Summary of Contents for Global Specialties 1368

  • Page 1 1368 1369 1370 Triple Output Programmable Power Supplies User Manual...
  • Page 2: Table Of Contents

    Limited One-Year Warranty ................. 27 Service......................... 28 © 2019 Global Specialties. All rights reserved. Unauthorized duplication of Global Specialties documentation materials is strictly prohibited. Customers are permitted to duplicate and distribute Global Specialties documentation for internal educational purposes only. globalspecialties.com...
  • Page 3: Safety

    Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design, manufacture, and intended use of the instrument. Global Specialties assumes no liability for a customer’s failure to comply with these requirements.
  • Page 4 Global Specialties Definitions MEASUREMENT CATEGORY - Classification of testing and measuring circuits according to the types of mains circuits to which they are intended to be connected. • MEASUREMENT CATEGORY I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the mains supply.
  • Page 5 Global Specialties for repair. Notify Global Specialties of the nature of any contamination of the instrument.
  • Page 6 Global Specialties Servicing Do not substitute parts that are not approved by Global Specialties or modify this instrument. Return the instrument to Global Specialties for service and repair to ensure that safety and performance features are maintained. Cooling fans This instrument contains one or more cooling fans. For continued safe operation of the instrument, the air inlet and exhaust openings for these fans must not be blocked nor must accumulated dust or other debris be allowed to reduce air flow.
  • Page 7: Compliance And Certifications

    Global Specialties Compliance and Certifications CE Compliance This product meets the essential requirements of the applicable European Directives as follows: • 2004/108/EC: Electromagnetic Directive (EMC) • 2006/95/EC: Low Voltage Directive (LVD) Disposal (Applicable in the European Union and other European countries with separate collection systems).
  • Page 8: Introduction

    Global Specialties Introduction Overview The Performance Series (Models 1368, 1369, 1370) are linear triple output power supplies with fully programmable isolated outputs. These outputs can be adjusted independently or combined in series or parallel to output higher voltage or current.
  • Page 9: Product Overview

    Global Specialties Product Overview Front Panel 1. USB interface to update firmware 2. Model name and number 3. LCD display, 480 × 272 pixels, 24-bit color, 4.3-inch color TFT LCD screen 4. Adjustment knob, selection tool and fine adjustments 5. Arrow keys for moving the cursor 6.
  • Page 10: Rear Panel

    Global Specialties Rear Panel 1. USB Interface for communication by PC via USB DEVICE 2. RS-232 Function is same as USB 3. Power socket connecting 115/230 V 50 Hz AC power supply 4. Remote test terminal adds same function as output in front panel, 4-terminal sampling function is added for remote sense 5.
  • Page 11: Display

    Global Specialties Display 1. Display menu to display the name of current page 2. Status and error information to display the error information or current working status. 3. Quick setting and display where to set the voltage, current. Real time sampling numbers and timing will also be displayed 4.
  • Page 12: Operation

    Global Specialties Operation • Power on the unit and let warm up for 25 minutes. • Select the proper connection cable/wire. • Connect the load to the [+,-] terminal on the power supply. • In Series or Parallel modes, the connection cable in the output terminal must be connected in series or parallel.
  • Page 13: Ch Set

    Global Specialties ENTER to input the voltage or current value. Output The CH1, CH2, CH3 buttons on the left of panel are matched with output on or off in the 3 channels. When the parameter is set on one channel press the output; the button will light up blue.
  • Page 14: Ch File

    Global Specialties • SAVE VALUE - Saves the parameters in the CH File for the current channel. CH File Press CH FILE to enter <CH1 FILE>. This page is used to recall or delete the saved data for the specified channel. Up to 40 groups of data can be saved for each channel including all outputs as a group.
  • Page 15 Global Specialties • USB-TMC • Baud Rate - the rate at which information is transferred. The higher the baud rate, the more sensitive the cable is to length and connection quality. • 4800 • 9600 • 19200 • 38400 • 57600 •...
  • Page 16: Specifications

    Global Specialties Specifications All specifications apply to the unit after a temperature stabilization time of 20 minutes over an ambient range of 20°C ± 5°C. Electrical Model 1368 1369 1370 Channel CH1 CH2 CH1 CH2 CH1 CH2 Voltage 0 to 30 V...
  • Page 17: Interface And Communication

    Global Specialties Mechanical and Environmental Model 1368 1369 1370 Weight (lbs) 17.9 28.7 Dimensions 17 x 10.5 x 4.5 19.2 x 10.5 x 4.5 ( L x W x H in) Normal Working Temperature (˚C) 0 to 40, <90% RH Reference Working Temperature (˚C)
  • Page 18 Global Specialties RS-232 Interface Signal Code Connector Pin Number Transmitted Data Recieved Data Signal Ground Common Note: The pin-out of these power supplies is the same as the standard 9-core connector. The 9-core DB type socket of RS-232 in these power supplies is as below:...
  • Page 19: Usb Tmc

    Global Specialties Communicate with PC Based on the table above, the pin-out of these supplies is the same as the 9-core connector of IMB AT. The user can make 3 wire connection cables (length is less than 1.5 m) by using double core shielding wire. For custom cables, please note Pin 4 and 6 and Pin 7 and 8 must be shorted in PC interface.
  • Page 20 Global Specialties • Open NI MAX and expand the “Devices and Interfaces” tree view Commands In this device, SCPI commands are referred. SCPI is a tree structure with 3 levels at most, where the top level is called subsystem command. Only subsystem is selected, then the bottom level is valid, use the colon to separate the structure.
  • Page 21: Usb Cdc Virtual Serial Port

    Global Specialties USB CDC Virtual Serial Port Select the bus mode to “USBCDC” on the power supply. Reboot is necessary for the choice to reflect on the PC if switching from USBTMC. This makes USB interface confi gure as Vcom Confi guration •...
  • Page 22: Scpi Command

    Global Specialties • Example: FUNC:IMP CPD, IMP is the command and CPD is its parameter. • There is no parameter in some command • Example: SYSTem:LOCK • Blank space (_) can’t be put in front or behind colon. • Example: FUNC_:_IMP CPD FUNC:IMP CPD •...
  • Page 23 Global Specialties • INSTrument:NSELect <1|2|3> • Select the channel to operate, which is the base of other commands • INSTrument:NSELect ? • Inquire the current channel • OUTPut output command • OUTPut <0|1|ON|OFF> • The command is used to control the on or off of output •...
  • Page 24 Global Specialties • TIMer timer set command • TIMer <0|1|ON|OFF> • The command is used to control the switch of timer • TIMer? • Inquire the status of timer, return 0 or 1 • TIMer: DATA<0 ~ 99999.9> • The command is used to set the time of selected channel, Min. value is 0 and Max.
  • Page 25 Global Specialties • MEASure measure command • MEASure:VOLTage? • The command is used to inquire the measured voltage value • MEASure:CURRent? • The command is sued to inquire the measured voltage value • MEASure:TIMer? • The command is used to inquire the current timer value •...
  • Page 26 Global Specialties • The command is used to inquire the status of beeper • MENu:PMEM <DEFault|USEr> • The command is used to set memory status, default and user • MENu:PMEM ? • Inquire the memory status • MENu:RWAIT <0s|5s|10s|20s|30s|60s> • The command is used to set the retest wait time •...
  • Page 27: Limited One-Year Warranty

    Global Specialties Limited One-Year Warranty Global Specialties (a brand of Cal Test Electronics) warrants these products to be free from defective material or workmanship for a period of 1 year from the date of original purchase. Under this warranty, Global Specialties is limited to repairing the defective device when returned to the factory, shipping charges prepaid, within the warranty period.
  • Page 28: Service

    Global Specialties Service If you have a need for calibration or repair services, technical, or sales support, please contact us: Global Specialties™ (A brand of Cal Test Electronic, Inc.) 22820 Savi Ranch Parkway Yorba Linda, CA 92887 800-572-1028 or 714-221-9330 globalspecialties.com...