Electrification Period - ACL 880 Operation Manual

Megohmmeter
Table of Contents

Advertisement

ACL 880 Operation Manual
ACL 880 in Auto Default
Operation Mode
5. The MODE button performs four useful functions:
a. Select mode of operation from AUTO (Go A) to Continuous (GO C) by pressing
MODE pad and TEST pad simultaneously.
b. Select manual test voltage for AUTO, or CONTINUOUS mode (see "Manual" section
above).
c. Bypass the automatic HOLD function and override displayed data by resetting the
instrument for the next measurement. The existing measurement in the display will
be discarded.
d. Initiate CAL? function. Calibration function is only to be used by qualified ACL
Staticide® trained technicians during calibration. Exit out of CAL? and do not try to
adjust.

ELECTRIFICATION PERIOD

Electrification Period (EP) is the time required for an instrument to make a measurement from
zero (0) within 5% of 1.0x10e12 ohms at 100 volts, plus 5 seconds. The EP is used for resistance
measurements equal to, or greater than 1.0x10e6 ohms. The 880 was designed to incorporate
variable EP to allow fastest possible, accurate measurements. Typical EP up to less than
1.0x10e9 ohms is typically less than 8.0 seconds. However, measurement conditions in the
factory environment vary considerably. To compensate for variables in materials and conditions,
ACL Incorporated
th
840 W. 49
Place
Chicago, IL 60609 USA
ACL 880 in Continuous
Operation Mode
Page 7 of 10
Rev. 08/22/19 MKB
Tel: 847-981-9212
info@aclstaticide.com
www.aclstaticide.com