Changing Color Splitter In P&C Reflector Module - Zeiss Axiostar plus Operating Manual

Transmitted-light microscope
Table of Contents

Advertisement

Axiostar plus
We recommend the following procedure if you
need to install filters without direction markings
(arrow):
Filters with reflective, dielectric layers should be
installed so that the reflective layer
Points out for an excitation filter (in relation
to the reflector module)
and
Points in for an emission filter (Fig. 2-28).
2.11.7
Changing color splitter in P&C
reflector module
Extreme care is required when attaching the
filter and color splitter to avoid damage to the
optical elements and to keep them clean.
We recommend ordering completely equipped
P&C reflector modules since changing the color
splitter has higher requirements.
To replace a color splitter:
Remove
P&C
reflector
reflector revolver.
Loosen
both
(2-29/1) with a screwdriver.
Hold together both halves of the reflector
module,
turn
against
position and set aside.
B 40-815 e 12/01
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Start-Up
Attaching epi-fluorescence illuminator
module
from
slotted
screws
the
installation
1
Locking filter (emission filter)
2
Reflective layer on emission filter
3
Color splitter
4
Reflective layer on the color splitter
5
Excitation filter
6
Reflective layer on the excitation filter
7
Illumination/Imaging beam path
Fig. 2-28 Inserting filter and color splitter
1
Slotted screws
2
Module half: emission
3
Module half: excitation
Fig. 2-29 Changing color splitter