Specifications - Ametek Crystal Pressure APM CPF Series Operation Manual

Advanced pressure module
Table of Contents

Advertisement

Specifications

Modules
Module
0 - 30%
Range
Gauge Full Scale
psi
± (% Full Scale)
30
0.0075%
100
0.0075%
300
0.0075%
psi
1000
0.015%
3000
0.015%
10000
0.015%
15000
0.015%
bar
± (% Full Scale)
3
0.0075%
10
0.0075%
30
0.0075%
bar
100
0.015%
300
0.015%
700
0.015%
1000
0.015%
kPa / MPa
± (% Full Scale)
300 kPa
0.0075%
1
0.0075%
3
0.0075%
kPa /
10
0.015%
MPa
30
0.015%
70
0.015%
100
0.015%
Accuracy (Gauge)
0 to 30% of Full Scale . . . . . . . . . . . . ± (0.0075% of Full Scale) or ± (0.015% of Full Scale)
30 - 110%
Gauge Full Scale
30 to 110% of Full Scale . . . . . . . . . . ± (0.025% of Reading) or ± (0.05% of Reading)
± (% of Reading)
0.025%
Vacuum . . . . . . . . . . . . . . . . . . . . . . . . . For 100 psi / 10 bar / 1 MPa and lower:
0.025%
0.025%
0.05%
0.05%
0.05%
0.05%
Accuracy specifications include all effects of linearity, hysteresis, repeatability, temperature, and stability for one year.
± (% of Reading)
0.025%
Note:
Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent
0.025%
recertification period.
0.025%
APM modules must be exercised and re-zeroed whenever exposed to significant changes in environmental conditions to achieve
0.05%
0.05%
these specifications. To exercise a module, cycle the module between zero (ambient barometric pressure) and the pressure of inter-
0.05%
est. A properly exercised module will return to a zero reading (or return to the same ambient barometric reading).
0.05%
!
CAUTION: Advanced Pressure Modules (APM) are not recommended for continuous use at high vacuum.
± (% of Reading)
0.025%
Operating and Storage Temperature
0.025%
0.025%
Operating Temperature Range . . . -20 to 50° C (-4 to 122° F)
0.05%
Storage Temperature Range . . . . . . -40 to 75° C (-40 to 167° F)
0.05%
0.05%
0.05%
*
± (0.06% of Full Scale
)
For 300 psi / 30 bar / 3 MPa:
*
± (0.06% of Full Scale
) ±1 LSD
*
Full Scale = -14 . 5 psi / -1 . 0 bar / -100 . 0 kPa .
Specifications
3
APM CPF Series Operation Manual

Advertisement

Table of Contents
loading

Table of Contents