Safety Precautions If this device is not used according to safety precautions, accidents can occur which may cause damage to the instrument, the operator or other items. Our design has included substantial consideration for product safety but please read the precautions contained in this user manual and use the product correctly.
■ Measurement Principles This device is designed to measure the whiteness of rice based on our own uniquely developed standard. The C-600 measures reflectivity and utilizes two blue LEDs as its light source. The light from the blue LEDs shine onto the sample surface at a 45 degree angle. Next, the amount of reflected light reflected is measured with a photo diode.
RS-232C Connector Exhaust Glass Filter <Included Parts - with part numbers> Whiteness Standard Plate Sample Case Sample Platter Fixed Quantity Shooter Sample Cup C-600-01 C-600-02 C-600-03 C-600-04 C-600-05 Main Unit Cover Glass Wipe Replacement Sponge Replacement Packing (x3) Replacement Fuse...
4. Description of Operation Keys <Operation Panel> Function Used to turn the printer output ON or OFF. Used as a backspace when inputting numerical values. Used to input standard plate values. Used to finalize numerical input. Used to input numerical values. Used for sensitivity calibration. Used to display the average of the measured values.
ON, the display will repeatedly show "INSERT STD. "If this should occur, insert the whiteness standard plate into the sample case slot on the main unit until it cannot be inserted any further. The C-600 is now ready for measurement.
6. Measurement [Caution] Before performing any measurements, use the glass wipe to clean the glass surface of the sample case and the main unit's glass filter. 6-1. Inserting the Sample for Testing Fixed Quantity Shooter Place the sample case on a flat surface and set the sample platter. Sample Platter Sample Case (2) Place the fixed quantity shooter on the sample platter.
Page 10
Sample Case 6-2. Insertion Into the Main Unit Insert the sample case into the sample case slot on the main unit until it cannot be inserted any further. [Caution] Make sure the sample case is facing the correct direction. 6-3. Measurement Once the sample case has been inserted into the main unit, Sample Case Slot the number of times measured and the whiteness value are...
7. Maintenance and Other Functions 7-1. Main Unit Glass Filter Handle If the glass filter on the main unit is dirty, you will not receive an accurate measurement. You must keep the glass filter Front Cover clean at all times when using the device. Use your finger to lift upwards on the handle on the inside of the front cover, in order to remove the front cover.
7-4. Inputting a Whiteness Standard Value If the device is turned ON, turn the device OFF. (2) Insert the whiteness standard plate into the main unit. (3) When you turn the power switch ON, "WARMING-UP" is displayed. (4) After a few moments, the light on the lower right side will begin to flash. When this happens, press the key.
Page 13
[Caution] Refer to the printer's user manual for more details about how to connect the device to the printer. [Printout Example] [Caution] Whiteness values outside the range of 5.0 to 69.9 can- not be output. RICE WHITENESS TESTER C-600 TIMES 39.6 39.5 39.7 39.6...
8. Errors When an error occurs with this device, an error number is displayed alongside an error message. Display Cause Solution Circuitry error. Repair is required. No whiteness standard plate has Inser t the whiteness standard been inserted when attempting to plate, then tr y per forming the perform a sensitivity calibration.
Page 16
Notes ● Copying some or all of the contents of this user manual without prior written consent is strictly prohibited. ● The contents of this user manual may be changed at any time in the future without any prior notice. ● The appearance and/or representations of the products and parts depicted in this user manual may not appear exactly as their actual counterparts, but this does not affect their operation or functionality. ● This user manual was intended to be written as clearly and accurately as possible. However, if you are unclear about anything in this user manual or notice any missing information, please contact us directly. ● We cannot be held responsible for any actions or effects resulting from the execution of any operations outlined in this user manual.