Alarms for clinicians; left ventricular assist device
Hide thumbs Also See for HeartMate III:

Advertisement

Quick Links

Priority
HA Z ARD
System Controller
Active
Screen
Symbols
+
+
+
+
+
+
+
IMPORTANT! The Pump Running (
Thoratec Corporation continually strives to provide the highest quality
products for mechanical circulatory support. Specifications may change
without notice. Therefore, please refer to the HeartMate III Instructions for
Use shipped with the HeartMate III for the most current information regarding
indications, contraindications, warnings, and cautions. Thoratec, the Thoratec
logo, and HeartMate are registered trademarks, and HeartMate III, Mobile
Power Unit, and HeartLine are trademarks of Thoratec Corporation.
©2017 Thoratec Corporation. 10009740.A 10/2017
Alarm Means
1 Check if the fixed speed setting
is below 4,000 rpm AND the
System Controller's backup battery
is not installed. Under these
Pump is off.
conditions, the pump can only
The Pump Running
be started from the System Monitor's
symbol is black.
Clinical or Settings screen by
pressing the Pump Start button.
Otherwise, press any button on
the System Controller to attempt
pump start.
1 Ensure that the Driveline is
connected to the System Controller.
Low flow, flow is
less than 2.5 lpm
1 Immediately reconnect the
Driveline is
disconnected.
Driveline to the System Controller
The Pump Running
and move the Driveline safety
symbol is black.
lock on the System Controller
to the locked position. Also check
that the Modular In-Line connector
is secure.
2 If the alarm persists after
reconnecting the Driveline,
press any button on the System
Controller to attempt pump start.
Both power cables
Immediately connect to a working power source
are disconnected.
(Power Module, Mobile Power Unit
14 Volt Lithium-Ion batteries).
System Controller
No active symbols (constant audio tone).
Hardware Fault
1 Immediately switch to the backup System Controller.
(Microcontroller
2 Provide the patient with a new System Controller.
Failure)
Low Battery,
Immediately connect to a working power source
Power Input is extremely
(Power Module, Mobile Power Unit, or two fully-charged
low with less than 5 min.
HeartMate 14 Volt Lithium-Ion batteries).
remaining.
) light is always green (ON) when the pump is running.
Thoratec Corporation
6035 Stoneridge Drive, Pleasanton, CA 94588
Tel (925) 847-8600 | Fax (925) 847-8574
HeartLine (800) 456-1477
www.thoratec.com
To Resolve Alarm
2 Switch to the backup
System Controller and
attempt to restart the pump.
3 Clinically evaluate the patient.
2 Ensure that a power
source is connected to
the System Controller.
3 Clinically evaluate the patient.
3 If the alarm still persists, check
if the fixed speed setting is below
4,000 rpm AND the System
Controller's backup battery is not
installed. Under these conditions,
from the System Monitor's Clinical
or Settings screen by pressing
the Pump Start button.
4 If the Driveline is connected and
the alarm persists, replace the
System Controller with a configured
or two HeartMate
®
™ ,
European Authorized Representative
St. Jude Medical
Coordination Center BVBA. The Corporate Village
Da Vincilaan 11 Box F1. 1935 Zaventem, Belgium
Tel +32 2 774 68 11 | 24-Hr Tel +32(0)2 200 66 45
Alarms for Clinicians
For additional information and
guidelines, refer to the
HeartMate III Instructions for
Use.

Advertisement

loading
Need help?

Need help?

Do you have a question about the HeartMate III and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Thoratec HeartMate III

  • Page 1 Therefore, please refer to the HeartMate III Instructions for Tel (925) 847-8600 | Fax (925) 847-8574 Coordination Center BVBA. The Corporate Village Use shipped with the HeartMate III for the most current information regarding indications, contraindications, warnings, and cautions. Thoratec, the Thoratec HeartLine (800) 456-1477 Da Vincilaan 11 Box F1.
  • Page 2 Replace the 11 Volt Lithium-Ion backup battery. Note: If replacing the battery does not resolve the alarm, the System Controller may need to be replaced or additional steps may be required. Call Thoratec Corporation with questions. Note: If replacing the battery does not resolve...