ACL 395 Operation Manual 
ACL 395 Resistivity Meter 
Meter is warranted for one year from the date of purchase on parts and labor.  
 
 
 
 
ACL Incorporated  
th
840 W. 49
 Place 
Chicago, IL 60609 USA 
OPERATION MANUAL 
Calibration is recommended every 12 months. 
Page 1 of 13 
 
 
 
 
Rev. 08/18/17 MKB 
Tel: 847‐981‐9212 
info@aclstaticide.com  
www.aclstaticide.com 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 395 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ACL 395

  • Page 1 ACL 395 Operation Manual    Rev. 08/18/17 MKB  ACL 395 Resistivity Meter  OPERATION MANUAL  Meter is warranted for one year from the date of purchase on parts and labor.   Calibration is recommended every 12 months.            ACL Incorporated     Tel: 847‐981‐9212  840 W. 49  Place  Page 1 of 13  info@aclstaticide.com   Chicago, IL 60609 USA    www.aclstaticide.com ...
  • Page 2 ACL 395 Operation Manual    Rev. 08/18/17 MKB    ACL 395 Resistivity Meter    The ACL 395 is a lightweight, pocket‐sized, auto‐ranging surface resistivity meter.  It is designed  to test conductive, dissipative, and insulative surfaces for electrical resistivity according to the  ESDA’s parallel resistivity probe method DIN EN 100 015/1 & ASTM D257.      If the meter is used with the 5‐lb probes available in the accessory kit, it will comply to  IEC 61340‐4‐1, ANSI/ESDA S4.1 and ANSI/ESDA S7.1.  ACL 395 Resistivity Meter includes:   * Tester  * Two accordion cables (stereo to banana)  * 9‐volt battery  * Certificate of calibration    ACL 396 Accessory Kit includes:  * Two 5‐lb probes  * Foam‐lined carrying case    Limits  * Resistivity: 10 ‐10  ohms/square  * Resistance: 10 ‐10  ohms  * Measuring voltage: 10v and 100v     ACL Incorporated     Tel: 847‐981‐9212  840 W. 49  Place  Page 2 of 13  info@aclstaticide.com   Chicago, IL 60609 USA   ...
  • Page 3 ACL 395 Operation Manual    Rev. 08/18/17 MKB      NTRODUCTION   The ACL 395 Resistivity Meter is an easy‐to‐use tester for measuring surface resistivity.  When  used with the ACL 396 Accessory Kit, the ACL 395 Resistivity Meter is a dependable audit kit for  conductive and dissipative surfaces.  This meter is designed for use in all facets of material  production including engineering, maintenance, quality control, incoming inspection,  manufacturing and research, or in sales departments for the testing of anti‐static mats, floor  finishes, paints, wrist straps, smocks, footwear, bags and containers.      When using the built‐in probes, the meter’s test values for surface resistivity are in ohms per  square (although they are displayed in ohms). When using the external 5‐lb probes, the meter’s  test values for resistance are in ohms.       ECADE  CALE    3  =  1 kilohm  4  =  10 kilohms  5  =  100  kilohms  5  3 x 10 =  300  kilohms  6  =  1 meg ohm  6  3 x 10 = ...
  • Page 4: Test Voltage

    ACL 395 Operation Manual    Rev. 08/18/17 MKB      EST  OLTAGE    The test voltages are 10v and 100v.  According to ESD Association (ESDA) standards S4.1 and  S7.1, 10v should be applied to surfaces with resistivity of less than 10  and 100v should be  applied to surfaces with resistivity of 10  or greater.   The ACL 395 will automatically simulate  the proper voltage according to the test measurement.     As defined by the ESDA, values indicate the following:      Voltage  Range   Definition     10 volt  < 10  ohms per square  Conductive    100 volt   – 10  ohms per square  Dissipative    100 volt  > 10  ohms per square  Insulative         ...
  • Page 5: Recording Data

    ACL 395 Operation Manual    Rev. 08/18/17 MKB    ECORDING  ATA    ANSI/ESDA and European CECC recognize the effect environmental conditions have on test  measurements and specify in their standards that they measured and recorded. It is possible to  test or manufacture a material at high humidity and pass all the test specifications, but when  the customer receives the material and uses it at a lower humidity or temperature the material  fails to pass the specifications. This can cause rejections and loss of product.     Both ESD S4.1 ESD Protective Worksurface sections 6.2.4 and ESD S7.1‐1994 Resistive  Characterization of Materials Floor Materials sections 5.2.4 and 5.3.3 require reporting of  temperature and humidity at the time of testing.         ALIBRATION   Calibration is recommended annually. The ACL 395 meter comes with a certificate of calibration  which verifies the calibration of the instrument using equipment that is traceable to National  Standards and CAD‐generated techniques. Meters can be sent back to ACL Inc. for calibration for  a lab fee or they can be sent to a certified calibration lab.  (See page 12 for calibration  instructions.)  The meters also come with a CE mark approval.        EASURING WITH  NTERNAL  ROBES    The parallel resistivity probe method complies with ASTM D257. It is used to give fast electrical  resistivity measurements on flat homogeneous materials.  It may be used on multilayered  materials, but this should be noted along with the temperature and humidity values on the data  report.       ...
  • Page 6: External Probes

    ACL 395 Operation Manual    Rev. 08/18/17 MKB  When the measurement is taken between the meter’s two rubber rails, the meter will indicate  the surface resistivity of the material being tested.     A. Prior to testing, make certain that surfaces to be tested are clean and free of  contaminants.    B. Allow the meter to acclimate to the atmosphere in which it will be used.  Adjustment to  new environmental conditions may take as long as a half hour.    C. Place the meter on the desired surface to be tested.    D. Press and hold the red integrated test  button with approximately five pounds  of applied force. The meter will display  the surface resistivity.     E. When using the built‐in probes, the  meter’s test values for surface resistivity  are in ohms per square (although they  are displayed in ohms).    F. The test value is indicated on the LED  display.  A decade will brighten to the  according test value.  Colors signify the  test value’s function (see page 3).      EASURING WITH  XTERNAL  ROBES  ...
  • Page 7 ACL 395 Operation Manual    Rev. 08/18/17 MKB  By connecting the 5‐lb probes to the ACL 395’s sockets, it is possible to measure Resistance  Point to Point (RTT), Resistance to Ground (RTG), and Volume Resistance.  Using these external  probes will allow compliance with various standards including ANSI/ESDA S4.1 for Worksurface –  Resistance Measurements, ANSI/ESDA S7.1 Resistive Characterization of Materials – Floor  Materials.      When auditing is finished, unplug the cables and store probes in the protective case.  Resistance Point to Point (RTT)    RTT measurements can be used for the evaluation of floors, chairs, carts, work surfaces and  other ESD‐controlled materials and products. Procedures vary regarding sample preparation,  probe preparation and spacing of the 5‐lb probes.  Select and read the correct test procedure or  standard for the desired measurement. Test procedures can be purchased from www.esda.org.      A. Connect one end of each banana test lead into the sockets of the meter.  Connect the  other end of the test coil cords into the 5‐lb probes.                                                                                         B. Place both probes on the material according to test procedures or standard being used.  C. Press the “TEST” button and the value will be displayed on the LEDs. The meter will  apply the correct voltage (10v or 100v) according to the value of what is measured. ...
  • Page 8 ACL 395 Operation Manual    Rev. 08/18/17 MKB      N  XAMPLE OF  EASURING   ON  ISSIPATIVE  LOORING           Taking routine measurements of tiles coated with dissipative floor finish is a key component of  proper floor maintenance.  Any problems that arise with the floor will be easily detected.  Keeping  a record of temperature, humidity and electrical properties will provide a reference, and will point  toward a blueprint of traffic patterns on the floor.  Good record keeping will ensure success when  developing and maintaining a maintenance program.      To obtain an average measurement of a floor, map out a 4’ x 4’ section and conduct five tests  (one at a time) within the square.  Conduct a test for each side of the square and a final test  diagonally through the center as shown in the drawing below.      In each RTT test, the 5‐lb probes are placed three feet apart (36 inches).  Connect the test leads  to the meter.  Attach a 5‐lb probe to the end of each lead and place three feet apart as indicated  below.  Press and hold the “TEST” button on the ACL 395 meter until a value is displayed.              RTT #1           ...
  • Page 9 ACL 395 Operation Manual    Rev. 08/18/17 MKB  (RTG)  EASURING  ESISTANCE TO  ROUND    RTG measurements can be used for the evaluation of floors, chairs, carts, work surfaces and  other ESD‐controlled materials and products.  Keeping a record of test results for temperature,  humidity and electrical properties will provide a reference.    :  N  XAMPLE OF  EASURING   ON  ISSIPATIVE  LOORING   For Testing Resistance on Floors, S7.1 requires a minimum of five RTG tests per 5,000 sq. ft.   Connect the leads for the external probes to the meter.           When the cables have been plugged into    the appropriate socket, the parallel    probes under the meter disengage.                ...
  • Page 10 ACL 395 Operation Manual    Rev. 08/18/17 MKB  :  N  XAMPLE OF  EASURING   ON  ISSIPATIVE  ABLE    A. To test RTG on a workstation, connect the first lead to the meter and to a 5‐lb probe.  Place  the probe on the work surface.  Connect the second lead to the meter and to a groundable  point (RTGP).     B. To attach lead to RTGP, slip the alligator clip to the lead and connect it to the ground snap or  connect the banana plug to a common point ground plug:                  b. Press the “TEST” button and the value will be displayed on the LED.  While performing test,  do not touch lead wires or probe. Avoid overlapping of lead wires. This will ensure accurate  readings.     c. Resistance values are in ohms. When recording test values also make note of the  temperature and humidity, as the environment can affect test results.          To Ground           ...
  • Page 11 ACL 395 Operation Manual    Rev. 08/18/17 MKB  Volume Resistance Measurement     Volume Resistance measures the electrical path through a material.       A.   Connect one end of each of the banana test leads into the sockets of the meter.   Connect the other end of the test coil cords into the 5‐lb probes.    B.  Place sample material on a conductive metal plate (such as stainless steel).  Place one of  the 5‐lb probes onto the material so that the material is sandwiched between the probe  and metal plate. (See below)     C.   Place the second 5‐lb probe on the conductive metal plate.    D.  Press the “TEST” button and the value will be displayed on the LED.  Volume Resistance  is measured in ohms‐cm.                                                                                      ...
  • Page 12: Calibration Instructions

    ACL 395 Operation Manual    Rev. 08/18/17 MKB        ALIBRATION  NSTRUCTIONS This meter should be calibrated every twelve months.  To verify if the tester is within  specification, a test resistance can be applied across the parallel bars using a resistance decade  box.  Calibration can be obtained by contacting ACL Incorporated.    PECIFICATIONS Dimensions:     70mm x 130mm x 35mm (approximate)    Weight:        103 grams (approximate)      Power:      Battery‐operated PP3 9‐volt    Connections:     2 x 3.5mm jack plug for earth connection    3   12+  Test Range:      to 10      5  9  Dissipative Range:  ...
  • Page 13: Test Equipment

    ACL 395 Operation Manual    Rev. 08/18/17 MKB    EST  QUIPMENT  * Resistance Decade Box  * Test Leads  The resistance decade box required will need a range of from 1 kilohm (10 ) to 999 meg ohms  ). Measurements greater than 10  can be calculated using CAD‐generated techniques, as  high resistances greater than 10  are difficult to verify with a test voltage of 9 volts unless 1%  10  resistors are used.  If using a decade box with resistors in the upper range (10 – 10 ), resistors  must be 1%.    Connect the test leads from the resistance decade box to the test probes of the meter, set the  decade box to the desired resistance i.e. 10  = 1 K, then press and hold the meter’s test button.  The 10  LED should light, 10 K 10  LED should light and so on. To measure the changeover point  between decades, increase the resistance of the decade box while pressing the meter’s test  button. Record the resistance when the next LED lights permanently (this is the changeover  resistance).  Example: The first green LED is illuminated 10  = 1 kilohm.  At 3 or 4 kilohms, 10  LED is  illuminated.  The changeover point is 3 or 4 kilohms. 10  = 10 kilohms so between 3 or 4 kilohms  and 30 or 40 kilohms will be the changeover points from 10  to 10 .     Please note that the meter has no internal parts to adjust, so verification of calibration   can be achieved by using the above process. If verification cannot be achieved the unit   should be returned to the supplier.