BIOTIX COBRA PIPETTE
need to be lubricated after autoclaving, except if grease
was removed during cleaning. Gravimetric checking is
recommended after every 5 autoclave cycles for single
pipettes and after 1 cycle for multichannel pipettes.
14 - SPECIFICATIONS
LabPRO pipettes are high quality. These pipettes are
compatible with universal tips.
Checking and recalibrating your pipette with the tips
that you use may be needed.
Each pipette is inspected and validated by qualified
technicians in accordance with DKSH Quality System.
DKSH declares that its manufactured pipettes comply
with the requirements of the ISO8655 standard, by type
testing.
The adjustment is carried out under strictly defined and
monitored conditions (described in the internal manu-
facturer procedure wich are based on the ISO8655).
Single models
Model
Volume
range (µL)
2µl
0.2 - 2
10µl
1 - 10
20µl
2 - 20
100µl
10 - 100
200µl
20 - 200
1000µl
100 - 1 000
5000µl
500 - 5 000
10mL
1000 - 10000 1000
24
SPECIFICATIONS - Error limits
Volume
Systematic error
(µL)
(µL)
0.2
± 0.026
2
± 0.033
1
± 0.035
10
± 0.110
2
± 0.11
20
± 0.20
10
± 0.39
100
± 0.80
20
± 0.55
200
± 1.60
100
± 3.3
1000
± 8.0
500
± 13
5000
± 33
± 33
10000
± 60
Random error
(%)
(µL)
≤ 0.013
± 13.2
≤ 0.015
± 1.7
≤ 0.013
± 3.5
≤ 0.044
± 1.1
≤ 0.033
± 5.5
≤ 0.066
± 1.0
≤ 0.11
± 3.9
≤ 0.17
± 0.8
≤ 0.22
± 2.8
≤ 0.33
± 0.8
≤ 0.7
± 3.3
≤ 1.7
± 0.8
≤ 3.3
± 2.6
≤ 8.8
± 0.7
≤ 6.6
± 3.3
≤ 17.6
± 0.6
(%)
≤ 6.6
≤ 0.8
≤ 1.3
≤ 0.4
≤ 1.7
≤ 0.3
≤ 1.1
≤ 0.2
≤ 1.1
≤ 0.2
≤ 0.7
≤ 0.2
≤ 0.7
≤ 0.2
≤ 0.7
≤ 0.2