Carbolite MTF 15 Installation, Operation & Maintenance Instructions Manual page 6

900c tube furnace
Table of Contents

Advertisement

MTF 9/15
Altering the Ramp Rate
It is possible to limit the rate of heating by setting a ramp rate. Press Scroll
(SetPoint ramp rate) is displayed. Use Down
The ramp rate sets the maximum rate of heating or cooling in degrees per minute. A value of
cancels the ramp rate, allowing heating and cooling at the maximum rate. When this feature is in
use, there is a "working setpoint" which can be viewed at any time by scrolling to
pressing
or
Fig 1 and fig 2 indicate the possible difference between running without and with a ramp-to-
setpoint value (depending on the load and the value used).
4.6
Navigation Diagram
Home
List
20.0
measured temperature;
*C
use
setpoint
OP
output power,
read only
w.SP
present only if
SPrr in use
m-A
manual/Auto
mAn=off, Auto-on
SPrr
setpoint ramp rate
OFF or value
tm.OP
timer mode
tmr
time remaining
dwell time for
dwel
timer
StAt
timer status
run or OFF
6
.
Input
List
*C
/
to access
CAL.P
CAL
Pnt.L
OFS.L
Pnt.H
OFS.H
or Up
to display and adjust the value.
Output
iP
OP.Hi
enter
password
power limit
setting, if
present
if user
calibration
user 2-point
calibration
until the legend
Access
List
List
oP
ACCS
codE
for factory access to
lists and parameters
not available to the
operator
SPrr
OFF
w.SP
and
MF09