Customer Wiring Connections - Process Sensors MCT Series Installation And Operation Manual

Multiple constituent transmitter
Table of Contents

Advertisement

11.0 Customer Wiring Connections

All wiring connections are made to three quick-disconnect connectors mounted on the rear of the sensor
enclosure. These connectors may have different signals coming to them from inside the sensor. The
following diagrams show the standard configurations for these connectors.
POWER
Terminal
#
1
2
Gnd
Symbol
ANALOG Connector (4-20mA outputs) – Standard configuration
Terminal
#
1
2
3
4
5
6
ANALOG Connector (0-10V outputs) – Alternate 1
Terminal #
1
2
3
4
5
6
Input
Live
Neutral
Ground
Signal
Gnd
4-20 mA
Gnd
2 (1 if sensor is in single mode)
4-20 mA
2 (1 if sensor is in single mode)
Gnd
3 (not operational in single mode)
4-20 mA
3 (not operational in single mode)
Signal
Gnd
0-10V
Gnd
2 (1 if sensor is in single mode)
0-10 V
2 (1 if sensor is in single mode)
Gnd
3 (not operational in single mode)
0-10 V
3 (not operational in single mode)
Constituent #
1
1
Constituent #
1
1
36

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mct 300