Troubleshooting - BHM Medical V3 Instructions For Use Manual

Table of Contents

Advertisement

Troubleshooting

WARNING: Do not attempt to open the V3 portable lift
cassette. Only a qualified technician is authorized to open it.
Alterations made to the portable lift by someone other than a
qualified technician may cause serious injury.
Problem
The red "service" light is on and
flashing.
The unit starts and stops
repetitively.
The lift emits a "beep" during use.
The unit may stop lifting the patient
but can still lower the patient.
Charger indicator on the lift
(yellow) does not light up when the
charger is connected to the lift.
The lift does not work when you
press the buttons on the handset.
Contact your local BHM Medical representative for maintenance.
If the load is greater than the unit's safe working load, it will not func-
tion due to an overload protection on the motor.
Batteries are low and need to be changed. Contact your local BHM
Medical representative.
Batteries are low and need to be recharged.
Check that the charger is plugged into a standard electrical outlet, and
that the outlet has power. The green indicator light confirms the func-
tionality of the charger.
If the emergency stop is activated, gently release the emergency but-
ton.
Check if the buttons on the lift are working. If they are working, the
problem may come from the handset. If they are not working, check
the charge on the lift.
Check to see if the handset is plugged in properly into the lift. The
handset may be slightly pulled out from its socket and yet appear as
though it is plugged in. Check the connection.
If, after testing all of the above, the lift will not operate, contact your
BHM Medical representative.