Specifications
Power :
•
Environmental :
•
•
•
•
•
•
Sensor Excitation :
•
Input :
•
Display :
•
Relay :
•
•
Analog Output :
•
•
Cable :
•
•
Accuracy :
•
Enclosure :
•
•
•
SITRANS CU02
Operating Instructions, 03/2017, A5E40873245_AA
see nameplate for voltage configuration (100/115/200/230 V ac ±15%, 50/60 Hz, 10 VA)
location:
altitude:
ambient temperature:
relative humidity:
installation category:
pollution degree:
26 Vdc nominal, 70 mA max
SITRANS Sensor 0 – 10 Vdc
liquid crystal
2 alarm/control relays
1 form 'C' SPDT contact per relay, rated 5 A at 250 V ac non inductive
isolated 4 - 20 mA
750 Ω load max
analog output
latch contact input:
±0.02V (display) or ±40μA (mA output)
55 mm W x 75 mm H x 110 mm D (2.2" W x 3" H x 4.4" D)
polycarbonate
mounting:
indoor
2000 m max
-20 to 50 °C (-4 to 122 °F)
80% for temperatures up to 50 °C
II
2
three 9 mm (0.35") digits multisegment graphic
for operation status
Belden 8760 18AWG shielded twisted pair or
equivalent
Belden 8760 18AWG shielded twisted pair or
equivalent
DIN rail (DIN 46277 or DIN EN50022)
•
wall / panel mount
•
2
7