10 10 10 10 Calibration and maintenance
Calibration and maintenance
Calibration and maintenance
Calibration and maintenance
The device must be calibrated every 6 months. This is necessary in order to ensure the
specified measuring precision (see specifications) during that period of time.
Calibration can only be performed by authorised specialists or EnviteC Service. The
equipment required for it (calibrator, flow meter, pump, service instructions, service software
and calibration solution) can be ordered from EnviteC-Wismar GmbH.
The device should be serviced every 12 months. This includes the following measures:
Calibration:
Calibration of the device
Testing:
Test whether device functions are in proper working order
Maintenance: Measures to keep the device in perfect working order
Repair: Measures to restore the specified device function
ATTENTION:
Calibration and maintenance must be carried out by trained service personnel only. It is recommended that a service
contract be concluded.
27
AlcoQuant® 6020 – Breathalyser – Operating Instructions