Step 7. Reinstall the Front Panel Assembly and Install the Front
Panel Jumpers
Before installing the front panel assembly onto the analyzer, lift and support the
CAUTION
front of the analyzer chassis.
Refer to
Figure 8
for this procedure. New parts are listed in
1. Make sure all 12 of the hex nuts on the front-panel cable connectors have been tightened
using a 5/16-in torque wrench set to 21-in lbs.
2. Reconnect the ribbon cable (item
3. Slide the front panel over the test port connectors being careful to align the power switch
and floppy disk drive to their corresponding front panel cutouts. Ensure that the ribbon
cable (item
) is located below the fan to prevent it from being damaged by the fan blades.
4. With a T-10 TORX driver, install the eight screws (item
5. Install the six semirigid jumpers (item
Figure 8
Front Panel Assembly Reinstallation
14
) to the A3 front panel interface board.
) on the front panel, and tighten to 10-in lbs.
Table 1 on page
4.
) in the sides of the frame.