LASER:
Hold down the MEAS. button and press the DOWN button at the same time to
turn the laser ON or OFF.
Note: Non-contact infrared thermometers are not recommended for use in
measuring the temperature of shiny or polished metals.
** The thermometer will automatically shut off if left to idle for more than 60 sec,
unless in PRB mode. (In PRB mode, it will shut off if left idle for more than 9
minutes.)
STORAGE & CLEANING
The thermometer should be stored at room temperature between
-4º to 149ºF (-20º to 65ºC). The sensor lens is the most delicate part of the
thermometer. The lens should be kept clean at all times. Care should be taken
when cleaning the lens using only a soft cloth or cotton swab with water or
medical alcohol. Allow the lens to fully dry before using the thermometer. Do
not submerge any part of the thermometer.
LCD ERROR MESSAGES
HI
LOW
Er2 Er3
Error 5-9: for all other error messages it is necessary to reset the
thermometer. To reset it, wait for auto power off, remove the
Er
battery and wait for a minimum of one minute, reinsert the battery
and turn it on. If the error message remains please contact the
Service Department for further assistance.
'Hi' or 'Lo' is displayed when the temperature being measured
Hi Lo
is outside of the measurement range.
BATTERY REPLACEMENT
(Remember to power off the unit before replacing the battery).
BATTERY OK
When the Low Battery icon displays, the batteries should be replaced.
(AAA batteries – 2 pcs.)
Dispose of used battery promptly and keep away from children.
SPECIFICATIONS
Non-Contact Temp. Range
Thermocouple Temp. Range -83.2º to 2552ºF (-64º to 1400ºC)
Full Range Accuracy
Resolution
Response Time (90%)
Distance : Spot
Battery Life
Dimensions
Weight (Including batteries) 13.62 oz (386.1 grams)
ATTENTION!
• DO NOT LOOK DIRECTLY INTO THE 2 LASER BEAMS, PERMANENT
EYE DAMAGE MAY RESULT.
• USE EXTREME CAUTION WHEN OPERATING THE LASER!
• NEVER POINT THE DEVICE TOWARDS ANYONE'S EYES!
• KEEP OUT OF REACH OF CHILDREN!
HI or LO is displayed when the temperature being measured
is outside of the settings of (HAL) High Alarm Level and (LAL)
Low Alarm Level.
'Er2' is displayed when the thermometer is exposed to rapid
changes in the ambient temperature. 'Er3' is displayed when
the ambient temperature exceeds -32ºF (0ºC) or 122ºF (50ºC).
The temperature should be allowed plenty of time (minimum
30 minutes) to stabilize to the working/room temperature.
BATTERY LOW
-76º to 1832ºF (-60º to 1000ºC)
+ 2% of reading or 2˚C (4˚F) whichever is greater
0.1ºF (0.1ºC)
1 second
50 : 1
Typically 14 hrs
1.85 x 7.71 x 7.90" (47 x 195.9 x 200.6 mm)
BATTERY EXHAUSTED
52225-B-INSTUNIV