ACL 880 Operation Manual Rev. 08/22/19 MKB ACL 880 MEGOHMMETER OPERATION MANUAL Meter is warranted for one year from the date of purchase on parts and labor. Calibration is recommended every twelve months. NOTE: THIS MANUAL IS NOT TO BE...
Page 2
Rev. 08/22/19 MKB ACL 880 Megohmmeter The ACL 880 Megohmmeter kit measures surface resistivity, resistance, temperature and humidity. It is designed to test conductive, dissipative, and insulative surfaces for electrical resistivity/resistance according to EOS/ESD, CECC, ANSI, ASTM and UL test procedures.
Page 3
Rev. 08/22/19 MKB INTRODUCTION The ACL 880 Megohmmeter is a dependable and easy-to-use audit kit for conductive and dissipative surfaces. This meter is designed to be used in all facets of material production including engineering, maintenance, quality control, incoming inspection, manufacturing, research, or sales departments for the testing of anti-static mats, floor finishes, paints, wrist straps, smocks, footwear, bags and containers.
Page 4
¾ inch spacing to accommodate dual banana plug connectors. These connectors are used for microprobes resistance measurement fixtures. Leads supplied with the ACL 880 are specially designed for this instrument. They have superior outer insulation and a high number of stranded conductors to minimize measurement errors.
RECOMMENDED USAGE ACL 880 instruments and fixtures shall be operated at 65 – 90° Fahrenheit (18 – 32° Celsius), 10 – 60 % Relative Humidity. ACL 880 recommends making measurements using its instruments and fixtures in an environment of 72 – 76° Fahrenheit (22 – 24° Celsius), 25 – 35% Relative Humidity.
Page 6
ACL 880 Operation Manual Rev. 08/22/19 MKB b. To start measurement in GO C, press & release TEST pad. Measurement sequence is conducted automatically and ends only when MODE pad is depressed and stops measurement. Green HOLD LED will illuminates. Yellow LED lights correspond to the voltage applied.
(0) within 5% of 1.0x10e12 ohms at 100 volts, plus 5 seconds. The EP is used for resistance measurements equal to, or greater than 1.0x10e6 ohms. The 880 was designed to incorporate variable EP to allow fastest possible, accurate measurements. Typical EP up to less than 1.0x10e9 ohms is typically less than 8.0 seconds.
The ACL 880 measurement process is quite fast and once a measurement is stable in the LCD, the operator may elect to use the stable displayed measurement. In this case the EP is manually determined by the operator.
OTHER THEORY. OUT OF WARRANTY REPAIRS ACL Staticide expressly warrants that for a period of 90 Days from the date of repair, by ACL Staticide, instruments and products will be free of defects in material (parts) and workmanship (labor). If ACL Staticide receives notice of such defect during the warranty period, ACL Staticide will replace or repair at its expense such parts which it determines to be defective.
Rev. 08/22/19 MKB LIMIT OF LIABILITY In no event will ACL Staticide or any authorized reseller be responsible or liable for special, incidental, or consequential losses or damages, under any legal theory including but not limited to contract, negligence, or strict liability. Fulfillment by ACL, Inc. of its express warranty obligations described above will be Buyer’s exclusive remedy and will be ACL, Inc.