In order to avoid interruptions in the command execution process:
Turn on software (
XON
/
XOFF
) flow control on the PC side.
Avoid long command lines (lines more than 96 characters).
Send the
*CLR
command before starting the communication procedure.
Check the error buffer sending the
FAULT?
„ „ „ „ „

REMOTE COMMANDS AND ERROR CODES

The following tables list all the commands, and their descriptions, accepted by the APM.
Common Commands
Command
Description
*CLS
(Clear Status) Clears the error queue.
*IDN?
(Identification Query) Returns the manufacturer, model number, and firmware revision of the module.
*RST
Resets the module to the power up state.
Module Commands
Command
Description
DAMP
Turns Damp on or off.
DAMP?
Returns if Damp is on or off.
FAULT?
Returns the most recent error code.
PRES_UNIT
Sets the pressure unit for the indicated display.
PRES_UNIT?
Returns the pressure unit from the indicated display.
VAL?
Returns the measured values.
ZERO_MEAS
Zeros the pressure module.
ZERO_MEAS?
Returns the zero offset of the pressure module.
CAL_STORE
This command must be sent anytime a parameter has changed.
command after the communication procedure is finished.
8
Communications
APM CPF Series Operation Manual