ACL 800 Operation Manual 
ACL 800 MEGOHMMETER 
Meter is warranted for one year from the date of purchase on parts and labor. 
Calibration is recommended every twelve months. 
NOTE: THIS MANUAL IS NOT TO BE    
 
 
 
 
ACL Incorporated  
th
840 W. 49
 Place 
Chicago, IL 60609 USA 
OPERATION MANUAL 
SEPARATED FROM METER!
Page 1 of 13 
 
 
 
 
 
Rev. 07/26/22 MKB 
 
Tel: 847‐981‐9212 
info@aclstaticide.com  
www.aclstaticide.com

Advertisement

loading
Need help?

Need help?

Do you have a question about the 800C and is the answer not in the manual?

Questions and answers

Summary of Contents for ACL 800C

  • Page 1 ACL 800 Operation Manual    Rev. 07/26/22 MKB  ACL 800 MEGOHMMETER  OPERATION MANUAL      Meter is warranted for one year from the date of purchase on parts and labor.  Calibration is recommended every twelve months.  NOTE: THIS MANUAL IS NOT TO BE             SEPARATED FROM METER!   ACL Incorporated     Tel: 847‐981‐9212  840 W. 49  Place  Page 1 of 13  info@aclstaticide.com   Chicago, IL 60609 USA    www.aclstaticide.com...
  • Page 2 ACL 800 Operation Manual    Rev. 07/26/22 MKB  ACL 800 Megohmmeter    The ACL 800 Megohmmeter kit measures surface resistivity, resistance, temperature  and humidity.  It is designed to test conductive, dissipative, and insulative surfaces for  electrical resistivity/resistance according to EOS/ESD, CECC, ANSI, ASTM and UL test  procedures.     The internal parallel electrodes comply with DIN EN 100 015/1 and ASTM D257.  The 5‐ lb electrodes can be externally connected for tests according to IEC 61340‐4‐1,  ANSI/ESDA S4.1 & ANSI/ESDA S7.1.   Optional concentric ring probe will allow  compliance to ANSI/ESDA S11.11.    ACL 800 Megohmmeter Includes:   * Tester  * Two 5‐lb probes  * One ground cord  * Two accordion cables (monaural to banana)  * 110‐volt power adapter (in the US only)  * 9‐volt battery  * Foam‐lined carrying case  * Certificate of calibration    Limits  * Resistivity: 10  ‐10  ohms/sq.  * Resistance: 10  ‐10  ohms  * Relative Humidity: 10% ‐ 90%   * Temperature: 32°F to 100°F or 0°C – 37.7°C if Celsius model was purchased  * Measuring voltage: 10v and 100v     ACL Incorporated     Tel: 847‐981‐9212  840 W. 49  Place ...
  • Page 3 ACL 800 Operation Manual    Rev. 07/26/22 MKB  INTRODUCTION      The ACL 800 Megohmmeter is a dependable and easy‐to‐use audit kit for conductive  and dissipative surfaces.  This meter is designed to be used in all facets of material  production including engineering, maintenance, quality control, incoming inspection,  manufacturing, research, or sales departments for the testing of anti‐static mats, floor  finishes, paints, wrist straps, smocks, footwear, bags and containers.      When using the built‐in electrodes, the meter’s test values for surface resistivity are in  ohms per square.  When using the external 5‐lb electrodes, the meter’s test values for  resistance are in ohms (although they are displayed in ohms/sq.).         ECADE  CALE    3  =  1 kilohm  4  =  10 kilohms  5  =  100  kilohms  6  =  1 meg ohm  7  =  10 meg ohms  8  =  100 meg ohms  9 ...
  • Page 4 ACL 800 Operation Manual    Rev. 07/26/22 MKB     OTE ABOUT  OLTAGE     In previous years, those desiring to measure resistivity or resistance followed the ASTM  D264, ASTM 991, NFPA 56A or NFPA 99 test standards. These procedures required one  to test at either 500 or 1000 volts. This caused concern regarding safety to the person  performing the tests. The ESDA standardized the test procedures so that lower voltages  could be used within specific ranges.      The ACL 800 meter uses a 9‐volt battery.  Some meters with 9‐volt batteries do not give  the accuracy needed to perform the test, especially at values higher than 10 .   The ACL  800 is built with a transformer that converts the 9‐volt charge from the battery to 10  volts or 100 volts (whichever value is selected).  The meter applies a constant charge  over the complete voltage range. Accuracy depends on applied voltage, temperature,  and humidity.      EMPERATURE AND  UMIDITY   Humidity and temperature affect the electrical properties of the material being tested.  The combination of low humidity and low temperature will give the highest electrical  resistance results and slowest dissipation times. At high humidity, a thin layer of water is  condensed on or absorbed into the material being tested. This is true of hydroscopic  additives that are incorporated into a material in order to increase the electrical  conductivity.  These additives will allow moisture to be absorbed into the materials to  which they are added.      At elevated temperatures, the mobility of free electrons increases, thereby also  increasing a material’s conductivity. This is especially true for carbon black, metallic  oxides, metals, and other substances. When the material is at a lower temperature,  built‐in stresses occur which may increase the resistance due to increased distance  between the conductive additives. Thus, humidity and temperature must be known.  ...
  • Page 5 ACL 800 Operation Manual    Rev. 07/26/22 MKB    ALIBRATION   Calibration is recommended annually. The ACL 800 meter comes with a NIST certificate  when ordered. After one year the meters can be sent back to ACL Inc. for a new NIST  certificate for a lab fee or it can be sent to a certified calibration lab (see page 11 for  instructions). The meters also come with a CE mark approval.      MEASURING WITH INTERNAL ELECTRODES     The parallel resistivity probe method complies with ASTM D257. It is used to give fast  electrical resistivity measurements on flat homogeneous materials.  It may be used on  multilayered materials, but this should be noted along with the temperature and  humidity values on the data report.    When the measurement is taken between the tester’s two conductive rubber rails  under the tester, the tester will indicate the surface resistivity of the material being  tested.     A. Prior to testing, make certain that surfaces to be tested are clean and free of  contaminants.    B. Allow the meter to acclimate to the atmosphere in which it will be used.   Adjustment to new environmental conditions may take as long as a half hour.    C. Place the meter on the desired surface to be tested.    D. Move switch to desired test voltage position, either 10 or 100 volts, depending  on the target range of the material.    E. Press and hold the test button with approximately five pounds of applied force.  ...
  • Page 6 ACL 800 Operation Manual    Rev. 07/26/22 MKB    MEASURING WITH EXTERNAL ELECTRODES           When the measurement is taken using the 5‐lb external probes, the tester will indicate  the resistance of the material being tested once the button is pressed (value is in ohms,  though the display says ohms/sq).      When a probe is plugged into the  appropriate socket, the associated   internal electrode is disengaged.    ▲ ▲ ▲ ▲ 12VDC External Electrodes    By connecting these probes to the ACL 800’s sockets it is possible to measure Point to  Point (RTT) Resistance, Surface to Ground (RTG), and Volume Resistance.  Using these  probes will allow compliancy with various standards including ANSI/ESDA S4.1 for  Worksurface – Resistance Measurements, ANSI/ESDA S7.1 Resistive Characterization of  Materials – Floor Materials.      When auditing is finished, unplug the cables and store probes in the protective case.  If  cables are left in the tester, the jacks will lose their elasticity as shown below:       ...
  • Page 7 ACL 800 Operation Manual    Rev. 07/26/22 MKB      Measuring Resistance Point to Point (RTT)    RTT measurements can be used for the evaluation of floors, chairs, carts, work surfaces  and other ESD‐controlled materials and products. Procedures vary regarding sample  preparation, probe preparation and spacing of the 5‐lb probes.  Select and read the  correct test procedure or standard for the desired measurement. Test procedures can  be purchased from www.esda.org.      A. Connect one end of each of the banana test leads into the sockets of the meter.   Connect the other end of the test coil cords into the 5‐lb probes.                                                                                                       A. Place both probes on the material according to test procedures or standard ...
  • Page 8 ACL 800 Operation Manual    Rev. 07/26/22 MKB      N EXAMPLE OF MEASURING   ON DISSIPATIVE FLOORING           Taking routine measurements of tiles coated with dissipative floor finish is a key  component of proper floor maintenance. Any problems that arise with the floor will be  easily detected.  Keeping a record of test results for temperature, humidity and electrical  properties will provide a reference and will point toward a blueprint of traffic patterns on  the floor.  Good record keeping will ensure success when developing and maintaining a  maintenance program.      To get an average measurement of a floor, map out a 4’ x 4’ section and conduct five  tests (one at a time) within the square.  Conduct a test for each side of the square and a  final test diagonally through the center as shown in the drawing below.      Each RTT test utilizes the 5‐lb probes placed three feet apart (36 inches).  Connect the  test leads to the meter.  Attach a 5‐lb probe to the end of each lead and place three feet  apart as indicated above.  Press and hold the “TEST” button on the ACL 800 meter until  a value is displayed.        RTT #1                  ...
  • Page 9 ACL 800 Operation Manual    Rev. 07/26/22 MKB  (RTG)  EASURING  ESISTANCE TO  ROUND    Resistance to ground measurements can be used for the evaluation of floors, chairs,  carts, work surfaces and other ESD controlled materials and products.  Keeping a record  of test results for temperature, humidity and electrical properties will provide a  reference.    :  N EXAMPLE OF MEASURING   ON DISSIPATIVE FLOORING   For Testing Resistance on Floors, S7.1 requires a minimum of five RTG tests per 5,000 sq.  ft.  Connect the leads for the external probes to the meter.   ▲ ▲ ▲ ▲ When the cables have been plugged in  12VDC External Electrodes the appropriate sockets, the parallel  probes under the meter disengage.     Attach one lead to a 5‐lb probe and place probe on the floor that’s being tested.  Attach  the other lead to an alligator clip and connect to ground or a groundable point (RTGP).   If using a ground adapter plug, plug the banana lead into the adaptor after the adapter  is plugged into the receptacle.       ...
  • Page 10 ACL 800 Operation Manual    Rev. 07/26/22 MKB      :  N EXAMPLE OF MEASURING   ON DISSIPATIVE TABLE MATS   A. To test RTG for a workstation, connect the first lead to the meter and to a 5‐lb  probe.  Place the probe on the work surface.  Connect the second lead from the  meter directly to a ground source.   If connecting to a groundable point, the test is  “resistance to groundable point” (RTGP) and not “resistance to ground” (RTG). Use  the alligator clip if necessary.     B. Press the “TEST” button and the value will be displayed on the LCD.  When  performing test do not touch lead wires or probe. Avoid overlapping of lead wires.  This will ensure accurate readings.     C. Resistance values are in ohms. Record temperature, humidity and resistance.      6  If the range of your work surface is 10 to 10 , measure at 100v.  The EOS standard says  6  11  6  9  the dissipative range is 10 – 10 but commonly 10 – 10 is considered a pass.       ...
  • Page 11 ACL 800 Operation Manual    Rev. 07/26/22 MKB        To Ground                                 Connect alligator clip to snap for RTGP                     ACL Incorporated     Tel: 847‐981‐9212  840 W. 49  Place  Page 11 of 13  info@aclstaticide.com   Chicago, IL 60609 USA    www.aclstaticide.com...
  • Page 12 ACL 800 Operation Manual    Rev. 07/26/22 MKB    EASURING  OLUME  ESISTANCE   Volume Resistance measures the electrical path through a material.       A.   Connect one end of each of the banana test leads into the sockets of the meter.   Connect the other end of the test coil cords into the 5‐lb probes.    B.  Place sample material on a conductive metal plate (such as stainless steel).  Place  one of the 5‐lb probes on the material so that the material is sandwiched  between the probe and metal plate. (See below)     C.   Place the second 5‐lb probe on the conductive metal plate.    D. Press the “TEST” button and the value will be displayed on the LCD.  Volume  Resistance is in ohms‐cm.                                                                                                                                                   ...
  • Page 13 ACL 800 Operation Manual    Rev. 07/26/22 MKB  CALIBRATION INSTRUCTIONS      The ACL 800 meter is calibrated to be most accurate in the most widely‐used range (10  ‐10 ).   The lab calibrates meters between 30 ‐ 40% RH at 65 ‐ 73°F.  If you are using the meter in  atmospheres different than above, it is strongly advised to recalibrate the meter to your  conditions.       1.  Purchase  1%  10 ‐10   ohm  resistors,  high‐accuracy  relative  humidity  hygrometer,  and  high‐accuracy thermometer.   2.  Open meter, being careful not to disturb or break the two wires connecting the power  button to the circuit board.  3.  Observe on the right lower side of the meter printed circuit board four calibration pots.  4.  ALLOW  THE  METER  TO  EQUILIBRATE  AND  NORMALIZE  IN  THE  ENVIRONMENT  FOR  2 ...