BW TECHNOLOGIES
GasAlertClip Extreme
2 Year Gas Detector
Instruction Sheet
Introduction
The GasAlertClip Extreme gas detector ("the detector") is a personal safety
device that warns when hazardous gas exceeds factory set alarm setpoints.
The detector stores and transmits gas alarm event data. It is your
responsibility to respond properly to the alarm.
Models with internal vibrator
alarm and event logging
GA24XT-H
Hydrogen Sulfide (H
GA24XT-M
Carbon Monoxide (CO)
GA24XT-S
Sulfur Dioxide (SO
GA24XT-X
Oxygen (O
a Safety Information - Read First
Warning: Substitution of components may impair intrinsic safety.
Warning: To prevent ignition of flammable or combustible
atmospheres, disconnect power before servicing.
⇒
Do not activate the detector after the date on the package.
⇒
This product is a gas detector, not a measurement device.
⇒
Perform a self-test each day prior to use.
⇒
Make sure the sensor grill is free of dirt and debris.
⇒
Make sure the sensor grill is not covered.
⇒
Periodically test the sensor's response to gas by exposing the
detector to a target gas concentration that exceeds the low alarm
setpoint. Manually verify that the audible and visual alarms are
activated.
⇒
Periodically calibrate the GasAlertClip Extreme O
This instrument contains a lithium battery. Do not mix with the solid
waste stream. Spent batteries should be disposed of by a qualified
recycler or hazardous materials handler.
Parts of the GasAlertClip Extreme
Gas Monitored
S)
2
)
2
)
2
A
B
C
D
E
F
G
.
2
c
Note:
Audible alarm
Visual alarm
Infrared download port
LCD display
Clip
Start/test button
Sensor and sensor grill
Display Features
A
Maximum exposure under alarm
B
Data transmission
C
High and low alarm setpoints
D
Self-test status
E
Gas monitored
C F
Alarm condition
G H
Life-remaining indicators
A H
Time since maximum exposure
Pushbutton
Pushbutton
Description
•
To turn on the detector, press and hold C for 5
seconds.
•
Press C within 24 hours of receiving a gas alarm
to view the maximum gas exposure.
•
When Test is displayed, press and hold C for
C
approximately 1 second to activate the self-test.
•
To calibrate the GasAlertClip Extreme O
and hold C for approximately 3 seconds.
•
To display the gas alarm setpoints, press C.
•
To transmit the data press C when Prn and
are displayed.
, press