 3 MY54050023 MY54130268 MY54220156 MY54250119 MY54370008 MY54370174 MY54050024 MY54150015 MY54220244 MY54250146 MY54370039 MY54370176 MY54050034 MY54150028 MY54220258 MY54250173 MY54370046 MY54380010 MY54050035 MY54150040 MY54220271 MY54250183 MY54370047 MY54380015 © Keysight Technologies 2015-2016 Modification Recommended Service Note Version 1.2 | April 25, 2016| Page 3 of 3...