Calibration
CO2 Calibration
Humidity Calibration
Temperature Calibration
PC Interface, Software Installation and Operation
This meter has the capability to connect to and communicate with a PC. To install and use
the software, please refer to the instructions provided on the supplied CD-ROM and/or the
instructions provided in the HELP Utility within the software program.
Check the software download page of the website
the PC software and its operating system compatibility.
Find Quality Products Online at:
1.
Press "
" to turn the meter on.
2.
Place the meter in a known CO2 reference for 10 minutes.
3.
Press the "SET" button 5 times until "C-01" is displayed in the second display.
4.
Press "►" or "◄" to select digits to adjust (flashing).
5.
Press "▼" or "▲"to adjust the display to the reference value.
6.
Press "" to store the value and exit the calibration mode.
1.
Press "
" to turn the meter on.
2.
Place the probe in a known humidity reference for 60 minutes.
3.
Press the "SET" button 6 times until "SET" and "CAL" are displayed on the LCD.
4.
Press "▼" or "▲" to adjust the display to the reference value.
5.
Press "" to store the value and exit the calibration mode.
1.
Press "
" to turn the meter on.
2.
Place the probe in a known temperature reference for 60 minutes.
3.
Press the "SET" button 7 times until "SET" and "CAL" are displayed on the LCD.
4.
Press "▼" or "▲" to adjust the display to the reference value.
5.
Press "" to store the value and exit the calibration mode.