Safety, Care, Accuracy Test And Disposal; Safety And Protection - SAN-UP BP 3ABO-H Instruction Manual

Table of Contents

Advertisement

Available languages

Available languages

Error
Description Potential cause and remedy
«
ERR 2 » Error signal
During the measurement, error signals were
detected by the cu , caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
ERR 3 » No pressure
«
An adequate pressure cannot be generated
in the cu
in the cu . A leak may have occurred. Check
that the cu is correctly connected and is
not too loose. Replace the batteries if
necessary. Repeat the measurement.
«
ERR 5 » Abnormal
The measuring signals are inaccurate and
result
no result can therefore be displayed. Read
through the checklist for performing
reliable measurements and then repeat
the measurement.*
«
HI
»
Pulse or cu
The pressure in the cu is too high (over
pressure too
300 mmHg) OR the pulse is too high (over
high
200 beats per minute). Relax for 5 minutes
and repeat the measurement.*
«
LO »
Pulse too low The pulse is too low (less than 40 beats per
minute). Repeat the measurement.*
BP 3ABO-H
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through
the information in «Section 1.» carefully.

8. Safety, Care, Accuracy Test and Disposal

Safety and protection

• This instrument may be used only for the purpose described
in this booklet. The manufacturer cannot be held liable for
damage caused by incorrect application.
• This instrument comprises sensitive components and must be
treated with caution. Observe the storage and operating
conditions described in the «Technical Speci cations» section!
• Protect it from:
− water and moisture
− extreme temperatures
− impact and dropping
− contamination and dust
− direct sunlight
− heat and cold
• The cu s are sensitive and must be handled with care.
• Only pump up the cu once tted.
EN
9