You can select the following units:
mmol/L
g/dL (default)
g/L
ES C
O K
5.3.6

Hgb mode and limits

The function of enabling the Hgb mode and setting the limits can
be found under "Menu", sub-menu "Hgb limits". This gives you the
ability to enable or disable the Hgb mode and to set the normal
range limits of each patient group. This setting influences all test
displays, including the results in the memory (data storage) and
the printer.
Limits can be set for the following groups of patients:
Adult Male - (13 – 18 g/dL) default
Adult Female - (11– 16 g/dL) default
Child - (11 – 16 g/dL) default
Limits can be set within 5 – 25.5 g/dL (0.31 – 15.81 mmol/L).
While the "Hgb limits" mode is activated, the lower limits must
show different values than the upper limits of all groups of patient.
Otherwise the "Hgb-limits" mode will be inactive.
5.3.7

Setting the display contrast

The LCD display contrast is strongly dependent for technical
reasons on the environmental temperature. You can set the
contrast by entering "Menu", sub-menu, "Contrast". Default
setting is "3". Higher numbers will darken the screen.
Select the appropriate entry using the
arrow buttons.
Press this button to reject the selection.
Press this button to confirm the selection.
HemoPoint
H2 User Manual
®
Further functions
TIP
39