Download Print this page

Agilent Technologies 8960 Installation Note page 4

Mdc upgrade for e5515b/c. kit part number: e5515bu-506 and e5515cu-506
Hide thumbs Also See for 8960:

Advertisement

Disassembly:
Remove outer cover and top cover for the 8960
1.
Remove handles and rear bumpers.
2.
Remove the screws on the rear panel holding outer cover.
3.
Slide outer cover off.
4.
Remove the 26 screws holding the top cover on and remove the cover.
Remove the old boards
1. Remove the old Measurement Down Converter module. Refer to the Replaceable
Parts List (E5515A/B/C/T) in the Troubleshooting and Repair section of the 8960
Assembly Level Repair area at:
location.
Assembly:
Install New Measurement Down Converter Module
1. Install the Measurement Down Converter as per diagram on page 5 of the
Replaceable Parts List document, referred to above.
Install Option Label
1. Install the option label, 7121-7884, on the back of the instrument.
a. Write 506 on the label if blank.
b. Don't cover up the old option label if the options on the old label are not
also on the new label.
Replace covers:
1. Replace the top cover and secure it with the 26 screws.
2. Slide the outer cover on.
3. Replace the outer cover screws on the rear panel.
4. Attach the rear bumpers and handles.
Adjust and Test and after Upgrade:
1. Turn on the test set and warm up for 30 minutes.
2. Perform the IQ Calibration for SOURCE 1 & 2
3. Perform the Burst Mod Offset 1 Calibration
4. Perform the Digital Average Power calibration
5. Cycle the power on the instrument
6. Perform the Spectrum Monitor Calibration
7. Run the Test Set Verify software
after the upgrade.
Questions and concerns: contact Agilent
Technologies Support:
Phone: (800) 827-3848 (U.S. & Canada)
Fax:
E-mail:
Web:
www.agilent.com/find/8960toolbox
www.agilent.com/find/8960toolbox
(509) 921-3848 (International)
(509) 921-3514
spokane_service@agilent.com
www.agilent.com/find/8960support
for board
to verify the test set
4

Advertisement

loading