Quality assurance document for the CT 200 measuring instrument If you have any questions do not hesitate to contact us at any time. Mail: m.kloeckner@xonox-tec.com // Phone: +49 6441 21049-13 Check of mechanical alignment 1) Place 10mm gauge block on the lens transducer and tighten the chuck.
Page 2
7) If you have big deviations, please check the situation/quality of the special measuring tips. It is important that you use only the Xonox measuring tips, no ruby or similar. 8) To control the mechanical axis, demount both measuring tips and put the measuring...
Page 3
9) Now you can check the alignment of the mechanical axis: Lead the upper and lower shaft together and move the measuring sleeve over the shafts. 10) When the alignment of the axis are correctly, the measuring sleeve can be moved without chopping.
Page 4
For a significant measuring value you can use a CaliBall. We make the quality assurance for the CT’s with a CaliBall Ø 25,400. Tolerance for the measurements are 4 µm. It is absolutely forbidden to set the dial gauge to zero when the measuring tips stay in that the alignment of the CT don’t be checked contact! In additional it is very important in this way!