Calibration - ABB MU 80 Product Manual

Robotics. motor units and gear units
Table of Contents

Advertisement

5 Calibration

When to calibrate
Mechanical units must be calibrated after installation, if parts of the transmission
are replaced, or if the resolver memory is lost.
Fine calibration defines the calibration zero position and updates the revolution
counters.
Robot calibration is described in Operating manual - IRC5 with FlexPendant, and
in the product manual for the robot.
Commutation of motors is described in Application manual - Additional axes and
stand alone controller.
Coordinated positioners and user frames are described in Application
manual - Additional axes and stand alone controller.
Calibration position
Any position can be defined as the calibration zero position but it must be possible
to move the axis to this position with good precision repeatedly. For example, this
can be a mechanical stop or another indicator that clearly shows the position.
Fine calibrating mechanical units
Use this procedure to fine calibrate the mechanical units.
1
2
3
4
5
Updating revolution counters
Use this procedure to update revolution counters.
1
2
3
4
Product manual - Motor Units and Gear Units
3HAC040148-001 Revision: K
Action
Move the mechanical unit to a suitable zero position
for calibration.
On the FlexPendant ABB menu, tap Calibration.
Tap to select the mechanical unit.
Tap Fine Calibration.
Tap to select the axis and then tap Calibrate.
Action
Move the mechanical unit to the zero position for
calibration.
On the FlexPendant ABB menu, tap Calibration.
Tap Rev. Counters and then tap Update Revolution
Counters.
Tap to select the axis and then tap Update.
© Copyright 2011 -2018 ABB. All rights reserved.
5 Calibration
Note
Mark the zero position for future
reference.
Note
73

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents