4) Should the bottom measuring tip touch the block first, you can use the torx screwdriver
and make a new adjustment of your lower mess point. Loosen the screw from the shaft
holder and move the shaft. Then fix the screw again.
5) Set the digital dial gauge to 10.0000 mm and make some measurements again. The
value must be 10.0000 mm.
6) To know how save your repeatability measurement is, you can form the arithmetic
mean value with this formula:
��
+��
1
2
�� =
n
7) If you have big deviations, please check the situation/quality of the special measuring
tips. It is important that you use only the Xonox measuring tips, no ruby or similar.
8) To control the mechanical axis, demount both measuring tips and put the measuring
sleeve on the shaft.
+⋯��
��