.6 Diagnostics

The MCT's software continually monitors several internal electronic variables to ensure that the sens
is operating correctly at all times. If any of these parameters go out of pre-set limits then the M
display will provide a warning message. Maintenance perso
p
arameters by selecting Diagnostics from the Main Menu.
Move DOWN to DIAGNOSTICS and then move right to enter the menu. The Diagnostics m
selection is not protected by a password as none of the parameters displayed may be altered.
Moisture= 12.89%
DIAGNOSTICS
The Diagnostics Menu allows the user to view the status of various internal set up parameters, voltages
and signal levels. Pressing the DOWN/UP key will scroll through the various parameters monitored by
the system.
Moisture = 12.89%
Thermistor: 3.20v
Moisture = 12.89%
Cooler Drive: 1.61v
Moisture = 12.89%
Gain: 156 (7.1v)
nnel may view the values of these
This parameter is the internal temperature
signal of the Pbs detector.
This parameter indicates the voltage of the
thermocooler being used to keep the
detector at the set temperature of 20ºC.
This is the gain factor of the pre-amp and
the strength of the largest filter signal on
the detector.