Page 1
Powder Whiteness Tester C-130 Operating Manual Thank you for purchasing this product. Please read the operating manual carefully and use this product properly.
Page 2
For safety precautions Improper use of the Powder Whiteness Tester in violation of the following safety notes may result in death, injury or damage to property due to fire, etc. While the safety of the product has been given considerable attention, read the precautions in the operating manual and use the instrument properly.
Table of Contents 1. Measurement Principles and Characteristics........4 2. Specifications ..................5 3. Part Names & Accessories ..............6 4. Description of Operation Keys .............7 5. Preparing for Measurement ..............8 6. Measurement ..................9 7. How to Make User Calibration Curve ...........12 8.
<Configuration Diagram> ■ Characteristics The new C-130 model is smaller and more compact than our previous C-100 model. Furthermore, using blue LED light sources allow for a longer light source lifetime reduced power consumption and reduced heat generation. In addition, the required sample size has been greatly reduced.
2. Specifications Measurement method : Reflectivity measurement Accepted samples : Various types of powder (Starch, Wheat flour, Tapioca flour, Cement etc.) *(starch is the base calibration curve, but the user can set their own calibrations) Measured item : Whiteness (Different from JIS definition) It is diffierent from "Haku Shoku Do"...
4. Description of Operation Keys <Operation panel> Function Select calibration curve and input numerical value (0). 〜 Used to input numerical values. Turn printer on/off. When changing numerical values, use this key as the backspace key. Switch between measurement mode and standard plate check mode. Measurement mode Use for normal sample measurements and when creating a user calibration curve.
5. Preparing for Measurement 5-1. Connecting the power cord [Caution] Power supply voltage may be different when using this product in other countries. Please only use a power cord designed for your country. Insert the power cord into the power supply connector on the rear of the main unit.
6. Measurement [Caution] Before performing any measurements, use the glass wiper to clean the glass filter of the main unit (refer to 8.1). 6-1. How to pack the sample (see the separate ‘Simple Guide to Sampling’) [Note] The following is the standard measurement method for starch. Different packing methods will apply to samples with larger diameters, viscosity, moisture, runny sample, etc.
Page 10
If the reading is outside of the measurement range, UNDER or OVER will be displayed. ■ Selecting the Calibration Curve The C-130 can have up to ten calibration curves. [01] is for the base calibration curve, and [02-10] provide for nine additional customer calibrations that can be stored and later retrieved.
Page 11
■ Average Value This device c an display the average value for 2- 9 measurements. After measuring 2-9 times, press the key and the average value, in addition to the number of measurements, will display. Displaying the average value deletes all previous values and makes the next measurement the first measurement of the next set.
In such cases, it is possible to calculate the reading by creating a user calibration curve. [Note] You can register up to 9 user calibrations from [02] to [10]. C-130 (Base Calibration Curve [01]) 7-2. Creating a calibration curve <Table 1> Sample preparation ◆ Calibration Curve (CC) Number: 03 Prepare 5 samples of various whiteness levels.
Page 13
(1~3), and then after pressing the next 2 whiteness values are displayed (4~5). [Note] The values shown in the [C-130] column are the measurements in the base calibration curve [01]. The values shown in the [USER] column are the values input by the user. Before calibration, both the [C-130] and [USER] values are the same.
Page 14
→ . Press to confirm Sample (1). The display will change from [USER] to [C-130], and the upper right number will change from [#1] to [#2]. Repeat [Procedure 1] and [Procedure 2] to input Samples (2) to (5). Once you have input Sample (5), press complete the calibration curve input operation.
(3) Clean the glass filter. Clean the drawer rails with cleaning brush (C-130-07). There is a ridge around a section of the glass filter on the bottom of the main unit. Be very careful not to damage the ridge.
Page 16
(3) When the printer is ON, readings for measurement periods as well as the reading after the key has been pressed can be printed. <Printout example> [Note] Measurement results that are outside of the measurement range cannot be printed. WHITENESS TESTER Model C-130 Calibration No.01 TIMES 83.2 83.4 82.9 83.1 83.2...
Page 17
Data bit length : 8-bit stop bit data bit Parity start bit Stop byte 1 byte Code ASCII (2) C-130 and RS-232C cable connection ■ Printer VZ-330 (for use with VZC-14 cable) TXD→ ② ③ C-130 VZ-330 ⑤ ⑤ 8-6. Error display If an error occurs with this instrument, an error number and message will be displayed.
Page 18
■ Calibration curve number : ■ Calibration curve name : ■ Sample name : C-130 Reading from Previous Machine Base Calibration Curve [01] or User Standard Method ① Sample ② Sample ③ Sample ④ Sample ⑤ Sample ■ Calibration curve number : ■ Calibration curve name : ■ Sample name : C-130 Reading from Previous Machine Base Calibration Curve [01] or User Standard Method ①...
Page 20
Notes ● Copying some or all of the contents of this user manual without prior written consent is strictly prohibited. ● The contents of this user manual may be changed at any time in the future without any prior notice. ●...