General Calibration/Adjustment Procedure

The following procedure is the recommended method to complete an
instrument calibration. This procedure is an overview of the steps
required for a complete calibration. Additional details for each step in
this procedure are given in the appropriate sections of this chapter.
1 Read "Test Considerations" on page 54.
2 Unsecure the instrument for calibration (see page 68).
3 Perform the verification tests, beginning on page page 55, to characterize
the instrument (incoming data).
4 Press
5 Select the "Test / Cal" menu.
6 Select Perform Cal.
7 Enter the Setup Number for the procedure being performed. The default
setup number is "1" and, from the front panel, the number will increment
as the procedures are performed.
8 Select BEGIN.
9 For setups that require an input, adjust the value shown in the display
to the measured value and select ENTER VALUE.
10 The setup will automatically advance to the next required value.
To cancel the adjustment procedure, select CANCEL STEP. The display
Note
will return to the setup number entry.
11 When finished, select END CAL.
12 Secure the instrument against calibration.
13 Note the new security code and calibration count in the instrument's
maintenance records.
General Calibration/Adjustment Procedure
on the front panel.
Chapter 4 Calibration Procedures
4
4
71