Summary of Contents for Dr. Trust Comfort

  • Page 5 Range & broadcasting Systolic Blood -pressure DiastolicBlood -pressure Measures Hypotension lower than 100 lower than 60 Consult your doctor Optimal BP between 100 and 1 20 between 60 and 80 Self-check Normal BP between 120 and 130 between 80 and 85 Self-check High To Normal BP between 130 and 140...
  • Page 9 Error No. Possible cause(s) ERR 1 No pulse has been detected. ERR 2 Unnatural pressure impulses influence the measurement result. Reason: The arm was moved during the Measurement (Artefact). ERR 3 The inflation of the cuff takes too long. The cuff is not correctly seated. ERR 5 The measured readings indicated an unacceptable difference between systolic and diastolic pressures.
  • Page 10 Malfunction Remedy The display remains empty when the 1. Check batteries for correct polarity and if necessary insert instrument is switched on although the correctly. batteries are in place. 2. If the display is unusual, re-insert batteries or exchange them. The device frequently fails to measure the blood 1.
  • Page 11 IEC60601-1-6:2010+A1:2013/ EN60601 - 1-6:2010+A1:2015 IEC60601-1:2005+A1:2012/EN60601 - 1:2006+A11:2011+A1:2013+A12:2014 IEC60601-1-2:2014/ EN60601-1-2:2015 IEC/EN60601-1-11:2015 IEC80601-2-30:2009+A1:2013/EN80601 - 2-30:2010+A1:2015 Some electrical and electrical equipments forbid to abandon TUV NO. and disposal at will Manufacturer’s name and Reading Instruction Book before address Inapplicable baby Type B equipment Protection against IP22 Cuff Connector...
  • Page 13: Immunity Test

    ±8 kV air ±8 kV air or ceramic tile. If floors are covered should be used no closer to any part of Dr Trust BP Check 61000-4-2 with synthetic material, the relative with MDI , including cables, than the recommended humidity should be at least 30 %.
  • Page 14 Separation distance according to frequency of transmitter m Rated maximum output power of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz d = 1.2×p d = 1.2×p d = 2.3×p 0.01 0.12 0.12 0.23 0.38...